1 / 9

Experiment 23

Experiment 23. Determination of the iron content with Spectrophotometry. Purposes 1. Learn to operate 722 Spectrophotometer. 2. Master to determine the iron content by use of sulfosalicylic acid method. Principles.

allanf
Download Presentation

Experiment 23

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Experiment 23 Determination of the iron content with Spectrophotometry Purposes 1. Learn to operate 722 Spectrophotometer. 2. Master to determine the iron content by use of sulfosalicylic acid method.

  2. Principles In this experiment, we use Fe3+ ions to react with sulfosalicylic acid and produce brown [Fe (Ssal) 2]- ions under a condition of HAc-NaAc buffer solution (pH=5). Determine the absorbance with λ=460nm. The Lambert-Beer law provides the mathematical correlation between absorbance and concentration. It is usually stated as

  3. The relationship between A and c serves as the basis for the quantitative analysis of a great many substances.

  4. λ=460nm · Standard curve · · Ax ········· · · ··········· · cx Standard Curve · • c0 c1 c2 c3 c4 c5 • A0 A1 A2 A3 A4 A5 · · · The graph should be a straight line ·

  5. Procedure • Prepare standard solutions and • a sample solution Prepare a series of standard solutions of different accurately known concentration. Get 6 volumetric flasks (10mL) and use the measuring pipets to transfer solutions into the volumetric flasks (according to the table). Prepare a sample solution of unknown concentration.

  6. Preparation of Fe3+ standard solutions and a sample solution

  7. 2. Determine of absorbances (A) According to the usage of 722-type spectrophotometer, use λ=460nm and use blank solution (1#) as reference solution to determine absorbance of each standard solution and a sample solution. Record the results.

  8. 3. Draw a calibration curve Draw a working curve. The abscissa is the concentration of Fe3+of standard solutions, the ordinate is the absorbance of the standard solution.

  9. 4. Determine the [Fe3+] of sample solution According to the determined absorbance of the sample solution, We can look up the concentration of Fe3+ in sample from the working curve drew by yourself. Calculate the concentration of Fe3+ in original undiluted sample. [Fe3+] (original undiluted sample) =[Fe3+] (diluted sample) ×10