Quantitative Imaging . Using imaging to analyze molecular events in living cells. Ann Cowan. FUNCTION OF MICROSCOPY. Function of any microscopy is NOT simply to magnify! Function of the microscope is to RESOLVE fine detail. Magnification makes objects bigger. Magnification.
E N D
Quantitative Imaging Using imaging to analyze molecular events in living cells Ann Cowan
FUNCTION OF MICROSCOPY • Function of any microscopy is NOT simply to magnify! • Function of the microscope is to RESOLVE fine detail.
Magnification makes objects bigger Magnification
Magnification in the microscope is not perfect; the magnified image is blurred by diffraction Magnification
RESOLUTION means objects can be seen as separate objects Resolution
RESOLUTION l d N.A. The resolution of a microscope is the shortest distance two points can be separated and still be observed as 2 points. Not resolved just resolved Well resolved MORE IMPORTANT THAN MAGNIFICATION !!
How to get better resolution? Image plane Objective lens specimen
specimen How to get better resolution? Image plane Objective lens
specimen How to get better resolution? Image plane Objective lens
WHAT DETERMES RESOLUTION? • Contrast is necessary to detect detail (edges) from background • Diffraction fundamentally limits resolution diffraction occurs at the objective lens aperture
IMAGE OF A SELF-LUMINOUS POINT IN THE MICROSCOPE maximum First minimum Light from each point of the object is spread out in the microscope because light diffracts at the edges of the lens = Airy Disk Objective lens
RAYLEIGH CRITERIONGenerally accepted criterion of resolution Single point sourcce Just resolved Just resolved Wel resolved Intensity Central maximum of one peak overlies 1st minimum of neighboring peak
What determines the distance between Peaks? Objective θ θ specimen The maximum angle of light collected by the objective lens. Larger angle of collection = Better resolution
Maximum angle of light collected from a point determines width of Airy Disk q specimen Objective lens Image plane Min distance between points: wavelength refractive index λ d sinq n Numerical Aperture (N.A.) = n sinq
Resolution therefore is given by: l d N.A. • To reduce d, and therefore achieve better resolution: • wavelength • N.A. • Light microscope: • maximum N.A. is 1.4, • for visible (e.g. green light), = 500 nm • thus best resolution is 0.2 um. Useful magnification is limited to 500-1000 X N.A., so about 1,000 X
Contrast is required to see objects Increasing Contrast light from an object must either be different in intensity or color (= wavelength) from the background light
AIRY DISK 255 INTENSITY 0 Z-POSITION
AIRY DISK 255 INTENSITY 0 Z-POSITION
AIRY DISK 255 INTENSITY 0 Z-POSITION
FWHM Z-POSITION INTENSITY Z resolution Z Resolution defined as FWHM = the full width at half maximal intensity of a z line of a point source For 1.4 N.A. lens, Z resolution ~ .5 um By Nyquist theorem, need to collect at 0.25 um Z steps
NA NA4 mag2 OBJECTIVE LENS • Resolution • Intensity • > corrections Intensity (For epiflourescence; for transmission it is NA2 of objective time NA2 of condenser)
Digital Images Are Arrays of Numbers Value at each point is the amount of light collect from each point in an image 2-D Image becomes array of intensity values (grey levels) from 0 -255 (for 8 bit image) or 0-4,126 for 12 bit image. Each point in the array is a pixel
How CCD cameras Make an image Figure 1. The pixels of a CCD collect light and convert it into packets of electrical charge Figure 2. The charges are quickly moved across the chip. Figure 3. The charges are then swept off the CCD and converted to analog electrical impulses, which are then measured as digital numerical values.
RGB (color ) IMAGE Display Red channel Green channel Blue channel
VOXELS ARE 3D PIXELS 2-D Image becomes array of intensity values (grey levels) from 0 -255 (for 8 bit image) or 0-4,126 for 12 bit image. Each point in the array is a pixel For successive Z section, 2D arrays are stacked into 3D arrays of values, each element is called a “voxel”
DIGITAL IMAGE MANIPUTATIONS (manipulating arrays of numbers in meaningful ways) • Frame averaging • (time averaging on CCD) 2 + =
DIGITAL IMAGE MANIPUTATIONS Output value Input value LUT (manipulating arrays of numbers in meaningful ways) • look up table (LUT) manipulations e.g. contrast stretching
DIGITAL IMAGE MANIPUTATIONS (manipulating arrays of numbers in meaningful ways) • image math e.g. ratio imaging =
Original image enhanced image background image enhanced - background image frame averaged enhanced - background
FLOURESCENCE Excited Energy States E Ground State lifetime t
EPIFLUORESCENCE First barrier filter Second barrier filter dichroic mirror objective lens specimen
Flourescence detection is linear and can be used to quantify relative or absolute amounts of molecules • If conditions are identical, 2X fluorescence = 2X amt of fluorophore • Because light in the microscope is spread out by diffraction, conditions within and between images are not always identical. • As with any measurement, need to be careful with measurements • Must be within linear range of detector (no 0’s, not above maximum level) • Must subtract background (generally cell-free area) • ALL conditions in microscope must be identical
Fluorescent Ion Indicators Fluorescence properties change when specific ion is bound. For example: fura-2 in low Ca2+ excitation maximum at 360nm fura-2 in high Ca2+ excitation maximum at 340nm ratio of fluorescence intensity at the two wavelengths is a measure of the concentration of Ca2+.
Image Math Bkgd corrected image 340ex Cell with 340ex Bkgd with 340ex _ = Cell with 360ex Bkgd with 360ex Bkgd corrected image 360ex _ =
Image Math Bkgd corrected image 340ex Ratio image (340/360) Bkgd corrected image 360ex
Dual Wavelength Ratios are Independent of the Amount of Fluorescent Indicator Ratioing helps eliminate bleaching and dye leakage artifacts and thus are sensitive only to the concentration of analyte
Dual Wavelength Ratios Normalize for Variable Thickness within a Sample (e.g. a cell under a microscope)
TIRF excites fluorescence only within a narrow region next to the substrate