An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. THEMIS – FGM – Status Report • Deliverables, Status • FGM Capabilities vs. scientific Requirements • Remaining tasks, Commissioning, inflight calibration Uli Auster and Karl-Heinz Glassmeier, IGEP, TUBS Werner Magnes, IWF, OEAW

  2. Deliverables / Status • Electronics (FGE) • Designed at IGEP, MagSon and IWF,manufactured at UCB • Calibrated in Germany /Austria • Delivered to UCB February 2005 • Sensor (FGS) • Designed and manufactured in Germany • Qualified and calibrated in Germany /Austria • Delivered to UCB February 2005 • S/C Integration • Facilities for functional & magnetic tests made available to UCB • Tests after each integration step done by UCB, data evaluation by IGEP at TUBS

  3. FGM capabilities / abs. Stability • Requirement: The absolute stability of the FGM shall be less than 1nT Dependency on sensor temperature: 2nT/170°C

  4. The relative stability of the FGM shall be less than 0.2nT/12h FGM capabilities / rel. Stability Test shows 12 hours registration. Sensor in the ferromagnetic shield of the THEMIS Calibration Unit (TCU) 0.2nT

  5. 0.1 0.09 0.08 0.03nT/Hz1/2 0.07 0.06 [nT/Sqrt(Hz)] 0.01nT/Hz1/2 0.05 0.04 0.03 0.02 0.01 0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1 1.1 1.2 1.3 1.4 1.5 1.6 1.7 1.8 1.9 2 [Hz] noise FGM capabilities / Noise • The FGM noise level @ 1Hz shall be less than 0.03nT/Hz1/2 Out of 6 sensors 5 sensors and all their components:noise ~0.01nT/Hz1/2 One sensor and one component:noise~0.015nT/Hz1/2 -selected as spare-

  6. The FGM digital resolution shall be less than 0.1nT The FGM science range shall exceed 0-1000nT FGM capabilities / Resolution Test shows Earth field variationsrecorded by two Themis magneto-meters Back ground field: 12000nTResolution: better 0.1nT 1 nT Offset: 1 nT

  7. Remaining tasks • S/C integration • Investigation of Low Telemetry timing problems • Further tests after each integration step • Determination of sensor alignment • Commissioning • FGM has to be switched on first !!!!!! • Measurement during boom deployment !!!!!! • Using small s/c distances after release for initial tests • Inflight calibration • Determination of 8 parameters by minimising spin signal • Determination of 3 parameters by IGRF fit • Determination of spin axis offset by field assumptions