ANAESTHETIC CONSIDERATIONS IN CPB BY- DR SUCHIT KHANDUJA

ANAESTHETIC CONSIDERATIONS IN CPB BY- DR SUCHIT KHANDUJA MODERATOR-PROF SURINDER SINGH . DEFINITION .

karis
Download Presentation

ANAESTHETIC CONSIDERATIONS IN CPB BY- DR SUCHIT KHANDUJA

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. ANAESTHETIC CONSIDERATIONS IN CPB BY-DR SUCHIT KHANDUJA MODERATOR-PROF SURINDER SINGH

  2. DEFINITION • “CPB is the technique whereby blood is totallyor partiallydiverted from the heart into a machine with the gas exchange capacity and subsequently returned to the arterial circulation at appropriate pressures & flow rates.”

  3. HISTORICAL ASPECTS • Legllois (1812) : “circulation might be taken over for short periods” • Dr.John Gibbon(Philadelphia) 1953 : “performed ASD repair with the aid of CPB for the 1st time with the survival of patient.”

  4. GOALS OF CPB To provide a still & Bloodless Heart with blood flow temporarily diverted to an Extracorporeal Circuit that functionally replaces the Heart & the Lung

  5. GOALS OF CPB RESPIRATION Ventillation Oxygenation CIRCULATION TEMP. REGULATION (Hypothermia) Low blood flowso ed blood trauma ses Body Metabolism.

  6. COMPONENTS OF CPB • TOTAL CPB : Systemic venous drainage CPB Circuit  External oxygenator heat exchanger External pump arterial filterSystemic circulation. • PARTIAL CPB : Portion of systemic venous return (Rt. Heart)  CPB .Undiverted blood  Rt. Atrium  Rt. Ventricle  Pul. Circulation  Lt. Atrium & Lt. Ventricle  Systemic Circulation.

  7. INTEGRAL COMPONENTS OF extracorporeal circuit • PUMPS • OXYGENATOR • Heat exchanger • Arterial filter • Cardioplegia delivery system • Aortic/atrial/vena cavalcannulae • Suction/vent

  8. PATIENT ARTERIAL LINE FILTER RESERVOIR ROLLER PUMP OXYGENATOR HEAT EXCHANGER

  9. ROLLER PUMP • Most commonly used. • Uses Volume displacement to create forward blood flow. • Non Pulsatile Blood Flow • By compressing Plastic Tubing b/w Roller & Backing Plate

  10. Properly set occlusion causes minimal haemolysis • Occlusion is 100% in cardioplegia &vent pumps • Each pump indepedently controlled by a rheostat • Larger tubing and lesser rotations cause minimal haemolysis

  11. Resistance= resistance of tubing+oxygenator+heatexcyhanger+filter+aorticcannulae+SVR • Usually line pr. depends on SVR and pump flow rate • Nl limit is 150-350 mm hg( >250 is seldom accepted)

  12. DISADVANTAGE of producing PULSATILE FLOW • Bubble Formation • Damage to Blood Components. ADVANTAGE : • Improved Tissue Perfusion • Better Preservation of Organ Function (Brain , Kidney)

  13. CENTRIFUGAL PUMP • Series of CONES that spin & propel blood forward by Centrifugal Force. • Safe • Reliable • Disposable • Simple to operate.

  14. CENTRIFUGAL PUMP • ADVANTAGE • No back pressure when tubing is temporarily obstructed / kinked • Doesn’t produce spatulated emboli from compression of the tubing • Cannot pump large amt.of gas / gas emboli. • Less blood trauma • High vol. output with moderate pressures • DISADVANTAGE • Inability to generate pulsatile flow • Potential discrepancy b/w pump speed & actual flow generated.

  15. Preferred over roller pumps in • Long-term CPB • In high-risk angioplasty patients • Ventricular assistance • Neonatal ECMO

  16. Pressure-regulated pump • Operates under passive filling • After&pre-load sensitive • Pump-chamberofpolyurethane+peristalticpump • Not yet fully evaluated

  17. OXYGENATOR • Where O2 & CO2 Exchange takes place. • Two Types : BUBBLE OXYGENATOR MEMBRANOUS OXYGENATOR

  18. BUBBLE OXYGENATOR • Gas exchange by directly infusing the gas into a column of systemic venous blood. • A) OXYGENATING CHAMBERS:bubbles produced by ventilating gas through diffusion plate into venous blood column • CO2 bubble & oxygen  plasma • Larger the No. of Bubbles ; Greater the efficiency of the oxygenator. • Larger bubbles improve removal of CO2 , diffuses 25 times more rapidly in plasma than O2 • Smaller bubbles are very efficient at oxygenation but poor in co2 removal

  19. DEFOAMING CHAMBER • Defoaming of frothy blood. • Large surface area coated with silicone • This es the Surface Tension of the bubbles causing them to burst.

  20. BUBBLE OXYGENATOR ADVANTAGE • Easy to assemble • Relatively small priming Volumes • Adequate oxygenating capacity • Lower cost. DISADVANTAGE • Micro emboli • Blood cell trauma • Destruction of plasma protein due to gas interface. • Excessive removal of CO2 • Defoaming capacity may get exhausted with time.

  21. MEMBRANOUS OXYGENATOR • Gas exchange across a thin membrane • Eliminates the need for a bubble-blood contact & need for a defoamer; so more physiological. • Blood damage is minimum • Ideal for perfusions lasting for >2-3 hours. • 2 types of membrane: • SOLID:Silicone • MICROPOROUS:polypropylene,Teflon &polyacrylamide

  22. MEMBRANOUS OXYGENATOR ADVANTAGE • Can deliver Air-O2 mixtures. • Hemolysis •  Protein desaturation •  Post-op bleeding • Better platelet preservation. DISADVANTAGE • Expensive • Large priming volume • Prolonged use pores may get blocked.

  23. CIRCUITS • Drains Venous Blood by gravity into oxygenator & returns the oxygenated blood under pressure to the systemic circulation.

  24. VENOUS DRAINAGE • Systemic venous blood (Rt.Heart)Oxygenator by • Direct Cannulation of SVC & IVC (BicavalCannulation) thru RA & joined to create a single drainage channel. • Single cannula into RA thru RA appendage. • Blood flow Oxygenator (Gravity) • Height Difference B/w Venacavae & Oxygenator > 20-30 cm.

  25. Complications • arrhythmia • bleeding • ivc/svc tear • cannulamalposition • low return • inadequate height • malposition • kink,clamp,air lock

  26. Size of cannula

  27. TUBINGS IN THE CIRCUIT • Non thrombogenic , Chemically Inert to prevent • clotting • Trauma to blood elements • Protein Denaturation • Smooth Internal Finish • Non Reactable Internal Surface • Durable to withstand high pressure & use of Roller pump

  28. Made of • PVC • Polyurethane • Silicone • I.D . Ranges from 3/16- 5/8 inches • HEPARIN BONDED CIRCUITS ARE AVAILABLE

  29. Disadvantages of plain circuits • Activation of platelets/coagulation factors • Post-op consumptive coagulopathyimmune reactions • More spallation Heparin coated circuits are • More hemo compatible • Cause less activation of platelets/white cells • Reduce heparin demand

  30. INTRACARDIAC SUCTION • Blood will enter the heart • Coronary venous Return • Retrograde flow in AR. • Bronchial Arteries • CARDIOTOMY SUCTION • Spilled Heparinised Blood is Scavenged & returned back to patient. • Handheld Suckers are used to return this blood.

  31. VENTRICULAR VENTING • LV Venting done to • Keep the operative field clear • Maintain Low LA & Pul.Venous Pressure • Remove air from Cardiac Chamber. •Blood from LV  Reservoir Bag

  32. RESERVOIR BAG • Collects the blood from VENOUS DRAINAGE & CARDIOTOMY SUCTION DRAIN PASSIVELY Reservoir Bag  Oxygenator Heat exchanger  Arterial Filter  Patient. • Volume in the bag should not be allowed to empty to prevent massive emboli.

  33. ARTERIAL RETURN • Ascending Aorta just proximal to Innominate Artery. • Femoral Artery in • Dissecting Aortic Aneurysm • For Reoperation • Emergency • Problems of Femoral Cannulation : • Sepsis • Formation of False Aneurysm • Development of Lymphatic Fistula.

  34. ARTERIAL CANNULA • Is the Narrowest part of the circuit. • Should be as Short as possible. • As Large as the diameter of vessel permits.

  35. Complications • Difficult Cannulation • Intramural Placement • Air embolism • Dislodgement of Cannula • Dissection • Arch Vessel Cannulation • Back Wall Injury

  36. MICROPORE FILTERS: • Remove Particulate Matter (Bone , Tissue , Fat , Blood Clots etc.) • Pore Size : 30 – 40  • ULTRAFILTRATION : • Remove the excess fluid from the CPB.

  37. PRIME FLUID • Ideally close to ECF. • Whole Blood NOT used : • Homologous Blood Syndrome. • Post Perfusion Bleeding Diathesis • Incompatibility Reactions. • Demand on Blood Banks. • Addition of Priming Fluid HEMODILUTION.

  38. COMPOSITION OF PRIME : • Balanced salt soln. RL 1250 ml • Osmoticallyactive agent 100 ml (Mannitol, Dextran 40 , Hexastarch) • NaHCO3 50ml • KCl10ml • Heparin 1ml

  39. PRIMING • Heme, nonheme • Decreases viscosity so better flow • Attenuates increased viscosity by hypothermia • Alters pharmacodynamics and kinetics of drugs • Decreases Hb but improves O2 delivery • Lowest acceptable value 8g/dl

  40. Prediction of initial haematocrit during CPB Predicted Hct = Pt. RBC volume before CPB /Pt. EBV + CPB prime volume • EBV • Infants 80-85 ml/kg • Children 75ml/kg • Adult (male) 70ml/kg • Adult (female) 65ml/kg • 1U packed cells = 0.7 x 350 = 245ml • IU whole blood = 0.4 x 350 = 140ml

  41. Amount of priming fluid • CVX CPCV = Pt. BV X PCV + PV X PCV • PT.BLOOD VOL. x PT. HEMATOCRIT = TARGET HCT X(PRIME VOL. + PT. BLOOD VOL.)

  42. PATHOPHYSIOLOGY OF CPB • THREE MAJOR PHYSIOLOGICAL ABERRTIONS ARE: 1.LOSS OF PULSATILE FLOW 2.EXPOSURE OF BLOOD TO NON-PHYSIOLOGIC SURFACES & SHEAR STRESSES. 3.EXAGGERATED STRESS RESPONSE.