1 / 29

Processing EMG

Processing EMG. David DeLion UNLV Biomechanics Lab. Why do we process EMG?. Raw EMG offers us valuable information in a practically useless form Raw EMG signals cannot be quantitatively compared between subjects

Thomas
Download Presentation

Processing EMG

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Processing EMG David DeLion UNLV Biomechanics Lab

  2. Why do we process EMG? • Raw EMG offers us valuable information in a practically useless form • Raw EMG signals cannot be quantitatively compared between subjects • If electrodes are moved raw EMG signals cannot be quantitatively compared for the same subject

  3. Raw Half-wave rectified Full-wave rectified Filtering Averaging Smoothing Integration Root-mean Square Frequency spectrum Fatigue analysis Number of Zero-crossings Amplitude Probability Distribution Function Wavelet Types of Signal Processing

  4. Removing Bias • Low amplitude voltage offset present in hardware • Can be AC or DC • Calculate the mean of all the data • Subtract mean from each data point

  5. Raw EMG • Unprocessed signal -Amplitude of 0-6 mV -Frequency of 10-500 Hz • Peak-to-Peak -Measured in mV -Represents the amount of muscle energy measured • Onset times can be determined • Analysis is mostly qualitative

  6. Rectification • Only positive values are analyzed -Mean would be zero • Half-wave rectification - all negative data is discarded, positive data is kept. • Full-wave rectification- the absolute value of each data point is used • Full-wave is preferred

  7. Filtering • Notch filter -Band reject filter; usually very narrow -For EMG normally set from 59-61 Hz -Used to remove 60 Hz electrical noise -Also removes real data! -Too much noise will overwhelm the filter

  8. Filtering • Band Pass filter -allows specified frequencies to pass -low end cutoff removes electrical noise associated with wire sway and biological artifacts -high end cutoff eliminates tissue noise at the electrode site -often set between 20-300 Hz

  9. Filtering • There are no perfect filters! • Face muscles can emit frequencies up to 500 Hz • Heart rate artifact can be eliminated with low end cutoffs of 100 Hz • Filters which include 60 Hz include the noise from equipment

  10. Averaging • Average EMG can be used to quantify muscle activity over time • Measured in mV • Values are averaged over a specified time window • Window can be moved or static • Moving windows are a digital smoothing technique • For moving windows the smaller the time window the less smooth the data will be

  11. Averaging • For EMG window is typically between 100-200 ms • Window is moved over the length of the sample • Moving averages introduce a phase shift • Moving averages create biased values -values are calculated from data which are common to the data used to calculate the previous value • Very commonly used technique

  12. Integration • Calculation of area under the rectified signal • Measured in Vs • Values are summed over the specified time then divided by the total number of values • Values will increase continuously over time • The integrated average will represent 0.637 of one-half of the peak to peak value • Quantifies muscle activity • Can be reset over a specified time or voltage

  13. Root Mean Square • Recommended quantification method by Basmajian and DeLuca • Calculated by squaring each data point, summing the squares, dividing the sum by the number of observations, and taking the square root • Represents 0.707 of one half of the peak-to-peak value

  14. Number of Zero Crossings • Counting the number of times the amplitude of the signal crosses the zero line • Based on the idea that a more active muscle will generate more action potentials, which will cause more zero crossings in the signal • Primarily used before the FFT algorithm was widely available

  15. Frequency Analysis • Fast Fourrie Transformation is used to break the EMG signal into its frequency components. • Frequency components are graphed as function of the probability of their occurrence • Useful in determining cutoff frequencies and muscle fatigue

  16. Fatigue Analysis • Isometric contraction • The two most important parameters for fatigue analysis are the median and mean frequency. • Median frequency decreases with the onset of fatigue • If fatigue is being measured it is important to have a large band pass filter

  17. Amplitude Probability Distribution Function • Illustrate variance in the signal • X-axis shows range of amplitudes • Y-axis shows the percentage of time spent at any given amplitude • Distribution during work should be bimodal -peak associated with effort -peak associated with rest

  18. Wavelet analysis • Used for the processing of signals that are non-stationary and time varying • Wavelets are parts of functions or any function consists of an infinite number of wavelets • The goal is to express the signal as a linear combination of a set of functions • Obtained by running a wavelet of a given frequency through the original signal

  19. Wavelet Analysis • This process creates wavelet coefficients • When an adequate number of coefficients have been calculated the signal can be accurately reconstructed • The signal is reconstructed as a linear combination of the basis functions which are weighted by the wavelet coefficients

  20. Wavelet Analysis • Time-frequency localization • Most of the energy of the wavelet is restricted to a finite time interval • Fourier transform is band limited • Produces good frequency localization at low frequencies, and good time localization at high frequencies • Segments, or tiles the time-frequency plane

  21. Wavelet Analysis • Wavelets remove noise from the signal • Signal energy becomes concentrated into fewer coefficients while noise energy does not

  22. Normalizing • There is no absolute scale so direct comparisons between subjects or conditions cannot be made • Maximum voluntary contraction levels are often used to compare EMG readings between subjects (i.e..50% MVC) • Relies on subject to give max effort

  23. Normalizing • Record contractions over a dynamic movement cycle • At least 4 repetitions are required • Peak values are averaged which creates an anchor point • Subsequent values are represented as a percentage of the anchor point

  24. Conclusion • EMG offers a great deal of useful information • The information is only useful if it can be quantified • Quantifying EMG data can be a qualitative process

  25. Thank You Any Questions?

  26. Bibliography • Kleissen, R.F.M, Buurke, J.H., Harlaar, J., Zilvold, G. (1998) Electromyography in the Biomechanical analysis of human movement and its clinical application. Gait and Posture. Vol. 8,143-158 • Aminoff, M.J. (1978) Electromyography in Clinical Practice. Addison-Wesley Publishing Company, Menlo Park, CA • Dainty, D.A., Norman, R.W. (1987) Standardized Biomechanical Testing in Sport. Human Kinetics Publishers, Champaign, IL • Cram, J.R., Kasman, G.S. (1998) Introduction to Surface Electromyography. Aspen Publishers, Gaithersburg, MD • Medved, V. (2001) Measurement of Human Locomotion. CRC Press, New York, NY

  27. Bibliography • Loeb, G.E., Gans, C. (1986) Electromyography for Experimentalists. The University of Chicago Press, Chicago, IL • Basmajian, J.V., DeLuca, C.J. (1985) Muscles Alive. Williams & Wilkins, Baltimore, MD • Moshou, D., Hostens, I., Ramon, H. (2000) Wavelets and Self-Organizing Maps in Electromyogram Analysis. Katholieke Universiteit Leuven • DeLuca, C.J. (1993) The Use of Surface Electromyography in Biomechanics. NeuroMuscular Research Center, Boston University • DeLuca, C.J. (2002) Surface Electromyography: Detection and Recording. Delsys Incorporated.

More Related