Targeted Left Ventricular Lead Placement to Guide Cardiac Resynchronisation Therapy: A Randomised Controlled Trial (TARG

Targeted Left Ventricular Lead Placement to Guide Cardiac Resynchronisation Therapy: A Randomised Controlled Trial (TARGET study). Khan FZ 1,2 , Virdee MS 2 , Palmer CR 3 , Pugh PJ 1 , O’Halloran D 1 , Elsik M 2 , Read PA 2 , Begley D 2 , Fynn SP 2 , Dutka DP 1.

carnig
Download Presentation

Targeted Left Ventricular Lead Placement to Guide Cardiac Resynchronisation Therapy: A Randomised Controlled Trial (TARG

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Targeted Left Ventricular Lead Placement to Guide Cardiac Resynchronisation Therapy: A Randomised Controlled Trial (TARGET study) Khan FZ1,2, Virdee MS2, Palmer CR3, Pugh PJ1, O’Halloran D1, Elsik M2, Read PA2, Begley D2, Fynn SP2, Dutka DP1 1Addenbrooke’s Hospital, Cambridge, UK 2Papworth Hospital, Cambridge, UK 3Centre for Applied Medical Statistics, Cambridge, UK

  2. Disclosures Disclosures: None Sponsorship: Papworth Hospital, Cambridge, UK Funding: NIHR UK Papworth Hospital, Addenbrooke’s Charitable Trust Trial Registration: ISRCTN19717943

  3. Introduction • Cardiac Resynchronisation Therapy (CRT) has evolved as established treatment for advanced heart failure symptoms, impaired LV systolic function and intraventricular conduction delay • 30-40% of patients fail to gain significant clinical benefit • Left ventricular (LV) lead position has emerged as an important determinant of response

  4. LV Lead Position in CRT Higher response when pacing the latest site of contraction1 Lower response when pacing areas of scar2 1 Ypenburg et al J Am Coll Cardiol.2009;53(6):483-490 2Bleeker et al Circulation 2006;113(7):969-976.

  5. Hypothesis • There have been no randomised controlled trials to investigate the impact of LV lead position • The TARGET study designed to prospectively assess the feasibility of a targeted approach to LV lead placement and the impact of LV lead targeting on CRT outcomes • We tested the hypothesis that targeted LV lead placement to the latest site of contraction and away from areas of scar would enhance CRT response when compared to usual (unguided) treatment

  6. Single blinded, randomised controlled trial, 2 UK centres Inclusion: Sinus rhythm, NYHA Class III/IV, LVEF<35%, QRS>120ms Baseline (all patients): LV volumes, EF, NYHA, 6MWT, MLHFLQ Randomisation CONTROL Group TARGET Group Standard CRT LV Lead Placement without echocardiographic guidance Targeted LV Lead Speckle Tracking Echocardiography used to identify optimal site and LV lead targeted to this site All CRT devices optimised using echo following implant Study Design

  7. Speckle Tracking Echocardiography to Identify Optimal Site – Latest Site

  8. Speckle Tracking Echocardiography to Identify Optimal Site – Scar Determination Speckle tracking radial strain imaging correlates with delayed enhanced CMR imaging for determination of scar1,2 In CRT patients, a <10% amplitude of radial strain at the LV pacing site has a high negative predictive value (91%) for response (LV reverse remodelling)3 1 Becker et al J Am Coll Cardiol 2008 51(15):1473-1481 2Delagado et al Circulation 2011 123 (1):70-8 3Khan et al J Am Soc Echocardiogr. 2010 23(11):1168-1176

  9. Step 1: Identify optimal site as the latest site with an amplitude >10% to signify freedom from scar Step 2: Coronary sinus venography in steep left anterior oblique (LAO) view (50-90º) and coronary vein closest to optimal site identified Step 3: LV lead placed to optimal site (anterior, lateral, posterior or inferior in LAO view and basal or mid LV in the RAO view) Step 4: LV lead position correlated with echocardiographic data and described as: Concordant (pacing the optimal site) Adjacent (1 segment away) Remote (≥2 segments away from the optimal site) LV Lead Targeting – TARGET Group

  10. Study Endpoints Primary endpoint • Comparison of response rates between groups (≥15% reduction of LVESV at 6 months) Secondary endpoints • Clinical response rates (≥1 improvement in NYHA class) • All cause mortality • Combined mortality & heart failure hospitalisation Statistics: 80% power to identify a 20% difference in response rates between groups (one-sided α value of 0.05)

  11. Assessed for eligibility (n=247) • Excluded (n= 27) • Inadequate images to perform speckle tracking echocardiography Randomized (n= 220) ALLOCATION • TARGET Group (n=110) • Died prior to receiving CRT (n=1)* • Failure to implant an LV Lead (n= 4) • CONTROL Group (n=110) • Died prior to receiving CRT (n=1) * • Failure to implant an LV Lead (n= 3) FOLLOW UP Lost to follow-up (n=6) * Died prior to 6 month follow up (n=3) Lost to follow-up (n=5) * Died prior to 6 month follow up (n=4) ANALYSIS *Excluded from analysis (n=7) All patients included for long term endpoints Total data analysed n=103 for primary and secondary endpoints *Excluded from analysis (n=6) All patients included for long term endpoints Total data analysed n=104 for primary and secondary endpoints

  12. Baseline Characteristics

  13. Latest Site and LV Lead Position

  14. LV Lead Targeting

  15. Procedural Characteristics

  16. Primary Endpoint Response Rates: TARGET vs. Control (70% vs. 55%, p=0.031) Absolute difference in the primary endpoint of 15% [95% CI (2%, 28%)]

  17. Secondary Endpoint Clinical Response Rates: TARGET vs. Control (83% vs. 65%, p=0.003)

  18. Target vs. Control

  19. All Patients: Effect of LV Lead Position

  20. All Patients: Effect of Scar

  21. Conclusions • The TARGET study is the first randomised controlled trial of LV lead targeting • Targeted LV lead placement is feasible and associated with enhanced CRT response • Concordant LV lead placement, baseline dyssynchrony and pacing away from areas of scar are strongly related to improved CRT outcomes