Ultrasound Guided Central Venous Cannulation (CVC)

Ultrasound Guided Central Venous Cannulation (CVC) . Shirley Lee MD CAEP 2008. Why Ultrasound guidance?. Traditionally, CVC mechanical complications occur up to 15% Insertion unsuccessful up to 12% Becoming standard of care . McGee and Gould, NEJM 2003; 348: 1123-1133

omer
Download Presentation

Ultrasound Guided Central Venous Cannulation (CVC)

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


    1. Ultrasound Guided Central Venous Cannulation (CVC) Shirley Lee MD CAEP 2008 Disclaimer - unless same folks as last talkDisclaimer - unless same folks as last talk

    2. Why Ultrasound guidance? Traditionally, CVC mechanical complications occur up to 15% Insertion unsuccessful up to 12% Becoming standard of care The use of ultrasound has obviated the absolute need for anatomical landmarks in the traditional method of CVC insertion. Ultrasound can readily identify the locations of veins relative to arteries and therefore can take into account for individual anatomic differences (Gordon AC, Troianos CA, Denys BG). The use of ultrasound in experienced hands decreases the number of attempts and arterial punctures compared with landmark method. It is also the preferred method in difficult cannulations such as those patients with obscure surface landmarks or hypotensive patients (Hind D, Calvert N, Miller AH, Keenan SP, Fyr WR). Several meta-analyses have been performed and national recommendations have been made by both the National Quality Forum in the United States and the National Institute of Clinical Excellence in the United Kingdom (The National Quality Forum, National Institute for Clinical Excellence). Merrer J, De Johnghe B, Golliot F, et al. Complications of femoral and subclavian venous catheterization in critically ill patients: a randomized controlled trial. JAMA. 286:700-7, 2001 The use of ultrasound has obviated the absolute need for anatomical landmarks in the traditional method of CVC insertion. Ultrasound can readily identify the locations of veins relative to arteries and therefore can take into account for individual anatomic differences (Gordon AC, Troianos CA, Denys BG). The use of ultrasound in experienced hands decreases the number of attempts and arterial punctures compared with landmark method. It is also the preferred method in difficult cannulations such as those patients with obscure surface landmarks or hypotensive patients (Hind D, Calvert N, Miller AH, Keenan SP, Fyr WR). Several meta-analyses have been performed and national recommendations have been made by both the National Quality Forum in the United States and the National Institute of Clinical Excellence in the United Kingdom (The National Quality Forum, National Institute for Clinical Excellence).Merrer J, De Johnghe B, Golliot F, et al. Complications of femoral and subclavian venous catheterization in critically ill patients: a randomized controlled trial. JAMA. 286:700-7, 2001

    3. Right pneumothorax, post IV insertionRight pneumothorax, post IV insertion

    4. CVC Complications Pneumothorax, arterial puncture, hematoma, malposition, increased skin punctures with bleeding complications, delay and failure to catheterize, thoracic duct injury (left sided approach), air embolism, arrhythmias, and death Delayed complications: infection and thrombosis

    5. Advantages Safer Markedly decreased pneumothorax rate Real-time visualization of target See needle enter target vein, avoiding adjacent arteries, nerves Tseng, Sadler et al, CARJ 2001;5(6):379-84 (i.e., 0/3267 at Foothills Hospital), high success (>99%) Tseng, Sadler et al, CARJ 2001;5(6):379-84 (i.e., 0/3267 at Foothills Hospital), high success (>99%)

    6. Evidence In ER: Prospective, randomized trial of 130 patients Complication rate: 4.6% vs. 16.9% Success rate: 93.9% vs. 78.5% Time not significantly different

    7. Evidence ICU – Prospective,randomized trial of 900 Success: 100% vs. 94% Carotid puncture 1% vs. 10% Pneumothorax 0% vs. 2.4% Hemothorax 0% vs. 1.7% Reduced time and number of attempts with US Karakitsos et al, Crit Care,2006;10(6):R162Karakitsos et al, Crit Care,2006;10(6):R162

    8. Jugular and Subclavian anatomy

    9. Landmarks: SCM muscle above, central vein larger, more irregular in shape vs carotid artery smaller, rounder, prominently pulsatile. Right side - vein compresses easily.Landmarks: SCM muscle above, central vein larger, more irregular in shape vs carotid artery smaller, rounder, prominently pulsatile. Right side - vein compresses easily.

    10. Subclavian vein

    11. Common indications for CVC Hemodynamic monitoring Administration of drugs likely to induce phlebitis Temporary cardiac pacemaker Hemodialysis Lack of peripheral venous access

    12. Technique 5-10 MHz probe - locate vein, ensure patency, then puncture blindly – but no safer than landmark technique Real-time visualization of needle tip helps prevent pneumothorax, arterial puncture Still need X-ray to document tip position, as catheter can still go wrong direction START by facing caudad, facing patient’s feet, probe marker to patient’s LEFT (not traditional patient’s right). SINGLE OPERATOR: Probe in non-dominant hand, needle and syringe in dominant hand. Beward catheter can go to other innominate vein, azygous, deep in right atrium! START by facing caudad, facing patient’s feet, probe marker to patient’s LEFT (not traditional patient’s right). SINGLE OPERATOR: Probe in non-dominant hand, needle and syringe in dominant hand. Beward catheter can go to other innominate vein, azygous, deep in right atrium!

    13. Technique Use either transverse or longitudinal orientation of ultrasound beam to needle path Transverse supposedly easier for novice Advantage of longitudinal: see needle through entire course With either, you will NOT see needle tip if out of plane Can use needle guides to help

    14. Predictors of difficult cannulation Emergency Placement Obesity Coagulopathy Intubated Hypotensive/Hypovolemic Edematous patient Known previous difficult cannulation

    16. Static vs Dynamic technique Static technique: Using 2-3 ultrasound planes, mark with felt pen Dynamic technique: 1 vs 2 person technique Use image as guide Observe needle throughout procedure, as it penetrates vessel Watch tip! 2 person process: Scan for vein first (non sterile), mark site. Then give probe to nurse to prep for sterility, while you prep sterile dress and patient and prep equipment, then probe handed back to you. While you scan patient, nurse preps for sterile dress, and when you find the vein, grab nurse’s hand to hold position (NB - sometimes better if nurse does not see imaging screen, so won’t keep moving around). You proceed with cannulation.2 person process: Scan for vein first (non sterile), mark site. Then give probe to nurse to prep for sterility, while you prep sterile dress and patient and prep equipment, then probe handed back to you. While you scan patient, nurse preps for sterile dress, and when you find the vein, grab nurse’s hand to hold position (NB - sometimes better if nurse does not see imaging screen, so won’t keep moving around). You proceed with cannulation.

    17. Venous access – easiest to more challenging 1st CHOICE: Internal/External Jugular 2nd CHOICE: Femoral – easy 3rd CHOICE: Subclavian/Axillary – harder due to location, more difficult to visualize while you puncture 4th CHOICE: Cephalic/Basilic/Antecubital – 4th choice: harder due to small size 4th choice: harder due to small size

    18. Jugular Vein Large, easy to see, good choice Trendelenberg, head contralaterally turned 30 degrees Put probe transversely across vein, just superior to clavicle btn two SCM heads, just superior to clavicle Bring needle in from laterally above probe (in same plane as transducer), aiming just slightly down to toes ~ 20-45 degrees (Posterior approach) Watch needle well away from vein, indenting vein wall, and pop through…and know where carotid is! Beware of anatomical variants: Widest diameter just superior to clavicle - is closer to lung, hence need steeper angle of 45-60 degrees to safely avoid lung. Widest diameter just superior to clavicle - is closer to lung, hence need steeper angle of 45-60 degrees to safely avoid lung.

    19. Subclavian Vein More challenging to see needle and vein at puncture site Jugular much easier, less risk of venous stenosis, thrombosis, catheter fracture from pinch off syndrome Place probe inferior to most lateral aspect of clavicle Puncture axillary-subclavian junction close to clavicle Infraclavicular or supraclavicular technique. Infraclavicular or supraclavicular technique.

    20. Femoral vein Also easy Orient transducer longitudinally, along course of vein, bring needle in from below, parallel to transducer and vein Valsalva often helps distend vein, bigger target Externally rotate leg to move artery more lateral NAVEL (lateral to medial leg) . Avoid going too distal for needle insertion, as femoral artery tends to become more anterior to femoral vein, the more distal you go from the inguinal ligament. NAVEL (lateral to medial leg) . Avoid going too distal for needle insertion, as femoral artery tends to become more anterior to femoral vein, the more distal you go from the inguinal ligament.

    21. U/S CVC Pitfalls 1. Failure to identify the vein correctly 2. Failure to locate the needle in tissue

    22. Tips Awake patient - Check position. If patient has moved after you have landmarked, this results in a change in anatomical position of the vein Centre vein in middle of the screen Lighten probe pressure, as may be collapsing vein Insert needle at sharper angle (45 - 60 degrees), to properly intersect with the vein directly under transducer Keep acoustic shadow and ring down artifact in center of vessel, if is off center, withdraw slowly and redirect, using depth markers to help guide needle insertion. ALSO to prevent drift of the transducer due to slippery gel, a steady US position can be maintained by resting BOTH hand and transducer onto the patient’s body (be careful not to place undure pressure that may collapse the target vessel during the procedure)Keep acoustic shadow and ring down artifact in center of vessel, if is off center, withdraw slowly and redirect, using depth markers to help guide needle insertion. ALSO to prevent drift of the transducer due to slippery gel, a steady US position can be maintained by resting BOTH hand and transducer onto the patient’s body (be careful not to place undure pressure that may collapse the target vessel during the procedure)

    23. 7 Steps to Success: Use adequate gel Confirm orientation of probe - conventionally probe head pointing to RIGHT (rub edge with finger, look at screen) Do preliminary US - find patent target vein Mark site (static vs direct technique) Consider local anesthetic Sterilize skin, sterile probe, sterile technique! Advance the needle!

    24. Sterile preparation of US transducer Apply non-sterile gel to probe Slip sterile sleeve over transducer, smooth all air bubbles away from scanning surface to prevent artifact Secure sleeve with rubber band Alternate: large sterile glove, with fingers folded over, palmar surface of glove is scan surface. Sterile gel applies outer surface of glove/sterile sleeve

    25. General Tips on CVC insertion Be aware that more than 3 failed attempts to cannulate the vein can result in a 6 fold increase in mechanical complication. (McGee) Aids to distinguish arterial vs. venous cannulation A pressure transducer can be attached to the needle cannulating the vessel to confirm the presence of venous waveforms and pressure Blood gases from the needle in the vessel can be measured and compared with known a known arterial sample

    26. Summary US guided procedures have a number of clinical utilities in the ED US can improve the safety of specific procedures and success rate Initially, can be more time consuming due to learning curve, but with practice, patience and good hand-eye coordination improve efficiency and efficacy of patient care

    27. References Abboud PAC and Kendall JL. Ultrasound guidance for vascular access. Emergency Clinics of North America. 22(3):749-773, 2004 Leung, Duffy and Finckh. Ann Emerg Med 2006;48(5):540-7 McGee DC and Gould MK. Preventing complications of central venous catheterization. NEJM. 348(12):1123-33, 2003 Miller AH, Roth BA, Mills TJ et al. Ultrasound guidance versus the landmark technique for the placement of central venous catheters in the emergency department. Academic Emergency Medicine. 9(8) :800-805, 2002.

    28. Procedure Video Reference http://content.nejm.org NEJM video - (look under Procedure videos on right side of webpage) – download to ipod, mem stick, etc.NEJM video - (look under Procedure videos on right side of webpage) – download to ipod, mem stick, etc.

    29. References National Institute for Clinical Excellence. NICE technology appraisal guidance No.49: guidance on the use of ultrasound locating devices for placing central venous catheters. London: NICE, September 2002. www.nice.org.uk/pdf/ultrasound_49_GUIDANCE.pdf (accessed 21 Apr 2004) The National Quality Forum. Safe Practices for Better Health Care. A consensus report.Washington, D.C. 2003. www.qualityforum.org (accessed 27 Jan 2005).