An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Image Reconstructionfrom Projections J. Anthony Parker, MD PhD Beth Israel Deaconess Medical Center Boston, Massachusetts Caveat Lector Tony_Parker@BIDMC.Harvard.edu

  2. Projection Single Slice Axial

  3. Single Axial Slice: 3600 collimator Ignoring attenuation, SPECT data are projections

  4. Tc-99m keV 150 100 80 60 50 x htl(140 keV) ≈ 4 cm Attenuation: 180o = 360o

  5. Coronal / Sagittal Axial Multiple simultaneous axial slices Cardiac Perfusion Data Collection Special Case - 180o

  6. 1 2 2 heads: 900 rotation = 1800 data Dual-Head General-Purpose Gamma Camera: 900 “Cardiac” Position

  7. Inconsistent projections “motion corrected”

  8. Original data

  9. Single Axial Slice: 3600 0 0 0 0 0 0 0 0

  10. 0 60 x projection angle 60 0 x x Sinogram: ProjectionsSingle Axial Slice

  11. 1 head 2 heads 12 min 12 min 24 min Uniformity & Motion on Sinogram

  12. Backprojection tails Reconstruction by Backprojection

  13. Backprojection 2 projections 2 objects

  14. projection tails merge resulting in blurring

  15. (1/r) backprojection lines add at the point tails spread point out Projection -> Backprojection of a Point

  16. Projection -> Backprojection

  17. Smooths or “blurs” the image (Low pass filter) ((Convolution with 1/r)) Nuclear Medicine physics Square law detector adds pixels -> always blurs Different from MRI (phase) Projection->Backprojection Smooths

  18. (Projection-Slice Theorem) “k-space (k,)” detail low frequency 2D Fourier transform spatial domain spatial frequency domain

  19. f(u,v) ≠ 0, single 0,v f(u,v) ≠ 0, single u,0 f(u,v) ≠ 0, single u = v Spatial Frequency Basis Functions

  20. 1/k (Density of slices is 1/k) one projection multiple projections (Fourier Transform of 1/r <-> 1/k) Projection -> Backprojection: k-space

  21. Projection -> Backprojection blurs with 1/r in object space k-space 1/k ( 1/r<-> 1/k) Ramp filter sharpen with k (windowed at Nyquist frequency) k k Image Reconstruction: Ramp Filter

  22. Low pass, Butterworth – noise Ramp – reconstruct Low Pass Times Ramp Filter

  23. Ramp filter exactly reconstructs projection Efficient (Linear shift invariant) (FFT is order of n log(n) n = number of pixels) “Easily” understood What’s Good about FPB

  24. New Cardiac Cameras Solid state - CZT: $$$, energy resolution scatter rejection, dual isotope Pixelated detector: count rate & potential high resolution poorer uniformity Non-uniform sampling: sensitivity potential for artifacts Special purpose design closer to patient: system resolution upright: ameliorates diaphragmatic attenuation

  25. Single photon imaging (i.e. not PET) Collimators: image formation Sensitivity / resolution trade-off Resolution recovery hype “Low resolution, high sensitivity -> image processing = high resolution” Reality - ameliorates low resolution Steve Moore: “Resolution: data = target object” Can do quick, low resolution image * not resolution from reduced distance due to design Collimator Resolution*

  26. Dual Head: Non-Uniform Sampling

  27. keV 150 100 80 60 50 htl(140 keV) ≈ 4 cm Activity Measurement: Attenuation

  28. Gd-153 rods T1/2 240 d e.c. 100% 97 keV 29% 103 keV 21% 2 heads: 900 rotation = 1800 data Attenuation Correction: Simultaneous Emission (90%) and Transmission (10%)

  29. Semi-erect: Ameliorates Attenuation

  30. Leaning Forward, < 500 Pounds

  31. Digirad: Patient RotatesX-ray Attenuation Correction

  32. keV 150 100 80 60 50 CT: Polychromatic Beam -> Dose

  33. X-ray tube: electrons on Tungsten or Molybdenum characteristic X-rays e- interaction: - ionization - deflection bremsstrahlung X-ray Tube Spectra

  34. Digirad X-ray Source: X-rays on Lead 74W 82Pb X-rays interaction - ionization - no 10 bremsstrahlung

  35. Digirad X-ray Spectrum

  36. New Cardiac Cameras D-SPECT CardiArc Digirad GE Detector CZT* NaI(Tl) CsI(Tl) CZT* Electronics SS* PMT PD*? SS* Pixelated Y N Y Y Collimation holes slits*? holes pinholes Non-uniform Y* Y* ~N Y* Limited angle Y Y N ~N Closer to pt Y Y Y ~N AC N CT? CT* CT Position ~semi semi erect supine

  37. breast lung Soft Tissue Attenuation: Supine

  38. breast Soft Tissue Attenuation: Prone

  39. breast post Soft Tissue Attenuation: Digirad Erect

  40. Sequential Tidal-Breathing Emission and Average-Transmission Alignment

  41. Non-uniform sampling -> sensitivity Special purpose design -> resolution Advantages Throughput at same noise Patient motion - Hx: 1 head -> 2 head Cost Non-uniform sampling -> artifacts History: 7-pinhole - failed 180o sampling - success Sequential emission transmission Sensitivity / Resolution Trade-Off

  42. What’s Wrong with FilteredBackprojection, FBP, for SPECT Can’t model: Attenuation Scatter Depth dependant resolution New imaging geometries (Linear shift invariant model)

  43. Solution Iterative reconstruction Uses: Simultaneous linear equations Matrix algebra Can model image physics (Linear model)

  44. But, exact solution for a large number of equations isn’t practical Projections as Simultaneous Equations(Linear Model)

  45. n projection backprojection estimate object data f p H - + + r model estimated data estimate error x ^ ^ ^ ^ fn f0 fn-1 ^ H A pn-1 A estimate + backprojected error ^ en-1 Iterative Backprojection Reconstruction

  46. n f p H - + + r x ^ ^ ^ ^ f0 fn fn-1 ^ H A pn-1 A ^ en-1 Reconstruction, H, can be Approximate

  47. n f p H - + + r x ^ ^ ^ ^ f0 fn fn-1 ^ H A pn-1 A ^ en-1 ^ Accuracy of Model, A, is Key

More Related