210 likes | 228 Views
The Crossvent 3 ventilator offers 3 modes of ventilation, from adult to pediatric range, with advanced features such as assist/control, volume and pressure control, sighs, SIMV, pressure support, and CPAP. Easily adjust settings using intuitive touch screen controls. Stay informed with LCD displays for key parameters. This portable ventilator is powered by a rechargeable battery, lasting up to 6 hours. Ensure patient safety with built-in alarms and safety features. Experience precise and reliable ventilation with Crossvent 3.
E N D
Modes – Touch Screen • Adult to pediatric range • Assist/Control – Volume Control • Assist/Control – Pressure Control • Sighs • SIMV • Pressure Support • CPAP
Turn knobs for setting: • Maximum pressure • PEEP • Flow
LCD Touch ScreenSelect Mode then press MAIN menu • Then press Rate & TV • adjust with up/down arrows • IT, ET & I:E Ratio displayed • Flow, PEEP & Max Press limit • adjust with dials
On the left side External Power Supply O2 Sensor ON/OFF Alarm Reset LED Charging Light Pt Circuit Pneumotach/Flow Sensor Input Airway Pressure Connector Exhalation Valve Connector
bmd CrossVent 3 • powered by a rechargeable NiMH battery • lasts 6 hours • BioMed 16vDC 3A charger/converter recharges internal battery and allows for AC operation • LCD screen has LCD charging indicator and visually large LOW BATTERY alarm
On the right side Gas Entrainment ON/OFF Bleed Exhaust Gas Supply Inlet (31 - 75 psig or 45 - 75 when using blender)
Gas Entrainment • When connected to 100% oxygen source gas with the Gas Entrainment ON, the oxygen concentration is diluted to 50%. • There will be some decrease in delivered flow rate with Entrainment ON under high load conditions
The rear panel • Maximum pressure relief valve (120 cm H2O) • Negative pressure relief valve – opens at - 4 cm H2O with machine shutdown
ALARMS • ALARM 1 • Peak Pressure • Respiratory Rate • Exhaled Tidal Volume (can be switched off) • Exhaled Minute Volume (can be switched off) • ALARM 2 • PEEP • Mean Pressure (can be switched off) • Oxygen (can be switched off)
To set Alarms Press ALARM 1 • Press High or Low Limits to Change • Adjust with up/down arrows
3rd Level Alarms Low Gas Supply Pressure Low Battery Ventilator Failure
With Alarm Condition: the value for the exceeded limit flashes red -1