If agglutination is observed, STOP the test and perform cell washing procedures before proceeding. - PowerPoint PPT Presentation


  • 66 Views
  • Uploaded on

RapidVet ® -H Blood Group Determination System for Identifying DEA 1.1 Positive and DEA 1.1 Negative Dogs.

loader
I am the owner, or an agent authorized to act on behalf of the owner, of the copyrighted work described.
capcha
Download Presentation

PowerPoint Slideshow about 'If agglutination is observed, STOP the test and perform cell washing procedures before proceeding.' - nizana


An Image/Link below is provided (as is) to download presentation

Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server.


- - - - - - - - - - - - - - - - - - - - - - - - - - E N D - - - - - - - - - - - - - - - - - - - - - - - - - -
Presentation Transcript
If agglutination is observed stop the test and perform cell washing procedures before proceeding

RapidVet®-HBlood Group Determination Systemfor Identifying DEA 1.1 Positive and DEA 1.1 Negative Dogs


If agglutination is observed stop the test and perform cell washing procedures before proceeding

Kit contains:auto-agglutination saline screen cards, patient test cards, pipettes, stirrers, diluent, instructions, interpretation guideand report card.Shelf-life: 24 months at room temperature


If agglutination is observed stop the test and perform cell washing procedures before proceeding

Prior to typing, check for auto-agglutination using the card provided in the kit. This card is used for three patients and, after use, should be returned to the kit box.

If agglutination is observed, STOP the test and perform cell washing procedures before proceeding.

If auto-agglutination does not occur, proceed to the next step.


If agglutination is observed stop the test and perform cell washing procedures before proceeding
Remove the Test Card from its plastic sleeve. Write the name/number of the dog and the testing date on the card.


Dispense 1 drop of diluent from the dropping bottle into each well
Dispense 1 drop of diluent from the dropping bottle into name/number of the dog and the testing date on the card.each well.


If agglutination is observed stop the test and perform cell washing procedures before proceeding

Aspirate a small amount of patient sample into a pipette and release 1 drop (50µl) into each well, taking care not to cross-contaminate the wells.


If agglutination is observed stop the test and perform cell washing procedures before proceeding

Using a release 1 drop (50NEW stirrer and pressing downward with the flat portion of the stirrer, spread and mix the materials within the entirety of the first well for about 10 seconds.


If agglutination is observed stop the test and perform cell washing procedures before proceeding
Using a release 1 drop (50NEW stirrer, spread and mix the materials within the entirety of the second well for about 10 seconds.


If agglutination is observed stop the test and perform cell washing procedures before proceeding
Using a release 1 drop (50NEW stirrer, spread and mix the materials within the entirety of the third well for about 10 seconds.


Rock the card using a transaxial motion for 1 minute
Rock the card, using a transaxial motion, for release 1 drop (501 minute.

Be sure materials are mixing and “rotating” within each individual well.

Be careful not to cross-contaminate.


Read the results noting whether agglutination has or has not occurred in the patient test well
Read the results, noting whether agglutination has or has not occurred in the “Patient Test” well.

DEA 1.1

DEA 1.1+


If agglutination is observed stop the test and perform cell washing procedures before proceeding

After the card has been read, take a digital photograph for a permanent record, or allow to dry and replace in its plastic sleeve as a permanent record.