Explore the impact of double event patterns on energy calibration shifts in X-ray datasets, examining singles versus doubles in various readout directions and illumination, using XSPEC for line fitting analysis. Identify systematic trends affecting calibration accuracy and propose corrections for improved data interpretation.
E N D
The problem with double event patterns (M. Brusa, K. Dennerl – MPE) EPIC CAL/OPS meeting - MPE
Classification of doubles SAS 2 Right SAS 3 Forward SAS 4 Left SAS 1 Backward readout direction split backward (SAS: „1“) Forward/Backward doubles: superposition of “true” doubles and “false” doubles (re-emission during read-out) cause shifts in energy calibration split right (SAS: „2“) split left (SAS: „4“) split forward (SAS: „3“)
SAS 2/4: (Left/Right doubles) resemble singles in illumination • SAS 1/3: (Forward/Backward doubles) different illumination the amount of forward and backward doubles vary across the detector
Q0 Q1 • 11 CALCLOSED observations (PN + FF + expo > 20 ks) spanning ~1000 revolutions from #84 to #1105 • Spectra extracted in 120 different positions (30x4 quadrants) for - SINGLES (PATTERN=0) - DOUBLES (PATTERN=1-4) • XSPEC: fit of the MnKα line (5.896 keV) and Al line (1.486 keV) tested against: - different matrices or models used - singles vs. doubles - backward vs. forward - MnKα vs. Al Q2 Q3
Examples of spectra (MnKα, Rev. #572, Q1) Singles (black) D-backward (red) D-forward (green) readout direction
Singles, Q1Single gaussian fit to MnKα REV #84 Mn line (5.896 keV)
Singles, Q1 REV #125 Mn line (5.896 keV)
Singles, Q1 REV #249 Mn line (5.896 keV)
Singles, Q1 REV #429 Mn line (5.896 keV)
Singles, Q1 REV #981 Mn line (5.896 keV)
Singles: different MATRICES used Singles: different MATRICES used • Best fit energy line does not depend on the actual matrix used REV #84
Singles: different MODELS used • Best fit energy line does not depend on the model used REV #84
Singles vs. Double (1 – backward) • REV #84 • Systematic trend • Position of SAS”1” inconsistent with values for the singles Singles (blue) D-backwards (red) readout direction
Singles vs. Double (1 – backward) • REV #125 • Systematic trend • Position of SAS”1” inconsistent with values for the singles Singles (blue) D-backwards (red) readout direction
Singles vs. Double (1 – backward) REV #572 Singles (blue) D-backwards (red) readout direction
Singles vs. Double (1 – backward) REV #857 Singles (blue) D-backwards (red) readout direction
Singles vs. Double (1 – backward)- “detector representation” - REV #84 “Curved shape” wrt singles Singles (blue) D-backward (red)
Singles vs. Double (1 – backward) REV #125 Systematic in all observations
Singles vs. Double (1 – backward) REV #249 Systematic in all observations
Singles vs. Double (1 – backward) REV #309 Systematic in all observations
Singles vs. Double (1 – backward) REV #429 Systematic in all observations
Singles vs. Double (1 – backward) REV #572 Systematic in all observations
Singles vs. Double (1 – backward) REV #605 Systematic in all observations
Singles vs. Double (1 – backward) REV #709 Systematic in all observations
Singles vs. Double (1 – backward) REV #857 Systematic in all observations
Singles vs. Double (1 – backward) REV #981 Systematic in all observations
Singles vs. Double (1 – backward) REV #1105 Systematic in all observations
Singles vs. Double (backward) vs. Double (left) Singles (blue) D-backward (red) Singles (blue) D-left (red) REV #125
Singles vs. Double (backward) vs. Double (forward) Singles (blue) D-backward (red) Singles (blue) D-forward (red) REV #125 No clear trend seen!
Al lineSingles vs. Double (1 - backward) Singles (blue) D-backward (red) REV #125
Mn vs. Al lines (singles vs Double 1- backward) REV #125 Al line MnKα line Smaller amplitude ..but similar shape!
Conclusions/Future developments • Singles: - Mn & Al line energy quite stable (in single observation) - Matrices/models adopted does not affect energy determination • Doubles: - most of problems caused by BACKWARD (systematic trend wrt to singles) - LEFT/RIGHT: similar to singles - FORWARD: no systematic trend… • Amplitude in Al line smaller than in Mn line - Energy AND spatial corrections needed