University of Pittsburgh Senior Design – BioE 1160

University of Pittsburgh Senior Design – BioE 1160.

pabla
Download Presentation

University of Pittsburgh Senior Design – BioE 1160

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. University of Pittsburgh Senior Design – BioE 1160 Incorporating a Pressure Sensor-Transmitter(RF) Complex on Transseptal Cannula for use with TandemHeart™ PTVA SystemTeam Members: Mihoko Hashimoto Lauren Kokai Nitin Narayana Katie O’CallaghanProject Advisors: David H.J. Wang, Ph.D. Douglas E. Smith, Ph.D. Dennis Kopilec Marlin H. Mickle, Ph.D.December 5, 2003

  2. Overview TandemHeartTM - A centrifugal pump that allows for fast deployment of oxygenated blood from the Left Atrium (LA) to the femoral artery and thus bypassing the Left Ventricle (LV). • Advantages: Reduced workload on LV, possible healing of LV, helps maintain blood circulation and organ perfusion during heart surgery, bridge to transplant. • Audience: Cardiologists and Cardiac Surgeons (maintaining blood flow), patients suffering from myocardial infarction or acute carditis, patients with deteriorating LV functionality waiting for a transplant, patients recuperating from heart surgery (increase speed of LV healing).

  3. Overview Problem: Unmonitored changes in LA pressure (below a certain pressure threshold) may pose a threat to the patient due to a potential “suck down” of tissue (overdrainage of LA). Cause tissue damage and reduce blood flow. Solution: Real time pressure readings in the LA can possibly prevent the tissue “suck down”. • Incorporate a pressure sensor into the tip of the Transseptal Cannula along with the use of RF-technology for signal transmission. • Effectively determine an appropriate pump speed depending on patient’s conditions without further invasive processes.

  4. Project Goals The ultimate goal of this project is to enable real-time measurements of Left Atrial Pressure (LAP) with TandemHeart use, by placing a barometric absolute pressure (BAP) sensor chip (Kavlico sCAP1) at the tip of the cannula. + Incorporation of RF transmitter complex to transmit the real-time measurements without need for wiring. + Determine durability of the device after coating using simulated conditions. High Level Timing Goals • Analysis of pressure sensor properties & manufacturing limitations • Development of sensor-signal transmitter (RF?) complex • Establish that device meets PDS under simulated clinical use conditions • Verify accurate signal transmission in intended clinical environment

  5. Future Project Opportunities Successful completion of this project could be used for future development of a safety pump speed adjust feature in the pump controller. • By developing an algorithm to respond to low LAP conditions, the controller on the TandemHeartTM will be able to auto-adjust (a vital safety feature) by decreasing pump speeds. The sensor-RF transmitter complex could also be applied for countless other internal biomedical devices!!

  6. Features & Benefits Pressure sensor  provide potential for accurate real-time monitoring of LAP to prevent “suck down” of LA  allow for more accurate/sensitive pump speed for individual patient’s condition (less need for highly trained system operator) RF transmitter  no need for wiring

  7. Group Object Tree

  8. Design Sketch Receiver Pressure Sensor 1.2mm RF Transmitter 1.5mm

  9. Design Alternatives • Pressure sensor: other manufacturers • Various sensor placements • Single sensor • Inside lumen / exterior • Various RF transmitter placements • Further away from cannula tip • Integrated on microchip • Potential need for extra transmitter outside body • Potential need to include amplification of signal in order to transmit through the chest wall • Pressure signal output via wire (instead of RF transmitter)

  10. Group Function-Means Tree

  11. Competitive Analysis Integrated pressure sensing capabilities don’t yet exist! Existing Designs (circulatory support): • Centrifugal flow pumps: • Thoratec, Arrow, MicroMedical, MedQuest, Kriton Medical, Terumo… • Other forms of circulatory assistance: • IABP (Intra-Aortic Balloon Pump)

  12. Strengths of our Design: Incorporated LAP-sensor on cannula tip no need for separate LAP line safety feature to prevent LA “suck down” real-time monitoring of patient LAP RF pressure signal transmission no wires for signal output future integration of auto safety adjust in controller Weaknesses of our Design: More complicated (more potential failure modes) Higher production cost Competitive Analysis

  13. Technology • Using new technology: • MEMS pressure sensor chip • RF data transmission (of pressure signal) • Engineering Models: • SolidWorks: design modifications • Simulated pressure/flow loop systems

  14. Team / Resources Advisors: • Drs. David Wang & Doug Smith, Dennis Kopilec (CardiacAssist, Inc.)  Medical device design, cannula design, clinical knowledge • Dr. Marlin Mickle (Dept. of Electrical Engineering, Univ. of Pittsburgh)  RF technology Locations: • Lab space for testing of design: CardiacAssist, Univ. of Pittsburgh • Swanson Center: design and RF systems

  15. Team / Resources Equipment: • Pump system supplies and testing equipment: CardiacAssist, Inc. • Pressure sensor: Kavlico, Inc. • RF transmitter: not yet identified Manufacturing: to be determined by CardiacAssist, Inc.

  16. Projected Task Schedule

  17. Personal Responsibilities Project Criteria: Milestone 1 (30%) Milestone 2 (30%) Milestone 3 (10%) Milestone 4 (10%) Milestone 5 (5%) Milestone 6 (7.5%) Milestone 7 (7.5%)

  18. Current Status Tasks completed as of 12/5/2003 • Group meeting with CardiacAssist -Determined scope of project and documented specific deliverables CAI requires from students at conclusion of project *Determined Product Design Specifications -Communicated project schedule and goals through Gantt chart, Function Means Tree, Object Tree • Group meeting with U of P advisor Marlin H. Mickle, Ph.D -Students presented summary of project goals and relevant design information, discussed student needs for assistance and obtained background info on RF technology and circuit integration components, as well as anticipated difficulties • Order placement for Kavlico sCAP 1 BAP sensor chips (5)

  19. Acknowledgements CardiacAssist • Drs. Doug Smith & David Wang, Dennis Kopilec • Project funding • Provision of pump system supplies & testing equipment • Lab space • Great project opportunity!! Dr. Mickle • RF technology expertise & project guidance re: electrical aspects • Lab space MG • Project development guidance • FDA background info (ad nauseum) • Moral support!

  20. Questions….?