1 / 27

VENTILATION PAR OSCILLATIONS A HAUTE FREQUENCE EN REANIMATION CHEZ L’ADULTE

VENTILATION PAR OSCILLATIONS A HAUTE FREQUENCE EN REANIMATION CHEZ L’ADULTE. Julien Bordes, le 25 janvier 2010 DESC Réanimation médicale. DEFINITIONS. VENTILATION PAR OSCILLATIONS A HAUTE FREQUENCE = HFO = HIGH FREQUENCY OSCILLATORY VENTILATION VENTILATION A HAUTE FREQUENCE: > 100/min

navid
Download Presentation

VENTILATION PAR OSCILLATIONS A HAUTE FREQUENCE EN REANIMATION CHEZ L’ADULTE

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. VENTILATION PAR OSCILLATIONS A HAUTE FREQUENCE EN REANIMATION CHEZ L’ADULTE Julien Bordes, le 25 janvier 2010 DESC Réanimation médicale

  2. DEFINITIONS • VENTILATION PAR OSCILLATIONS A HAUTE FREQUENCE = HFO = HIGH FREQUENCY OSCILLATORY VENTILATION • VENTILATION A HAUTE FREQUENCE: > 100/min • Fessler H et al, Resp Care 2007 • AUTRES TYPES DE VENTILATION A HAUTE FREQUENCE (HF) • VENTILATION PAR PRESSION POSITIVE A HF • VENTILATION PAR JET A HF • VENTILATION PAR PERCUTION A HF

  3. HFO: PRINCIPES • INSUFFLATION A HAUTE FREQUENCE DE VOLUMES COURANTS INFERIEURS A L’ESPACE MORT ANATOMIQUE • MAINTIEN D'UN NIVEAU CONSTANT DE PRESSION ALVEOLAIRE VENTILATEUR 3100B SENSORMEDICS

  4. RATIONNEL • CONCEPT DE VENTILATION PROTECTRICE • FAIBLE VOLUME COURANT (VT) • =LIMITER LA SURDISTENSION ALVEOLAIRE • RECRUTEMENT ALVEOLAIRE : « OPEN LUNG » • =LIMITER L’OUVERTURE/FERMETURE ALVEOLAIRE

  5. HFO: MECANISMES OSCILLATION DE LA MEMBRANE  Vt 3-10 Hertz Puissance 20-40 L/min D’après D. Jallal et al, Réanimation 2004 PRESSION DE DISTENSION 25-35 CM H20

  6. HFO: MECANISMES Pression cm H2O 20- 15- 10- 5- 0- ----------PIC -----------------------Paw ------------------------Paw ---------PIC PEP VMC HFO

  7. HFO: MECANISMES • OXYGENATION ET VENTILATION SONT REGLEES INDEPENDEMMENT • DETERMINANT DE L’OXYGENATION ARTERIELLE • PRESSION MOYENNE • FIO2 • DETERMINANT DE L’EPURATION DE CO2 • FREQUENCE • AMPLITUDE

  8. HFO: MECANISMESECHANGES GAZEUX 2 dispersion de Taylor 4 asymetrie des profils de vitesse 1 ventilation alvéolaire directe par convection 5 oscillations cardiogéniques 3 convection par effet Pendeluft D’après D. Jallal et al, Réanimation 2004 6 diffusion moléculaire de l’O2 et du CO2

  9. TRAVAUX CLINIQUES • HFO = mode de ventilation dans le SDRA • Technique alternative, « rescue ventilation » • HFO versus VMC? • Peu de données…

  10. TRAVAUX CLINIQUES HISTORIQUEMENT • EN NEONATALOGIE ET EN PEDIATRIE • Elective high frequency oscillatory ventilation versus conventional ventilation for acute pulmonary dysfunction in preterm infants. • Cochrane Database Syst Rev. 2009 Jul 8;(3):CD000104. • DIMINUTION DE L’INCIDENCE DES PATHOLOGIES PULMONAIRES CHRONIQUES

  11. CHEZ L’ADULTE AU COURS DU SDRAETUDES DE COHORTE 1997 2001 2003 2000 IO = index d’oxygénation = [FIO2 x mean airway pressure x 100]/PaO2)

  12. ETUDES PROSPECTIVES RANDOMISEES HFO VS VMC • Etude prospective randomisée multicentrique • Objectif = équivalence des 2 techniques • 148 patients • HFO n=75 • Pmoy = Pmoy en VMC + 5cm H2O, F=5 Hz • VMC n=73 • Pression contrôlée, Vt = 6-10ml/kg

  13. PAFI Pmoy Devenir à J30

  14. ETUDES PROSPECTIVES RANDOMISEES HFO VS VMC Critical Care, 2005 • Etude prospective randomisée multicentrique • Objectif = pronostic à J30 • 61 patients • HFO n=37 • Pmoy = Pmoy VMC + 5cm H2O, F=5 Hz • VMC n=24 • Vt = 8ml/kg

  15. HFO OXYGENATION ET INFLAMMATION 2005 • Etude prospective randomisée • Comparer HFO vs DV • Echanges gazeux • Inflammation • 39 patients • 3 groupes: • HFO DD • HFO DV • VMC DV

  16. HFO OXYGENATION ET INFLAMMATION 2005 HFO DV VMC DV HFO DD

  17. 2005

  18. HFO ET DECUBITUS VENTRAL • Hypothèse: HFO prévient la baisse du PaFi après DV • Etude prospective randomisée • Comparer HFO vs VMC après DV • 43 patients • 3 groupes: • DV en VMC puis DD en VMC n=15 • DV en VMC puis DD en HFO n=13 • DD en VMC puis DD en HFO n=15 2007

  19. HFO ET DECUBITUS VENTRAL 2007 L’HFO après le DV permet de maintenir le bénéfice du recrutement alvéolaire « OPENING THE LUNG AND KEEPING THE LUNG OPEN »

  20. SYNTHESE DES TRAVAUX CLINIQUESHFO et SDRA • Amélioration des échanges gazeux • Utilisation précoce de l’HFO • Absence de réponse = surmortalité • Complémentaire du décubitus ventral • Effet sur l’inflammation? • Effet sur la mortalité? 2004

  21. HFO: PRECAUTIONS D’EMPLOI • SEDATION PROFONDE ET CURARISATION • RISQUE DE PNEUMOTHORAX • Controversé • 21,8%!!! CHEST 2004

  22. HFO: PRECAUTIONS D’EMPLOI • RETENTISSEMENT HEMODYNAMIQUE • ASSOCIATION AVEC D’AUTRES TECHNIQUES • Décubitus ventral • Monoxyde d’azote Varkul et al , Successful Use of Combined High-frequency Oscillatory Ventilation, Inhaled Nitric Oxide, and Prone Positioning in the Acute Respiratory Distress Syndrome, Anesthesiology 2001.

  23. L’HFO EN PRATIQUE • HFO si échec de la VMC • Quels critères? • FIO2 > 60% et PEP = 15 cmH20 • IO > 25 • Acidose sévère avec Pplat > 30 cmH20 et FR>25/min D’après D. Jamal et al, Réanimation 2004 • Quels critères? • FIO2 > 70% et PEP = 14 cmH20 • pH < 7,25 avec vt = 6ml/kg et Pplateau > 30 cm H20 D’après H. Fessler al, Crit Care Med 2007

  24. HFO: REGLAGES Pression moyenne Pmoy VMC + 5 DeltaP = PaCO2 + 20cmH20 Débit de 40L/min TI =33% FIO2 1 Fréquence 5Hz

  25. CONCLUSION • LIMITER LES LESIONS PULMONAIRES ASSOCIEES A LA VMC • IMPACT SUR LE PRONOSTIC? • MODALITES? • Très hautes fréquences > 10Hz Fessler et al, CCM 2008 • Modalités de recrutement Pellicano et al, ICM 2009 • TECHNIQUES ADJUVANTES? • Combinaison des techniques Lubnow et al, J of Crit Care 2010

  26. Comparison of Two Methods of High Frequency Oscillatory Ventilation in Individuals With Acute Respiratory Distress Syndrome Outcome Measures: Number of ventilator free days [ Time Frame: Measured at 28 days ] Changes in plasma concentration of IL-6 [ Time Frame: Measured at 3 days ] High-Frequency Oscillatory Ventilation Associated With Inhaled Nitric Oxide in Children

More Related