1 / 37

Charles E. Hill Emory University School of Medicine

Converting BCR-ABL1 to the International Scale: Standardizing the Measurement of Relative Gene Expression. Charles E. Hill Emory University School of Medicine. Outline. Measuring BCR-ABL1 The International Scale Approaches for calibrating to the IS

zorion
Download Presentation

Charles E. Hill Emory University School of Medicine

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Converting BCR-ABL1 to the International Scale: Standardizing the Measurement of Relative Gene Expression • Charles E. Hill • Emory University School of Medicine

  2. Outline • Measuring BCR-ABL1 • The International Scale • Approaches for calibrating to the IS • Challenges of relative gene expression testing

  3. Disclosures • No financial interests/compensation from any of the companies discussed • Assay kits were provided for evaluation by both Asuragen and Ipsogen

  4. Chronic Myelogenous Leukemia • 1.6 cases per 100,000 • 1.75:1 male:female • BCR-ABL1 translocation (Ph chromosome) • Molecular monitoring of BCR-ABL1 has become standard of care SEER data 2012

  5. Why test? • BCR-ABL1 targeted by small molecules • Prognosis • Monitoring response • Prediction of relapse

  6. Monitoring BCR-ABL1 • Karyotype • Fluorescence in-situ Hybridization • PCR • RT-PCR • Competitive RT-PCR • Nested RT-PCR • qRT-PCR

  7. Sensitivities • Karyotype – 1:20 • FISH – 1-2:200 • RT-PCR – 1:100,000

  8. Response to Therapy Hughes, NEJM, 2003, 349:1423

  9. Prediction of Relapse Press, Clin Cancer Res, 2007, 13:6136

  10. NCCN Guidelines • From NCCN 2.2013, qPCR should be performed: • At diagnosis • Every 3 months when responding to therapy. After CCyR, every 3 months for 3 years, then every 3-6 months • For rising transcript levels (1 log) with MMR, repeat in 1-3 months

  11. The IRIS Trial • The International Randomized Study of Interferon versus STI571 • Major Molecular Response = 3 log decrease from average level at diagnosis • Three laboratories harmonized results (Australia, UK, and USA)

  12. BCR-ABL1 in Practice • Many labs test for BCR-ABL1 • Testing and reporting are variable • Many report change from diagnostic (local) baseline, but not calibrated the same • Results not generally comparable between labs

  13. Why is testing variable? • Sample volume and integrity • Processing and extraction • Reverse transcription • Control gene (BCR, ABL, GUSB, G6PDH, β2M) • Primers • Quantification standards • Instrument and analysis

  14. Harmonizing Results • Median measurement of 30 shared baseline samples established as 100% • 3 log decrease from baseline (0.1%) is Major Molecular Response • For example, if median ratio BCR-ABL/BCR is 0.75, MMR = 0.00075 • International Scale defined by these two points

  15. Getting to the IS • 3 IRIS trial labs use IS by definition (Adelaide, Seattle, London) • IRIS trial samples used to define 100% IS are not available to individual labs • Sample exchange with a laboratory calibrated to the IS?

  16. Pre-WHO Standard • Branford, et al., Blood, 112:3330, 2008, “Desirable performance characteristics for BCR-ABL measurement on an international reporting scale to allow consistent interpretation of individual patient response and comparison of response rates between clinical trials,” • Exchange of patient samples with IS calibrated reference lab to determine Conversion Factor • Requires many samples and periodic re-assessment • Very burdensome for few reference labs

  17. Calibration Pre-WHO Standard Branford, Blood, 2008, 112:3330

  18. What does an IS do? • An International Scale helps standardize quantification • An IS does not improve pre-analytic issues • An IS does not reduce inherent assay variability • An IS does improve reporting

  19. BCR-ABL1 WHO Standard • 1st WHO International Genetic Reference Panel for quantitation of BCR-ABL translocation by RQ-PCR • Intended for manufacturers/labs to develop secondary standards • 4 ampules of freeze dried cells • K562 cells (e14a2) diluted to different ratios in HL60 cells • Spanning relevant range of %IS values

  20. WHO BCR-ABL1 Standard

  21. Secondary Standards • Asuragen – ARQ IS Calibrators and BCR\ABL1 Quant • Ipsogen – BCR-ABL Mbcr-IS assay

  22. Asuragen ARQ IS Calibrator Panel http://www.asuragen.com/Diagnostics/US/Products/qRT-PCR_Oncology/ARQ_IS_CAL/arq_is_cal.aspx

  23. Asuragen Calibrators http://www.asuragen.com/Diagnostics/US/Products/qRT-PCR_Oncology/BCR_ABL1/

  24. Ipsogen Calibrator • Single point calibrator • Tied to WHO Standard • Plasmids • Set to approximate 0.1% IS (MMR cutoff) • Specific for Ipsogen assay

  25. Correlation of two IS based assays

  26. Comparison to Predicate Test

  27. Comparison to Predicate • Very similar results for both calibrated to the IS • IS tests are 0.4 and 0.35 lower than LDT • Very important to communicate this to clinicians when changing reports • Alternatively, report old result and IS

  28. Advantages of Secondary Standards • Allow calibration to WHO standard • Simpler than exchanging many samples with reference lab • Manufacturer’s QC and lot-to-lot control • Periodic checks for drift are more feasible

  29. Control Genes and Quantification • ABL1 is the most commonly used control gene • BCR was used as control gene for IRIS trial labs • GUSB used by some • EAC found ABL1, GUSB, and B2M suitable (Beillard, Leukemia, 2003, 17:2474) • BCR was not reported in study by EAC

  30. ABL1 as Control

  31. BCR as Control

  32. Other genes as Controls • Do not participate in the translocation • Should be expressed constitutively and not vastly over/under expressed compared to transgene • GUSB and B2M

  33. Limit of Detection • Depends on transgene copies and control gene level • What is minimum control gene for acceptable sensitivity? • Calculate minimum control gene level to detect MMR?

  34. “Complete” Molecular Response CMR = no detectable BCR-ABL transcripts Press et al, Clin Cancer Res 13, 6136 (2007)

  35. Practical LOD Issues • Balance sensitivity with not rejecting too many specimens • For LOD=5 copies BCR-ABL, need 50,000 copies of control gene to get 4.0 log • For 5.0 log need 500,000 copies control gene • LOD is affected by total RNA input (5 copies in isolation easier to detect than 5 copies with background nucleic acid)

  36. Summary • The WHO Standard and development of IS have better standardized testing and reporting of BCR-ABL1 • Development of assays/calibrators tied to the IS make transition much simpler • Are more sensitive assays needed?

  37. Thank You • Ruan Ramjit, Kaiser Permanente • Karen Mann, Emory • International BCR-ABL Standardization Group • Emmanuel Labourier, Asuragen • Mary Christopher, Ipsogen