An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Gas Analyzer Theory of Operation Gaseous Pollutants mini-course TAMS Center February 2009

  2. Course materials can be found at: http://itep68.itep.nau.edu/itep_downloads/ in the gaseouspollutants folder

  3. Ozone Analyzer Theory of Operation

  4. The ozone layer in the stratosphere absorbs the vast majority of ultraviolet light entering Earth’s atmosphere. We definitely need ozone in the stratosphere. UV

  5. Ground level ozone, however, is known to have significant adverse impacts on the health of humans and ecosystems. Ground level ozone is what we monitor. Ozone isn’t typically emitted directly; it is formed by atmospheric chemical reactions in the presence of sunlight (photochemical reactions). Major precursors are reactive oxides of nitrogen and volatile organic compounds.

  6. Ozone analyzers are photometers. They use the fact that ozone absorbs UV light to measure ozone. By the way, don’t smoke around any gas analyzer

  7. Ozone is most efficient absorbing UV light near the 254 nm wavelength. Sample gas fills the absorption tube (measurement cell) and ultraviolet light at near 254 nm is sent through the sample gas; absorbance is measured. Illustration is from Teledyne API 400E Operator’s manual One problem – other gases absorb at 254 nm.

  8. Solution: Add a reference cycle to the sample cycle - use a manganese dioxide scrubber to selectively remove ozone from the sample gas, measure absorbance due to interferants, and subtract interferants from sample cycle. Illustration is from Teledyne API 400E Operator’s manual

  9. The above text is from Teledyne API 400E Operator’s manual

  10. The above text is from Teledyne API 400E Operator’s manual Manganese dioxide scrubbers can adsorb interferants and rapid humidity changes can cause release of interferants impacting measurement . Ozone conditioning Nitrogen oxide titration

  11. A very important chemical reaction: Notes: NOx = NO + NO2 NOy = NOx + all other reactive oxides of nitrogen NO reacts with ozone driving ozone concentrations downNO2 reacts with VOCs in sunlight and produces ozone

  12. Ozone, NOx, NO2 and NO data, Quapaw Nation

  13. Cherokee Nation Tahlequah Site 7/19/06 through 7/21/06

  14. The 5th and 6th is Saturday, Sunday

  15. Why does a remote location have higher 24-hour averages? It has few NO sources to titrate ozone down at night. Remote areas typically have a much less pronounced diurnal pattern (daytime and nighttime concentrations are much closer).

  16. Wind Direction Power Plant Over 2 miles Monitoring Site

  17. Wind Direction Ozone titration downwind of NO source Ozone Concentrations NO2 + VOCozone produced Power Plant

  18. NOx Analyzer Theory of Operation:Chemiluminescence

  19. Sample gas w/NO O3 Light leaks can impact measurement PMTs are sensitive and can be damaged by light Illustration is from Teledyne API 200E Operator’s manual

  20. Illustration is from Teledyne API 200E Operator’s manual

  21. Illustration is from Teledyne API 200E Operator’s manual

  22. Illustration is from Teledyne API 200E Operator’s manual

  23. NOx Cycle goes through converter Only NO reacts with ozone Dark Cycle bypasses RX Cell MolyConMolybdenum Converter NO2 Converter NO NO Sample Air NOx = NO + NO2 PMT Rx Cell NOx NO NO2 NO2 converted to NO Valves NO Excess Ozone NOCycle NO Cycle passes sample air to reaction chamber Ozone reacts with NO Only NO reacts – NO2 isn’t measuredNO2 is calculated: NOx – NO = NO2 Ozone generator Dry filtered air hv in reaction cell is measured by PMT

  24. SO2 Analyzer Theory of Operation:Fluorescence

  25. (UV Light) Dark period every 30 minutes for PMT offset (API) Illustration is from Teledyne API 200E Operator’s manual

  26. Illustration is from Teledyne API 200E Operator’s manual

  27. CO Analyzer Theory of Operation:Gas Correlation

  28. Illustration is from Teledyne API 360U Operator’s manual