190 likes | 307 Views
This document outlines the quality control procedures for pH and ICP (Inductively Coupled Plasma) techniques utilized at Clemson University. It includes step-by-step checks including the use of blanks and standard samples, verification of electrode performance, maintenance of equipment, and adjustments for gas flow rates. Attention is given to the cleaning and inspection of components to ensure accurate readings. Regular data checks and the stability of equipment parameters are emphasized to maintain the integrity of laboratory results.
E N D
pH and ICP Techniques Clemson
Scoop 2 checks • One blank • Double scoop 6 samples • Tag set with lab number, check, and blank positions pH quality control checks
Two QC check samples • Two samples at beginning, middle, and end of set • Cross Check electrodes #1 and #2 1 2 3 Quality control row
ck • Check both QC check values at beginning and within set • Check all double scooped samples at beginning, middle, and end of set • Buffer run - checks • Buffer run - blank (1:1, water:buffer) End of run verification
Friday • clean out electrode filling solution • fill with water • Ultrasonic cleaner • Store in buffer pH 7 for the weekend Maintenance
Monday • Take water out of electrode • Refill with KCl filling solution • Verify flow of KCl through filter • Refresh buffer solutions • Electrode (from Fisher) • Accumet rugged bulb pH combination electrode • 13-620-183A Maintenance – cont.
Check B data since April 1, 2010 • 115 continuous reads over a 2.5 month period • Rinse pH < 4.5 and > 7.0 • NAPT value 5.15 • Lab average 5.15 • stdev 0.036 pH Check values
Two checks within the set • One blank ICP quality control checks
Probe – Leeman Labs • Phone: 603-886-8400 • Part number 902-00017 • Check all tubing for cleanliness and kinks • Change pump winding tube when needed • Adjust pump winding clamp - look for smooth flow as sample travels though tube Sample entry
Aspirate yttrium stock solution • Look for the tip of the red bullet • Adjust nebulizer gas flow rate ARCOS TJA Nebulizer gas flow
Keep clean • Take apart to rinse out nebulizer and spray chamber parts • TJA – use cleaning wire • ARCOS • SeaSpray Nebulizer • Twister Spray Chamber • Cross Flow ARCOS ARCOS TJA Nebulizer and spray chamber
ARCOS – clean as needed – salt buildup • TJA – clean ~once a year TJA ARCOS Torch
Transient scan – run standard • Check to see when the peak intensities stabilize • Clemson – 35 second total flush • 25 seconds at 60 rpm • 10 seconds at 30 rpm 25 seconds P K Time (seconds) using pump speed of 60 rpm Setting the flush timeARCOS - Crossflow
18 seconds 25 seconds Crossflow spray chamber Twister spray chamber Transient scan on ARCOS
Options for changing flush time • Spray chamber • Tubing length • Pump speed • Sample loop P – 100 ppm Ca – 100 ppm K – 100 ppm Na – 50 ppm 45 seconds 25 seconds Time in seconds using pump speed of 60 rpm Transient scan going to rinseARCOS - Crossflow
Run time scan • Check time for stability • Clemson soil method • pump speed 200 rpm • flush time 30 seconds Ca - 40 seconds Pump speed 175 rpm Setting the flush time - TJA
Look at several recommended lines • Clemson - Ca line – 317.933 for soil Ca – 396.847Saturated signal Ca – 315.887 Line selection – ARCOS
Set peak position • Select clean area for background subtraction • Look for uncontaminated blank Peak position Cu – 324.754 Blank correction blank Method development - ARCOS
Run profile when needed • On the AROCS - check intensity of P158.599 line • If significant UV intensity drop – clean window and/or lens to optics • Check new standard against old • Read reference sample at start, mid point, and end of run • Verify 2 check samples and blank within set Instrument profile and data verification