An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. HBF-500 Clinical evaluation test 1. Objective To evaluate the measurement accuracy of HBF-500, and to compare HBF-500 with the other companies products (Tanita BC-532 and SOEHNLE Body Balance NewYork). 2. Research institute Christian-Albrechts-Universität zu Kiel 3. Subjects (1) Body fat percentage (%FAT) : 54 females and 52 males (2) Skeletal Muscle percentage (%SM) and resting metabolism (RM): 51 females and 49 males (3) Visceral fat level (VFL) : 50 females and 50 males Table 1 Characteristics of the subjects

  2. HBF-500 Clinical evaluation test 4. Result of Body Fat percentage Figure 1 Correlation of measured value and DXA (Female) Figure 2 Correlation of measured value and DXA (Male)

  3. HBF-500 Clinical evaluation test 5. Result of skeletal muscle percentage Figure 3 Correlation of measured value and MRI (Female and Male) 6. Result of visceral fat level Notes : MRI Visceral fat level =MRI Visceral fat area(cm2) / 10 Figure 4 Correlation of measured value and MRI (Female and Male)

  4. HBF-500 Clinical evaluation test 7. Result of resting metabolism Figure 5 Correlation of measured value and Exhalation analysis (Female and Male)

  5. Appendix 1 Tanita BC-532 Visceral fat level and Muscle mass A1-1. Result of Visceral fat level Comment : It is presumed that the error of TANITA is larger than that of HBF-500. Notes : MRI Visceral fat level =MRI Visceral fat area(cm2) / 10 “S.E.E” is not be described, because the calculation formula and the reference value of TANITA are uncertain, Figure A1-1 Correlation of measured value and MRI (Female and Male) A1-2. Result of Muscle mass Comment : The result of TANITA is different from the skeletal muscle mass, because TANITA uses DXA for the reference value. TANITA’s muscle mass = weight – fat mass – bone mass Notes : “S.E.E” is not be described, because the calculation formula and the reference value of TANITA are uncertain, Figure A1-2 Correlation of measured value and MRI (Female and Male)

  6. Appendix 2 SOEHNLE Body Balance Muscle mass A2. Result of Muscle mass Comment : It is presumed that the error of SOEHNLE is larger than that of HBF-500. Notes : “S.E.E” is not be described, because the calculation formula and the reference value of SOEHNLE are uncertain, Figure A2 Correlation of measured value and MRI (Female and Male)

  7. Appendix 3 HBF-400 Clinical evaluation test Result of Body Fat percentage Figure A3-1 Correlation of measured value and DXA (Female) Figure A3-2 Correlation of measured value and DXA (Male)