UCCS Mechanical Engineering Senior Design Team Final Review. Jake Gelsinger David Hills Derick Jensen Ben Placzek. Overview. Brief Procedure Introduction Vein Analysis Syndaver Labs Synthetic Vein Tissue Testing FEA Synthetic Vein Tissue Analysis

Download Presentation

UCCS Mechanical Engineering Senior Design Team Final Review

An Image/Link below is provided (as is) to download presentation

Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server.


- - - - - - - - - - - - - - - - - - - - - - - - - - E N D - - - - - - - - - - - - - - - - - - - - - - - - - -

Presentation Transcript


Uccs mechanical engineering senior design team final review

UCCS Mechanical Engineering Senior Design TeamFinal Review

Jake Gelsinger

David Hills

Derick Jensen

Ben Placzek


Overview

Overview

  • Brief Procedure Introduction

  • Vein Analysis

    • Syndaver Labs Synthetic Vein Tissue Testing

    • FEA Synthetic Vein Tissue Analysis

    • FEA Vein Environment Modeling Analysis

  • Prototype

    • Assembly/Construction Progress

    • Proof of Concept


Vein extraction procedure

Vein Extraction Procedure

  • Prepare the Pacemaker Pocket

    • Detach Pacemaker Leads

  • Prepare Leads

  • Insert LLD (Lead Locking Device)

    • Expand LLD to lock in place

  • Deploy Laser Over Lead and LLD

  • Lase through all Fibrous Tissue

  • Problem Lies in Fibrous Occlusions


Syndaver labs tissue testing

Syndaver Labs Tissue Testing

  • Testing Procedure

    • Lowering tissue onto laser tip (12 french)

    • Reading force tissue exerts on laser on a scale

    • Increase force by lowering tissue until rupture

  • 18 Tests were conducted

    • 6 tests at three different angles were tested

    • Tried to model the angles that may occur during surgery

    • Each test lasted an average of 1 minute and 53 seconds


Sythetic vein fea

Sythetic Vein FEA

  • COMSOL was the software used

  • Limits of Software

    • Do not know true material properties of Syndaver tissue

    • Cannot model tearing

    • Cannot model soft tissue

      • Non-linear

      • Non-homogeneous

      • Transient response


Vein environment fea

Vein Environment FEA

  • Software Used was SolidWorks

  • Limits of Software

    • Cannot model tearing

    • Unknown material properties (Poisson’s Ratio)

    • Cannot accurately model vein constraints

  • Cannot model soft tissue

    • Non-linear

    • Non-homogeneous

    • Transient response


Prototype

Prototype

  • Design made to minimize cost while preserving durability and usability

  • Construction was difficult to get coordinated

    • Mind Studios vs. Machine Shop

  • End Result

    • Proof of Concept Design

    • Recommendations