An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Team Measuring Ocular Microtremor

  2. The Brainstem Has 3 components 4 cranial nerves on each section

  3. Assessing Brainstem Activity • “High Tech” • BIS Technology • Expensive (sensors & trained technicians) • Can’t use if swelling from head injuries • Glasgow Coma Scale • “Low Tech” • Cranial Nerves

  4. Unconscious Patients • Very uncomfortable procedures to assess intactness of cranial nerves • Flush 60 oz. of ice cold water in ear

  5. Ocular Microtremors (OMT) • OMT have been shown in clinical studies to accurately reflect brainstem activity • Ocular motor nerves originate in the brainstem where sensory signals are pre-processed and sent to the cerebrum • Oculo-motor nerves originate in the brainstem • These signals are necessary for consciousness

  6. Objective 1 Packaging & Architecture Incorporate a disposable element into the sensor mounting apparatus. Redesign and re-package the electronics for miniaturization and packaging for in-line dongle-type element on the cable.

  7. Objective 2 Signal Processing - Signal processing for signal amplification and/or noise reduction

  8. Objective 3 Integration Into Patient Overhead Monitor Research and understand the standard(s) used to integrate into this technology Implement in software with potential hardware modifications for the physical interface.

  9. Current Trend Analysis Software

  10. Component Overview Waveform Generator Bedside Monitor (Philips MP-60/70 or Agilent V24/26) Test MUX OMT Simulator Philips VueLink Module – M1032A Sensor (Provided) Analog Input/ Preprocessing Circuitry - Amp Microprocessor with Integrated A/D, UART, and DSP For MCU, currently looking at Atmel AVR32 with 12-bit A/D and integrated DSP functions,50 MHz clock. Power Supply XRAM

  11. OMT Sensor • Composed of a piezoelectric transducer, and a surface mount IC amplifier • The sensor generates voltage when it undergoes stress, in this case eye movement • The piezoelectric transducer was designed to be really sensitive to slight amounts of stress

  12. Power Management • The current “Blackbox” is powered by a medical grade power supply which provides ±12V for the sensor, and 5V for the processing unit • The Atmel AVR32 processor has a maximum voltage rating of 3.6V • The interface module to the bedside monitor will be powered by the monitor itself

  13. Power Management continued • Originally the bedside monitor was going to be used to power both the processor and sensor • Power options • Dual PS/2 – output voltage 5V ±10% • Dual MIB/RS232 – output voltage 5V ±5% • USB 2.0 – low power mode, output voltage 4.4V • It is more convenient to construct our own custom power supply that is capable of powering both the sensor and processor

  14. Power Flow Sensor Signal processing unit Monitor interface Power supply Bedside monitor

  15. Sample Signal until XRAM Buffer Full Broad Overview of Software Perform DSP Algorithm on Buffer Contents Extract Necessary Data From DSP Output Send Wave Samples, Freq, and Amp. to UART(Monitor)

  16. Sampling • Sampling will be interrupt driven, and will continue until a buffer a buffer is filled. Buffer must be large enough for an adequate analysis of signal. • When the first buffer is full, the DSP algorithm will be started on the signal and sampling will continue to a secondary buffer of equal size. Sampling and DSP will alternate between buffers so no data is lost. • The buffer size and sampling rate will be configurable over the UART. XRAM Buffer 1 A/D Buffer 2 DSP DSP Output Bus

  17. FFT Method FFT – Generate Output Array in XRAM Search Valid Range of Array for Max Energy -An FFT is performed on the contents of the sample buffer, with spectrum data saved to an array. -The array locations corresponding to the valid OMT range are searched for the maximum energy, giving the Frequency and Amplitude of the OMT signal. Amplitude/ Frequency Output Zero Out Locations Outside of OMT Range (Filter) -An inverse FFT on only the valid OMT frequency range is required to reconstitute a time-domain waveform that can be sent to the bedside monitor. Inverse FFT to Obtain Filtered Waveform in Time Domain Filtered Waveform Output

  18. Peak-Detection Method Digital HPF/LPF/Notch Filter Routines Modify Buffer • Digital filter routines modify the contents of the buffer, giving a filtered signal that could be sent to the bedside monitor. • Positive-to-negative changes in slope are counted, giving the frequency. The amplitudes at which this occurs are averaged. Filtered Waveform Output Peak-Count and Amplitude-Averaging Algorithm Amplitude/ Frequency Output Wavelets – Short Oscillations of a given frequency. Can be compared to our signal to determine if energy at that frequency is present.

  19. Bedside Monitor Interface • Philips VueLink Module 1032A • RS-232 Input – Open Interface Protocol • Allows 1 Waveform (OMT Signal), 2 Numerics (Amplitude and Frequency), Diagnostic Messages (ex. OMT out of valid range), and Alarms (ex. OMT below certain value) http://www.sentec.ch/fileadmin/documents/manuals/EN-HB-005928-b-SDM_VueLink_Installation_Manual.pdf

  20. Testing and Data Mining • Important in order to determine the pattern and/or trend of OMT signals • Determining a pattern is vital to the signal processing component • The process involves a subject (one of us) to lie down and have the sensor mounted on a closed eyelidusing surgical sensitive tape • Due to a lack of FDA approval and other required certification, we will not be able to gather data on a large scale basis

  21. Vivian’s OMT

  22. Mark’s OMT

  23. Division of Labor

  24. Schedule First Milestone Second Milestone Pre-Expo Grants Throughout Semester

  25. Budget

  26. Interference Concerns • Accidental patient movement • Electrical noise from other equipment in the ICU room • Noises in excess of 95 dB • Heartbeat

  27. Contingencies • Software implementation of display if bedside monitor is not able to be acquired • Scaling number of processors to ensure proper operation • Purchasing additional components to ensure minimal downtime especially before expo

  28. Risks • During application of sensor patients’ eye may be agitated • Additional cord is hanging off of patient • Possibility of dongle removing sensor from patients’ eye if too heavy • Components may be destroyed if wired incorrectly

  29. Extension possibilities • Add through connections in order to minimize hanging cables • Adding accelerometer to measure patient movement to find bad samples

  30. Questions?