basic patient monitoring for anesthesia bis monitors n.
Download
Skip this Video
Loading SlideShow in 5 Seconds..
Basic Patient Monitoring For Anesthesia BIS Monitors PowerPoint Presentation
Download Presentation
Basic Patient Monitoring For Anesthesia BIS Monitors

Loading in 2 Seconds...

play fullscreen
1 / 5

Basic Patient Monitoring For Anesthesia BIS Monitors - PowerPoint PPT Presentation


  • 309 Views
  • Uploaded on

Basic Patient Monitoring For Anesthesia BIS Monitors. Dr. Jeffrey Elliot Field HBSc , D.D.S. , Diplomat of the National dental Board of Anesthesia, Fellow of The American Dental Society of Anesthesia. Bispectral Index (BIS Monitors). Proper Electrode Placement.

loader
I am the owner, or an agent authorized to act on behalf of the owner, of the copyrighted work described.
capcha
Download Presentation

PowerPoint Slideshow about 'Basic Patient Monitoring For Anesthesia BIS Monitors' - kirby


An Image/Link below is provided (as is) to download presentation

Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server.


- - - - - - - - - - - - - - - - - - - - - - - - - - E N D - - - - - - - - - - - - - - - - - - - - - - - - - -
Presentation Transcript
basic patient monitoring for anesthesia bis monitors

Basic Patient Monitoring For AnesthesiaBIS Monitors

Dr. Jeffrey Elliot Field

HBSc, D.D.S. , Diplomat of the National dental Board of Anesthesia,

Fellow of The American Dental Society of Anesthesia

proper electrode placement
Proper Electrode Placement
  • Wipe the forehead with alcohol and dry with gauze.
  • Either side of the head can be used except in the case of stroke were you would use the unaffected side.
  • Electrode # 1 is placed in the center of the forehead 1.5 inches above the bridge of the nose
  • Next electrode # 4 is placed directly above the eyebrow
  • Place electrode #3 on the temple between the outer canthus of the eye and the hairline
  • Now electrode #2 will be in the proper position to be secured.
values for bis monitoring
Values for BIS Monitoring
  • General anesthesia associated with no awareness under anesthesia occurs from BIS values of 40-60.
  • Deep Sedation occurs with BIS Values of 60-80
  • In special needs patients work around 80-90 for sedation.