Results from the PMT test bench in BCN (10-Oct-03)

Loading in 2 Seconds...

play fullscreen
1 / 14

Results from the PMT test bench in BCN (10-Oct-03) - PowerPoint PPT Presentation


  • 100 Views
  • Uploaded on

Results from the PMT test bench in BCN (10-Oct-03). J.Garra, R.Graciani, E.Graugés, X.Xirgu. 8 stage PMT. CFD base: The output of 1 channel studied to perform a premilinary analysis on: Signal Shape Gain Number of photo-electrons. CFD active base. Signal Shape.

loader
I am the owner, or an agent authorized to act on behalf of the owner, of the copyrighted work described.
capcha
Download Presentation

PowerPoint Slideshow about ' Results from the PMT test bench in BCN (10-Oct-03)' - breena


An Image/Link below is provided (as is) to download presentation

Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server.


- - - - - - - - - - - - - - - - - - - - - - - - - - E N D - - - - - - - - - - - - - - - - - - - - - - - - - -
Presentation Transcript
results from the pmt test bench in bcn 10 oct 03

Results from the PMT test bench in BCN (10-Oct-03)

J.Garra, R.Graciani, E.Graugés, X.Xirgu

8 stage pmt
8 stage PMT
  • CFD base:
    • The output of 1 channel studied to perform a premilinary analysis on:
      • Signal Shape
      • Gain
      • Number of photo-electrons
signal shape
Signal Shape
  • Some oscillations in the signal shape are supposed to be caused by some connection(?) problem.
gain vs photo statistics
Gain vs Photo-Statistics
  • Possible saturation effects might be observed starting around 50 phe and above
slide6
Gain
  • Our desired HV working point should be ~ 600 Volts obtaining an approximate gain of 30k to 50k
8 stage pmt1
8 stage PMT
  • BCN base:
    • A preliminary study has been performed with a LED source (no fibre)
    • Gain and Photo-statistics have been looked at from the output of a 1 channel (#64)
    • Now the setup is ready to get measurements for all the channels at once.
    • Fatal accident ended up with the 8-stage PMT becoming unusable anymore
slide11
The oscillations present in the first attempt (CFD) could be partially removed (with the BCN base) by when the setup has been re-done very carefully and assuring a common grounding for the (ADC/Black Box/Active Base) system.
  • Still to be tried if that also cured the oscillation observed using the CFD base
12 stage pmt
12 stage PMT
  • This PMT has already incorporated the active base by Hamamatsu together with the (PMT/Base) system encapsulation
  • Scans of 50x50 measurements each have been performed for different angles to ensure the geometrical consistency of the setup (PMT plane parallel to the fiber scaning plane): 0,90,180 and 360 deg. scans
  • Also scans with different light intensities have been taken to test the uniformity in the linear response of each channel.
2d scans
2D Scans
  • Different angles
  • Different light I
to do list
To do list
  • Let the new team get more experienced
  • Complete a final version of the fine scans (Motor lost steps, etc…).
  • Understand any the level of saturation of the signal
  • Single photo-electron measurement using a 2-stage amplifier board (work in progress)