An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Analyze Effects of Multifocal Intraocular Lenses on Potential Acuity Meter and Perimetry Testing G. Yuricek Aylin Sarac S.A. Erzurum, MD, FACS

  2. Financial Interests The authors have no financial interests in any of the test devices mentioned in this study.

  3. Purpose: • Evaluate the effectiveness of potential acuity meter (PAM) after multifocal lens implantation. • Evaluate effects of multifocal lenses on visual field testing.

  4. Methods: • 4 types of lenses were evaluated • ReSTOR +4, ReSTOR IQ +3, REZOOM, and monofocal lens • Monofocal lenses used were SA60AT, SN60WF, and Clarflex

  5. Methods: • All patients underwent visual exam • All patients underwent Humphrey visual field 24-2 on each eye separately. • All patients were evaluated on the PAM. • No dilation was performed on any patient.

  6. Results: • 46 eyes of 23 patients were included in the study. • 12 eyes monofocal lenses • 10 eyes ReSTOR IQ • 12 eyes ReSTOR • 12 eyes REZOOM

  7. Results: • PAM was done on all patients. Only one patient due to small pupils could not perform test. • Remaining 44 eyes all achieved 20/25 visual acuity or better on the PAM testing.

  8. Humphrey Visual Field: Examples of ring effect

  9. Visual Field Results: Monofocal Group • 8/12 eyes normal visual field • 1/12 eyes ring effects • 1/12 reduced sensitivity

  10. Visual Field Results: ReSTOR +4 Group • 8/12 eyes normal visual field • 3/12 ring effect • 1/12 superior visual field defect attributable to ptosis

  11. Visual Field Results: ReSTOR IQ Group • 8/10 eyes normal visual field • 2/10 eyes ring effect

  12. Visual Field Results: ReZOOM Group • 6/12 eyes normal visual fields • 5/12 eyes ring effect • 1/12 mild superior field defect attributable to mild ptosis

  13. Conclusion: PAM Results • Comparing monofocal lenses to ReSTOR, ReSTOR IQ, and ReZOOM lenses appeared to show no differences in results. • Thus, poor PAM results should indicate pathology for decreased visual performance.

  14. Conclusion: Visual Field • Appeared to be reliable in all groups. • Slight concern that ReZOOM lenses induce a higher incidence of ring effects. • ReSTOR and ReSTOR IQ lens appeared to be equal to monofocal lenses in visual field performance.