fresenius medical care peritoneal dialysis connections l.
Download
Skip this Video
Loading SlideShow in 5 Seconds..
Fresenius Medical Care Peritoneal Dialysis Connections PowerPoint Presentation
Download Presentation
Fresenius Medical Care Peritoneal Dialysis Connections

Loading in 2 Seconds...

play fullscreen
1 / 12

Fresenius Medical Care Peritoneal Dialysis Connections - PowerPoint PPT Presentation


  • 1245 Views
  • Uploaded on

Fresenius Medical Care Peritoneal Dialysis Connections. Baxter to stay  safe Connection. Connections using Fresenius Medical Care products.

loader
I am the owner, or an agent authorized to act on behalf of the owner, of the copyrighted work described.
capcha
Download Presentation

PowerPoint Slideshow about 'Fresenius Medical Care Peritoneal Dialysis Connections' - Sophia


An Image/Link below is provided (as is) to download presentation

Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server.


- - - - - - - - - - - - - - - - - - - - - - - - - - E N D - - - - - - - - - - - - - - - - - - - - - - - - - -
Presentation Transcript
connections using fresenius medical care products
Connections using Fresenius Medical Care products

This inservice is not intended to replace the advice or training from your physician. Refer to the Instructions For Use (IFU) for detailed device description, instructions, contraindications, warnings, and precautions.

Fresenius Medical Care, DEL-CLAMP, Stay●Safe, Safe-Lock, DELFLEX and FREEDOM are trademarks of Fresenius Medical Care Holdings Inc. or its affiliated companies. All other trademarks are the property of their respective owners.

fresenius products
Fresenius Products

Baxter is a trademark of Baxter International Inc. or its affiliates.

Instructions on how to use Baxter products were reproduced from the manufacturer’s literature and were current at the time of print. Instructions are included for the patient's convenience but Fresenius Medical Care North America does not specifically endorse the use of these products or assume any liability related to their use. Should there be any differences between the instructions listed in this manual and the instructions for use provided with the medical devices referenced, the instructions provided with the medical device should be taken as accurate."

connections using fresenius medical care products4
Connections using Fresenius Medical Care products

Note: for clarity the model is shown not wearing personal protective equipment (PPE). Follow your facility’s policies and procedures for use of personal protective equipment.

It is vital to maintain sterility of the connectors during connection procedures.

baxter patient to stay safe system
Baxter patient to staysafe system

A staysafe Extension Set with Luer-Lock patient connector will allow the patient with a Baxter PD system to use staysafesupplies while in the hospital.

staysafeExtension Set with Luer-Lock patient connector

staysafe System

Baxter PD patient

baxter patient to stay safe system6
Baxter patient to staysafe system

Gather supplies

staysafe Extension Set with Luer-Lock patient connector

12 inch (P/N # 050-95004) Or

18 inch (P/N # 050-95005)

staysafesolution set and related supplies for transfer set connection and exchange procedure.

baxter patient to stay safe system7

ent extension set

Baxter patient to staysafe system
  • Remove the blue sterility protector from the end of the Fresenius staysafe Extension Set .
  • Remove the mini-cap from the Baxter patient extension set.
  • 3. Attach the Fresenius staysafe Extension Set to the end of the Baxter extension set. Proceed with the staysafe CAPD or APD exchange.

staysafe Extension Set

Baxter Extension Set

1

2

3

titanium luer lock adapter to stay safe system
Titanium luer lock adapter to staysafe system

staysafe Extension Set with Luer-Lock patient connector

12 inch (P/N # 050-95004) Or

18 inch (P/N # 050-95005)

staysafesolution set and related supplies for transfer set connection and exchange procedure.

Remove the blue sterility protector from the end of the Fresenius stay● safe extension set and attach to the luer lock catheter adapter.

discharge instructions for all connections
DISCHARGE INSTRUCTIONSFOR ALL CONNECTIONS
  • The patient should return to his/her home or clinic with the additional extension in place. This practice will allow the patient to return home and resume dialysis without interruption.
  • Instruct the patient to remove any adaptors added to their extension set at the beginning of the next exchange or as directed by the PD clinic nurse.
slide10

Choose the type of catheter or transfer set and the type of system utilized. Look down the column and across the row to the adapter needed. Click on the icon in the box to take you to the procedure

slide11

Please contact your Fresenius Clinical Sales Representative at

1-800-662-1237 if you have questions regarding the use of a Fresenius PD system or adapter.

slide12

CD Available through Customer Service

800-323-5188 (P/N # 100676-01)