shanon
Uploaded by
4 SLIDES
204 VIEWS
60LIKES

Optimizing ICalization Process for Instrument Calibration

DESCRIPTION

This guide outlines the procedure for effective ICalization and instrument warm-up. Begin by loading the ICalization solution and ensuring it is accessible in the designated location. If the falcon tube is low, refill it using the Genesis-ICAL solution. Next, clean the sample introduction system by flushing it with 3% HNO3 for 5 minutes, followed by Q-water for an additional 5 minutes. Finally, use the program to navigate to the ICal solution, and select “ICALization” from the “System” menu once the autosampler is positioned correctly.

1 / 4

Download Presentation

Optimizing ICalization Process for Instrument Calibration

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. ICP • ICALization

  2. Start • Start and warm up the instrument • Find the ICalization solution and have it open in the ICal location • Note: If the falcon tube appears to be empty or near empty it can be refilled with the solution from the bottle labeled Genesis-ICAL • Load your method

  3. Cleaning • Flush the sample introduction system with 3% HNO3 for 5 minutes • Flush with Q-water for 5 minutes • Reason: The instrument introduction system must be clean.

  4. Using the Program • In the window, right click on the ICAL solution location/spot and select “move to position” (Left click the ICAL circle then right click for the drag down menu). • Once the autosampler has moved to and been submerged in the ICAL solution select the function “ICALization” in the “System” menu.

More Related
SlideServe
Audio
Live Player
Audio Wave
Play slide audio to activate visualizer