1 / 21

VALIDATION OF MOIST HEAT STEILIZATION

2012-05-20. Moist Heat Sterilization. 2. Definition. ? Sterilization

zhen
Download Presentation

VALIDATION OF MOIST HEAT STEILIZATION

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


    1. 2012-05-20 Moist Heat Sterilization 1 VALIDATION OF MOIST HEAT STEILIZATION JM Tech. Do-Young Ahn

    2. 2012-05-20 Moist Heat Sterilization 2 Definition ? Sterilization “The act or process, physical or chemical, that destroys or eliminates all viable microbes including resistant bacterial spores from a fluid or a solid.” Examples of sterilization methods are : steam treatment at 121?, dry heat at 230?, flushing with a sterilizing solution such as hydrogen peroxide (H2O2) or ozone (O3), irradiation, and filtration. ?Sterility “The reduction of anticipated levels of contamination in a load to the point where the probability of survival is less than 10-6.”

    3. 2012-05-20 Moist Heat Sterilization 3 Definition ?D-value The time in minutes required for a one-log or 90% reduction of a specific microbial population under specified lethal conditions. For steam sterilization it is determined at a constant temperature ?z-value The number of degree of temperature change necessary to change the D-value by a factor of 10. Z-value measures the rate of change in the D-value as a function of temperature. Z=(T2-T1)/Log(D2/D1), D?? 1/10? ?????? ???? ????? Z-value measures the rate of change in the D-value as a function of temperature. Z=(T2-T1)/Log(D2/D1), D?? 1/10? ?????? ???? ?????

    4. 2012-05-20 Moist Heat Sterilization 4 Definition ?F value(lethal rate, instantaneous Fo) The F value is a measurement of sterilization effectiveness. F(T,z) is defined as the equivalent time at temperature T delivered to a container or unit of product for the purpose of sterilization, calculated using a specific value of z. ?Fo value(accumulated Fo) The term "Fo " is defined as the number of equivalent minutes of steam sterilization at temperature 121.1°C delivered to a container or unit of product calculated using a z-value of 10°C. Fo = ? 10^((121-T)/z)*?t

    5. 2012-05-20 Moist Heat Sterilization 5 Methodology ?Overkill Sterilization ?Provides a minimum 12 log reduction of a resistant BI w/ a known D-value of not less than 1 minute. ?Required minimal information on the bioburden ?Bioburden/Bioindicator Sterilization ?Provides a probability of survival of less than 1 in 106 for the bioburden as demonstrated using a resistant BI w/ a known D-value. ?BI may not be inactivated ?Requires information on the numbers and heat resistance of the BI. ?Requires ongoing monitoring or control over bioburden.

    6. 2012-05-20 Moist Heat Sterilization 6 Methodology ?Bioburden Sterilization ?Provides a probability of survival of less than 1 in 106 for the most resistance bioburden expected in the load. ?Requires information on the numbers and heat resistance of the BI. ?Requires ongoing monitoring or control over bioburden.

    7. 2012-05-20 Moist Heat Sterilization 7 Sterilizer Cycle ?Gravity Displacement ?Difference of density Density of air at 20? = 1.2 g/l Density of steam at 100? = 0.6 g/l ?Effectiveness of air elimination depends on the rate of steam supply Air pocket : too rapidly Diffusion into the steam : too slowly, more difficult to remove ?Specially designed steam trap permitting the passage of large volume of air

    8. 2012-05-20 Moist Heat Sterilization 8 Sterilizer Cycle ?Prevacuum cycle ?A more effective method ?By means of a mechanical vacuum pump or a steam eductor ?Vacuum as low as 15~20 mmHg, apply for 8~10 min. ?Pulsing cycle ?A series of alternating steam pulses followed by vacuum excursions ?Air-steam mixture ?Terminal sterilization of large volume parenterals ?Air injection required to compensate the great expansion of air or nitrogen in the head space above the liquid ?Well mixed chamber : fan, raining effect by external pump w/ cooling

    9. 2012-05-20 Moist Heat Sterilization 9 Cycle Development ?Consider factors into account ?Nature of the load : porous materials, heat sensitivity of the products ?Type of the sterilizer ?Employed containers and closures ?Heat stable product : overkill approach ?Heat liable product : bioburden approach Bioburden studies : number of microorganisms D-value studies : only highly resistant spore formers, BIER(biological indicator evaluator resistometer) Inoculate the spore into the actual solutions For solid materials, precut strips

    10. 2012-05-20 Moist Heat Sterilization 10 Preparing for Validation ?Temperature sensing devices : ? T type thermocouples(copper-constantan) encased in flexible sheaths ? Premium grades of wire having ?0.1? accuracy ?Temperature standards ? RTD traceable to the National Bureau of Standards , IPR, HTR ?Calibration of thermocouples ?At two temperatures : 0 ?, 130 ? ?Correction factors ?Stability : ?0.03? ?Accuracy : ?0.5?

    11. 2012-05-20 Moist Heat Sterilization 11 Preparing for Validation ?Autoclave ?Validation nozzle and adaptor ?Data logger : digital output and multi-channel device ?BIs or biological challenges ?Loads

    12. 2012-05-20 Moist Heat Sterilization 12 Validation Protocol Protocol should include ?Objectives of the validation ?Responsibilities of validation personnel and operating department personnel ?Identification and description of the sterilizer and its process control ?Identification of SOPs :equipment ?Calibration of instrument : SOPs and/or description ?Identification and calibration of the temperature monitoring equipment

    13. 2012-05-20 Moist Heat Sterilization 13 Validation Protocol ?A description of the following studies ?Bioburden determination studies(if applicable) ?Empty chamber heat distribution studies ?Container mapping studies(if applicable) ?Loaded chamber heat penetration studies ?Microbiological challenge studies ?Evaluation of drug product cooling water(if applicable) ?Integrity testing of vent filter ?Acceptance criteria ?References ?Review and approval

    14. 2012-05-20 Moist Heat Sterilization 14 Heat Distribution Studies ?To demonstrate the temperature uniformity and stability of the sterilizing medium throughout the sterilizer ?Conduct on both the empty and loaded chamber with max. and min. load configurations ?Acceptance criteria : Less than ±1?of the mean temperature ?Conduct 3 runs to obtain consistent results ?Distribution of the thermocouples : geometrical representatives, exhaust drain, adjacent to the control sensor ?At least 10 probes, normally 15~20 probes Temperature uniformity of the heat distribution studies may be influenced by Type Size Design Installation of the sterilizerTemperature uniformity of the heat distribution studies may be influenced by Type Size Design Installation of the sterilizer

    15. 2012-05-20 Moist Heat Sterilization 15 Heat Distribution Studies ?At loaded chamber heat distribution test, the thermocouples should be positioned in the same locations used for empty chamber heat distribution ?Avoid contacting solid surfaces ?Do not place within any containers ?Data should be obtained at regular intervals

    16. 2012-05-20 Moist Heat Sterilization 16 Container Mapping ?To determine the coolest point within the liquid filled container ?Temperature mapping should be conducted on all the different container types, sizes and fill volume to be validated ?The number of the thermocouples used depends on the container volume ?Possible to use a single thermocouple at different positions, and can be conducted in a smaller autoclave or retort ?Penetration thermocouples should be positioned at the cold spot having lowest temperature or Fo Container mapping studies can be conducted in a small autoclave or retort. The temperature profile of the container should remain constant among different sterilizing chamber.Container mapping studies can be conducted in a small autoclave or retort. The temperature profile of the container should remain constant among different sterilizing chamber.

    17. 2012-05-20 Moist Heat Sterilization 17 Heat Penetration Studies ?To determine the coolest point(s) within the specified load and configuration, and to assure that these points be consistently exposed to sufficient heat lethality ?Prior to conduct heat penetration studies, determine max. and min. load configurations ?Probed container at the cold spot should be distributed uniformly throughout the load ?Penetration thermocouple are positioned at points within the process equipment suspected to be the most difficult for steam heat penetration

    18. 2012-05-20 Moist Heat Sterilization 18 Heat Penetration Studies ?Lethal rate can be determined from the temperature data by the following formula : L = log-1(To-Tb)/z = 10^((To-Tb)/z) ?A summation over time of the lethal rate at a series of temperature(accumulated lethality) Fo = ? 10^((121-T)/z)*?t ?Regard to product stability

    19. 2012-05-20 Moist Heat Sterilization 19 Microbial Challenge Studies ?Biological challenges are employed during heat penetration studies in order to demonstrate the degree of process lethality provided by the sterilization cycle ?Microorganism frequently utilized ?Overkill : Bacillus stearothermophilus and Clostridium sporogenes ?Bioburden : Calibrated BIs from environmental and process isolates such as E. coli ?Type of BI : Spore strips or spore suspension into the suspending medium ?Microbial challenge studies are conducted concurrently with the heat penetration studies

    20. 2012-05-20 Moist Heat Sterilization 20 Validation Report ?Common elements of all reports : ?Identification of the task report by number ?Reference to protocol ?A brief summary of the range of operational conditions experienced and how they were controlled ?A procedure for maintaining control within the approved range ?A summary and analysis of the experimental results ?A brief description of any deviation ?Conclusion ?Review and approval ?Cycle development reports are not usually a part of the validation report

    21. 2012-05-20 Moist Heat Sterilization 21 Maintenance of Validation ?A routine calibration program for all instruments critical to the operation of the sterilizer and its support system ?A preventative maintenance program including periodic operational rechecks and comparison to OQ record ?Routine monitoring of bioburden and periodic BI challenges(optionally) ?Operating records and equipment logs ?Process and equipment change control procedures including review to establish whether additional validations are required ?On-going validation

    22. 2012-05-20 Moist Heat Sterilization 22 Controversial Issues ?Incubation of the sterility test : 7 days vs. 14 days ?USP provide information concerning critical parameters for Parameteric Release ?Reduction extent and frequency of revalidation ?Verification of D-value of BIs ?Use of alternative to B. stearothermophilus as a BI

More Related