1 / 36

Determination of Water by the Karl Fischer Titration: Theory

Determination of Water by the Karl Fischer Titration: Theory. Program. Motivation Volumetric KF titration one an two-component reagents resolution and detection limits Coulometric KF titration cell with or without diaphragm resolution and detection limits

leigh-duffy
Download Presentation

Determination of Water by the Karl Fischer Titration: Theory

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Determination of Water by theKarl Fischer Titration: Theory

  2. Program • Motivation • Volumetric KF titrationone an two-component reagentsresolution and detection limits • Coulometric KF titrationcell with or without diaphragmresolution and detection limits • Indication, control algorithm, termination parameters • KF titration: important points • Support

  3. Why measure water or moisture? Sugar: too much moisture will not flow Flour: too little moisture dust explosion Butter: max 16.5% water content by law Drugs: too much moisture decomposition Compact Disc: too much moisture bad music quality Brake Fluid: too much water brake do not work Kerosene: too much water blocked tubing

  4. Methods for the Determination of Water Drying oven Balance with IR /Halogen / Microwave heater Thermogravimetry / DSC Spectroscopy (IR, MS) Chromatography Karl Fischer Titration

  5. Karl Fischer Titration: Why? • Fast (e.g. 1...2 minutes) • Selective for water • Accurate and precise (0.3% srel) • Wide measuring range : ppm to % Coulometric KF Volumetric KF

  6. Karl Fischer German petrochemist,1901 – 1958 Publication:1935 Bunsen reaction: 2 H2O + SO2 + I2 = H2SO4 + 2 HIPyridine happened to be around in the Lab

  7. KF Titration • KF Reaction SO2 + RN + ROH ------> (RNH)SO3R a sulfite compound (RNH)SO3R + H2O + I2 + 2RN ------> (RNH)SO4R + 2(RNH)I a sulfate compound • Summary H2O + I2 + SO2 + 3RN + ROH ----->(RNH)SO4R + 2(RNH)I • The solvent (generally methanol) is involved in the reaction • A suitable base keeps the pH 5 - 7

  8. Solvent log K 4 side reactions optimal 2 0 slow pH 2 4 6 8 10 optimal pH 5 - 7 pH range buffer needed

  9. Volumetric / Coulometric Titration + - Volumetric Karl Fischer TitrationIodine is added by burette during titration. Water as a major component:100 ppm - 100 % Coulometric Karl Fischer TitrationIodine is generated electrochemically during titration. Water in trace amounts:1 ppm - 5 %

  10. Volumetric KF Titration • Iodine is added by burette during titration. • Water as a major component: 100 ppm - 100 %

  11. Volumetric KF Titration • One - component reagent • Titrant:I2 , SO2, imidazole, methanol and diethylene glycol monoethyleter • Solvent:Methanol • Two - component reagent • Titrant:I2 and Methanol • Solvent:SO2, Imidazole, Methanol -> fast reaction, chemically stable, higher cost

  12. Volumetric KF Reagents • Titrant Concentration • 1-2-5 mg H2O/mL • Titer stability -----> Check by Standardization • Standardization materials Water 100% Sodium tartrate 15.66% Standard solution 5 mg/mL Water Standard 1% (10 mg/g)

  13. Air Humidity Air humidity: 0.5 - 3 mg water / 10 mL air Tropical countries: Air conditioning Well sealed titration cell Conditioning of the titration stand Protect titration stand, titrant and solvent from ingress of water.

  14. Drift determination The titration stand is not 100 % tight against air humidity. Drift determination The drift is the amount of water entering into the titration stand per minute. 1 - 20 µg H20 / minute Automatic drift compensation in the result calculation.

  15. Resolution and Detection Limit Volumetric Karl Fischer Titration Resolution of burette: 10,000 steps Detection limit : 50 x Resolution Burette size: 5 mL Titrant: 5 mg H20/mL Resolution: 2.5 µg H20/step Detection limit: 125 µg H20 For 5 g sample: 25 ppm Titrant: 2 mg H20/mL Resolution: 1 µg H20/step Detection limit: 50 µg H20 For 5 g sample: 10 ppm

  16. Coulometric KF Titration - + • Iodine is generated electrochemically during titration • Water in trace amounts: 1 ppm - 5 %

  17. Coulometric KF Titration Cathode Anode – + • Titration cell and reagents Generator electrode Double platinum pin electrode Anolyte (sulfur dioxide, imidazole, iodide, different solvent for different application - methanol, ethanol with chloroform, octanol, ethyleneglycol ) Catholyte (similar or modified solution) Diaphragm

  18. Coulometric KF Titration + – AnodeIodine production by oxidation 2 I- I2 + 2 e- Cathode H2 2 H+ + 2 e- H+ H - Side reaction:Reduction of sulfur components. After 1 - 2 weeks, smells like mercaptans Change catholyte every week! - I I- • Same reaction as volumetric KF Titration • but Iodine is produced just in time from iodide

  19. Coulometry Theory Charles Augustin de Coulomb 14.6.1736 - 23.8.1806 One Coulomb C is the quantity of charge transported by an electric current of one Ampere (A) during one second (s). 1 C = 1 A • 1 sAbsolute method, no standardization! • To produce one mol of a chemical compound, using one electron, 96484 C are required. • 2 I- ions react to form I2 which in turn reacts with water • 1 mol of water (18g) is equivalent to 2 x 96484 C or 10.72 C/mg water.

  20. Filling the Titration Cell Cathode Anode – + The level of the anolyte should be 3 - 5 mm higher than the level of catholyte so that the flow is from the anolyte compartment to catholyte compartment. Catholyte  Low drift value With stirring the level difference of anolyte and catholyte will be stable. Anolyte Catholyte:Fill in 5 mL catholyte. Anolyte: Fill in ~ 100 mL anolyte

  21. Filling the Titration Cell Cathode Anode – + If the catholyte level is higher or at the same level as the anolyte, there is a flow of moisture into the anolyte compartment. Catholyte always contains water! Catholyte  High drift value Anolyte

  22. With or Without Diaphragm What are the differences?

  23. With Diaphragm – – + + I- I I I Iodine is only in the anode compartment and reacts with water. It is possible that iodine can go to the cathode and convert to iodide. I- I- - - - With or Without Diaphragm Without Diaphragm

  24. Without Diaphragm H+ H I I- - - Iodine I2 can go to the cathode and convert to iodide. – + Prevention: • Small cathode surface less chance to contact iodine • high stirrer speed iodine reacts faster with water • high iodine production speed hydrogen protects cathode Only a little less accuratefor samples with very low water content. • bigger sample error has no effect

  25. Without Diaphragm – + R-NH2 + H2O R-NO2 H+ H I I- - - The hydrogen produced at the cathode is a very good reducing agent. Easily reducible samples (nitrocompounds) get reduced, which produces water.  too high result Not recommended for easily reducible samples: e.g. nitrobenzene, unsaturated fatty acids, etc.

  26. Without Diaphragm • A little bit less accuracy for very small water content (< 50 µg/sample). • Not recommended for easily reducible samples: nitro compounds, unsaturated fatty acids, etc. • Titration cell easier to clean. • Long-term drift value more stable. • Only one reagent. Titration cell without diaphragm is the standard set-up for:Hydrocarbons, halogenated hydrocarbons, alcohols, esters, ethers, acetamides, mineral oils, edible oils, ethereal oils

  27. Resolution and Detection Limit + - Coulometric Karl Fischer Titration Resolution: 0.1 µg water Detection limit: 5 µg water for 5 g sample  1 ppm Measuring range: 10 µg - 100 mg water/sample 1 ppm - 5 % water

  28. Coulometry versus Volumetry volumetry 100 % srel < 0.5 % Not suitable for coulometry 10 % 1 % 1000 ppm srel 5 - 0.5 % 100 ppm srel > 5 % 10 ppm Not suitable for volumetry 1 ppm • Repeatability coulometry srel < 0.5 % srel 5 - 0.5 % srel > 5 %

  29. KF Indication Principle (1/2) Ipol = 20µA U = 650mV 2 • Bivoltametric indicationconstant current at the double platinum pin electrode ==> polarization current (Ipol) • During titration: • I2 reacts with water • no free I2 in the solution • high potential

  30. KF Indication Principle (2/2) Ipol = 20µA U = 84mV 2 e e + - I2 I2 2I- I2 + 2e- -> 2 I- 2 I- -> I2 + 2e- • At endpoint • all water has reacted with I2 • After the endpoint • free I2 in the solution • I2 is reduced to I- at the cathode • ionic conductivity occurs and the measured potential drops • potential change = endpoint

  31. KF Control: Titrator Algorithm KF Classical E/mV Control range EP V/mL KF Fuzzy logic E/mV EP V/mL • Karl Fischer Fuzzy Logic Control DL31/38 • No control band required (typical 300 mV) • The titrant addition rate depends on: • the distance to the endpoint EP • the potential change/increment • Advantages: • Simpler control: Only two control parameters Vmin , Vmax (smallest/largest increment) • Faster, more accurate, and better precision even at low water content(toluene: n = 5, 115 ppm, srel 0.17% )

  32. KF Control: Termination Parameters (1/3) • Delay time • the actual potential is lower than the EP for a defined time after the last titrant increment • typical delay : 15 - 20 sec • Note:Adapt the smallest increment to the drift and to the concentration of the titrant E (mV) EP 15 s t(s)

  33. KF Control: Termination Parameters (2/3) Drift (µg/min) EP abs. drift stop = 30 µg/min • Absolute drift stop • the actual drift is less then the predefined value • typical value : 30 g/min • Note:Adapt the value to the initial drift t(s)

  34. KF Control: Termination Parameters (3/3) Drift (µg/min) Rel. Drift stop = 20 µg/min Initial drift t(s) • Relative drift stop • the sum of the initial and the relative drift has been reached • typical value : 15 g/min • independent from the initial drift and of titrant concentration • ideal with side reactions that cannot be suppressed otherwise

  35. Karl Fischer Titration : Checks • Relevant points to be checked • System tightness : Check carefully • Ambient moisture : Drift determination • Stability of titrant : Standardisation • Side reactions : Check literature • Sample handling : Accuracy, precision • Free water only : Sample preparation

  36. Complete Solution : Solutions and Support • Application brochures • Internet databases www.titration.net