An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Filter Wheel Closed Calibration Observations Kip Kuntz Johns Hopkins University and NASA/Goddard Space Flight Center EPIC Operations and Calibration Meeting - BGWG Palermo – 12-13 March 2007

  2. Filter Wheel Closed Observations Regular calibration observations are needed for sensitive tracking of variations in the quiescent particle background. As the CCDs age this is likely to become more important. Anomalous states are perhaps becoming more prevalent. FWC Data: Currently there are 700-1000 ks of data for the CCDs in their standard states. However, very little data from the last year. For most chips which have anomalous states they are detectable in corner data, therefore they can be tracked in normal observations (on or off).

  3. Anonymous States However, some anomalous states are not detectable in corner data (known as an anonymous anomalous state). This has been identified in MOS1-4 and may occur in others Therefore there is a need for regular short FWC exposures to detect anonymous states. They occur primarily for E<1.0 keV, and the MOS1-4 anonymous state strongly effects ~1/9th of the chip. Typical FWC rates in the 0.3-1.0 keV band are 1-2 counts ks-1 in one ninth of a CCD. Anomalous states have count rates of ~10 counts ks-1for the same area

  4. MOS1-4 Anonymous State Images from MOS1-4 in 0.3-1.0 keV an the anonymous anomalous state. Values are the counts ks-1 for one ninth of the CCD Monthly FWC exposures of 10ks would produce 20 counts for each CCD ninth, enough for some tracking of the state and sufficient to produce 2160 counts/chip/year (enough for spectral analysis) in the standard state