An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. A3DULTRASOUND-BASED TRACKING SYSTEM FOR PROSTATE BIOPSY DISTRIBUTION QUALITYINSURANCE AND GUIDANCE. PhD Thesis Michael Baumann Supervisors Jocelyne Troccaz Vincent Daanen

  2. Context of this thesis Outline • TIMC laboratory • specializing in computer-assisted medical interventions for more than twenty years now • many clinical and industrial collaborations • Pitié-Salpétrière hospital, urology department • active support of this work and very inspiring exchanges • clinical data acquisition on more than 70 patients now • Koelis SA • industrial partner • objective: commercialize products based on prostate tracking ¯ Introduction ●○○○ Prostate Tissue Tracking ○○ Registration Framework ○○○○○○○ Experiments and Results ○○○ Discussion ○ Conclusion/Applications ○○

  3. Bladder Prostate Rectum Urethra Seminal Vesicles Introduction Outline • Prostate • Prostate Cancer • most frequent cancer in men • ~220.000 new cases in US (2007) • ~345.000 new cases in EU25 (2006) • second cause of cancer death for men • 27.000 deaths in US (2007) • 87.400 deaths in EU25 (2006) • slow growing disease • affects mostly elder men (>50 years) ¯ Introduction ●○○○ Prostate Tissue Tracking ○○ Registration Framework ○○○○○○○ Experiments and Results ○○○ Discussion ○ Conclusion/Applications ○○

  4. DRE Introduction Outline • Prostate Specific Antigene (PSA) screening • biological tumor marker • sensitivity for 4ng/ml threshold: 68-83% (clinically significant cancer) • specificity: ~30%  false positives! • Digital Rectal Exams (DRE) • highly varying sensitivity in clinical studies reported: 18% to 68% • specificity: 4% to 33% • complementary to PSA screening • Prostate Biopsies • Sensitivity: 60-80 % (clinically significant cancer) • Specificity: >95% (histological analysis) • invasive  programmed only if DRE/PSA positive • dilemma: false negatives  repeated biopsies ¯ Introduction ●○○○ Prostate Tissue Tracking ○○ Registration Framework ○○○○○○○ Experiments and Results ○○○ Discussion ○ Conclusion/Applications ○○

  5. 2D TRUS probe with needle guide corresponding 2D US image with needle trajectory longitudinal cut Introduction Outline • Prostate Biopsies • 2D transrectal ultrasound (TRUS) control • needle guide on probe • guide aligned with longitudinal plane of probe ¯ Introduction ●○○○ Prostate Tissue Tracking ○○ Registration Framework ○○○○○○○ Experiments and Results ○○○ Discussion ○ Conclusion/Applications ○○

  6. coronal plane Introduction Outline • Biopsy targets • prostate cancer is isoechogenic • systematic targets • McNeal’s 3-zone model: central zone (CZ), transition zone (TZ), peripheral zone (PZ) • 68% of cancer can be found in peripheral zone • Systematic 12-core protocol • clinical representation in (pseudo-)coronal plane ¯ Introduction ●○○○ Prostate Tissue Tracking ○○ Registration Framework ○○○○○○○ Experiments and Results ○○○ Discussion ○ Conclusion/Applications ○○ coronal plane

  7. Prostate Motion Problem Outline • Prostate motion • main challenge for any prostate tissue tracking system • displacements and deformations • Transrectal biopsy specific: probe-related motion • end-fire probe • deformations and displacements due to probe pressure • Neighboring organs (diaphragm motion, rectal and bladder filling) • minor impact during prostate biopsies ¯ Introduction ●●○○ Prostate Tissue Tracking ○○ Registration Framework ○○○○○○○ Experiments and Results ○○○ Discussion ○ Conclusion/Applications ○○

  8. Prostate Motion Problem Outline • Patient motion • (small) deformations • displacements with respect to surrounding tissues • displacements with respect to operating room (pelvis movements!) ¯ Introduction ●●○○ Prostate Tissue Tracking ○○ Registration Framework ○○○○○○○ Experiments and Results ○○○ Discussion ○ Conclusion/Applications ○○

  9. Introduction Outline • Biopsy and Target Localization Problem • only rudimentary knowledge about biopsy position • at all stages of intervention! • Pre-interventional stage/planning • n-core protocol target definition highly approximate • targets have to be mentally mapped into patient anatomy • During intervention: target localization problem • difficult to aim invisible target under 2D control • ultrasound: few structural information • 2D: no depth information • prostate motion ¯ Introduction ●●●○ Prostate Tissue Tracking ○○ Registration Framework ○○○○○○○ Experiments and Results ○○○ Discussion ○ Conclusion/Applications ○○ finding the target : what do we aim exactly?

  10. Introduction Outline • Target localization problem (ctd) • there exist better targets than systematic protocol • high quality cancer distribution atlas available [Shen’01] • simulations: biopsy sensitivity > 96% with only 6 needles (transperineal access) • suspicious lesions identified on IRM • repeated biopsy series • avoid already sampled tissues (negative targets) • how to aim these targets? • After intervention : sample localization problem • where were the samples taken exactly? • quality control? • are there unsampled regions? • difficult to map histological cancer information back to anatomy • difficult to use histological information for focal treatment planning ¯ Introduction ●●●○ Prostate Tissue Tracking ○○ Registration Framework ○○○○○○○ Experiments and Results ○○○ Discussion ○ Conclusion/Applications ○○

  11. Existing Solutions Outline • Magnet Resonance imaging-based approaches • objective : target suspicious lesions detected on MR images • biopsy under MRI control • instruments calibrated with MR frame • Beyersdorff [05], Musil, Krieger et al. [04,05,07], Stoianovici [07] • IRM compatible biopsy acquisition instruments/robot • pro: possibility to aim IRM targets • con: cannot detect/compensate patient movements • would require high resolution, real-time MRI • con: diagnosis: cost-benefit ratio unsatisfying • several millions of biopsies/year in US and EU ¯ Introduction ●●●● Prostate Tissue Tracking ○○ Registration Framework ○○○○○○○ Experiments and Results ○○○ Discussion ○ Conclusion/Applications ○○

  12. Existing Solutions Outline • Probe tracking + registration based approaches • 2D transrectal ultrasound • track US probe with optical or magnetic tracking system • identifies view cone motion • register 2D tracking images with free-hand reference volume • identifies prostate motion • Xu et al. [07]: Magnetic probe tracking + registration • pro: can compensate smaller rigid prostate-movements • con: free-hand volume with end-fire probe  low accuracy • con: rigid registration • con: registration of lateral biopsy images not robust (partial gland problem) • con: difficult to compensate large pelvis movements ¯ Introduction ●●●● Prostate Tissue Tracking ○○ Registration Framework ○○○○○○○ Experiments and Results ○○○ Discussion ○ Conclusion/Applications ○○

  13. Outline of the Presentation Outline • Prostate Tissue Tracking and Guidance • clinical and scientific objectives • soft-tissue tracking • Prostate Image Registration • registration framework • multi-resolution techniques • image distance metric (rigid) • probe movement model • rigid refinement • elastic registration framework • forces for elastic registration • Experiments and Results • registration success rate • accuracy • biopsy maps and targeting accuracy study • Discussion • Conclusion and Potential Applications • clinical and scientific contributions • potential applications ¯ Introduction ●●●● Prostate Tissue Tracking ○○ Registration Framework ○○○○○○○ Experiments and Results ○○○ Discussion ○ Conclusion/Applications ○○

  14. Objectives Outline • Scientific objectives • prostate tissue tracking • establish tissue correspondence • with respect to a reference space • goal: establish correspondence between • biopsy site planning • reference space • needle position during intervention • Clinical objectives • more sophisticated targets • MRI, statistical cancer atlas, unsampled zones when repeating biopsies • guide clinician to target • feed-back to clinician about exact sample position • immediately and after intervention • biopsy maps Introduction ●●● ¯ Prostate Tissue Tracking ●○ Registration Framework ○○○○○○○ Experiments and Results ○○○ Discussion ○ Conclusion/Applications ○○

  15. Image-based Prostate Tracking Framework Outline • Proposed Solution • 3D ultrasound-based • hybrid registration • image-based • a priori model based • deformation estimation • no probe tracking • miniminal overhead for clinician, no segmentation Introduction ●●● ¯ Prostate Tissue Tracking ●● Registration Framework ○○○○○○○ 3D ultrasound view cone Experiments and Results ○○○ Discussion ○ Conclusion/Applications ○○ anchor volume tracking volume tracking volume:  acquired during intervention:  “contains” sample trajectory anchor volume:  acquired before intervention  defines the reference space needle projection:  projection into anchor volume  projection can lead to curbed trajectories biopsy map:  contains projections of all samples guidance:  target projection into tracking volume registration: establishment of correspondences for identical tissues present in both images

  16. Registration Framework Outline • Image Registration • Optimization (minimization) problem • φ = transformation model • T = template/transformed image (R3  R) • R = reference/fixed image • D[.] = cost functional • Problems • registration only efficient with local minimization (downhill search) • successful local minimization requires • locally unimodal cost functional • start point inside the convex region • the more degrees of freedom (DOF) of φ, the more difficult to find unimodal region of D[R,T,φ] ! Introduction ●●● Prostate Tissue Tracking ●● ¯ Registration Framework ●○○○○○○ Experiments and Results ○○○ Discussion ○ Conclusion/Applications ○○

  17. Registration Framework Outline • Proposed approach Introduction ●●● 3 DOF 6 DOF ~ 125.000 DOF Probe kinematics based rigid presearch Refinement of rigid estimate Prostate Tissue Tracking ●● Elastic estimation ¯ Registration Framework ●○○○○○○ optimization techniques Experiments and Results ○○○ variational optimization parametric local optimization parametric systematic search Discussion ○ multi-resolution approaches loss-containing multi-resolution techniques Conclusion/Applications ○○ voxel intensity based image distance metrics SSD with local intensity shift multivariate correlation coefficient a priori models linear elasticity bio-mechanical probe insertion endorectal probe kinematics inverse consistency

  18. Multi-Resolution probe kinematics elastic rigid refinement Outline • Multi-resolution approach • Gaussian pyramid • registration performed on different resolution levels • Coarse resolutions and information loss Introduction ●●● Prostate Tissue Tracking ●● ¯ Registration Framework ●●○○○○○ Experiments and Results ○○○ Discussion ○ Conclusion/Applications ○○ level n level n+1 US-specific: complex image masks problematic when computing level n+1 from level n ≤ 50% of fine-grid voxel  mask coarse-grid voxel else  use average of available voxel attention: introduces, however, local information shifts

  19. Multi-resolution probe kinematics elastic rigid refinement Outline • 50-percent rule • use it for pyramid construction • for interpolation • for every other computation on multiple voxels • gradient computation (image distance metrics!) • Gaussian smoothing • Conclusion • Makes high-speed volume to volume registration possible • reliable registration on very coarse levels • Disadvantage • introduces small local information shifts Introduction ●●● Prostate Tissue Tracking ●● 50% rule level 5 standard level 5 level 1 ¯ Registration Framework ●●○○○○○ Experiments and Results ○○○ Discussion ○ standard level 5 50% rule level 5 Conclusion/Applications ○○ level 1

  20. Distance metric (rigid) probe kinematics elastic rigid refinement Outline • Image distance metric (Rigid Registration) • correlation coefficient (CC) based • well-proven for monomodal registration • multivariate application • intensity image + gradient magnitude image • more robust results on coarse levels Introduction ●●● Prostate Tissue Tracking ●● ¯ Registration Framework ●●●○○○○ Experiments and Results ○○○ Discussion ○ Conclusion/Applications ○○ raw image gradient magnitude

  21. Probe kinematics probe kinematics elastic rigid refinement Outline • Challenge • probe used to guide needle  view cone motion • adds up to prostate motion • motion too large for capture range of image distance metric • direct downhill/local registration only ~30-40% success rate • Observations • probe head always in contact with rectal wall in front of prostate • if not, no prostate image or needle trajectory outside prostate • anal sphincter heavily constrains probe motion • fix point for probe motion • most important rotations occur around probe axis (when switching lobe) Introduction ●●● Prostate Tissue Tracking ●● ¯ Registration Framework ●●●●○○○ Experiments and Results ○○○ Discussion ○ Conclusion/Applications ○○

  22. Probe kinematics probe kinematics elastic rigid refinement Outline • Model of endorectal probe kinematics: • approximate prostate capsule with ellipsoid from bounding box • estimate rectal probe fix point • admit only positions for which • the probe axis lies on the fix point • the probe origin lies on the membrane • 3 degrees of freedom only • can be exhaustively explored in reasonable time! • Advantages • Makes solution independent of external tracking system! • Solves patient motion problem! Introduction ●●● Prostate Tissue Tracking ●● ¯ Registration Framework ●●●●○○○ Experiments and Results ○○○ Discussion ○ Conclusion/Applications ○○

  23. Rigid Refinement probe kinematics elastic rigid refinement Outline • Refinement of rigid estimate Introduction ●●● rigid registration of 5 best transformations provided by probe kinematics Prostate Tissue Tracking ●● high quality registration of best result ¯ Registration Framework ●●●●●○○ Experiments and Results ○○○ classical local/downhill search algorithm: Powell-Brent Discussion ○ Conclusion/Applications ○○ high quality local search: from coarse to fine high speed: optimize on coarsest level

  24. Elastic Registration probe kinematics elastic rigid refinement Outline • Prostate deformations • relatively small (several millimeters) • strongest near probe head • difficult to estimate: • few image information near probe head • Transformation model: displacement field • Framework • : linear elastic potential • regularizes/smoothes displacement field • minimal when no deformation  strong regularizer • : SSD variant to measure image distance • : bio-mechanical simulation of probe insertion • : inverse consistency constraints Introduction ●●● Prostate Tissue Tracking ●● ¯ Registration Framework ●●●●●●○ Experiments and Results ○○○ Discussion ○ Conclusion/Applications ○○

  25. Elastic Registration probe kinematics elastic rigid refinement Outline • Elastic regularization: solution scheme • variational approach • necessary condition for solver u* of cost function: • Gâteaux derivative at u* vanishes for all perturbations ψ • Euler-Lagrange equations for linear elastic regularization: • trick: separate force computation and regularization • accumulate forces • solve Euler-Lagrange equations • then we get an elliptic boundary value problem of the form • trick: introduce artificial time to obtain iterative gradient descent scheme Introduction ●●● Prostate Tissue Tracking ●● ¯ Registration Framework ●●●●●●○ gradients of distance metrics gradient of linear elastic potential Experiments and Results ○○○ Discussion ○ Conclusion/Applications ○○

  26. Force terms probe kinematics elastic rigid refinement Outline • Image based forces • correlation coefficient: statistically not robust when locally computed • SSD • assumes identity between R and Tû • does not correspond to reality! • changes in ultrasound gain, probe pressure and ultrasound direction • Local intensity shift model • additive model: • b estimated with Gaussian convolutions on R and T • resulting force term Introduction ●●● Prostate Tissue Tracking ●● ¯ Registration Framework ●●●●●●● Experiments and Results ○○○ Discussion ○ Conclusion/Applications ○○

  27. Force terms probe kinematics elastic rigid refinement Outline • Bio-mechanical probe insertion model • model of probe-related tissue displacements • Interpret displacement differences as forces in the estimation process Introduction ●●● Prostate Tissue Tracking ●● ¯ Registration Framework ●●●●●●● Experiments and Results ○○○ Discussion ○ Conclusion/Applications ○○

  28. Force terms probe kinematics elastic rigid refinement Outline • Inverse consistency forces • Observation: forward and backward estimation u and v not symmetric: • Zhang’s approach [’06] • estimate u and v simultaneously • enforce inverse consistency by minimizing • alternating optimization process: • resulting force term Introduction ●●● Prostate Tissue Tracking ●● ¯ Registration Framework ●●●●●●● Experiments and Results ○○○ Discussion ○ Conclusion/Applications ○○

  29. Registration Framework Outline • Proposed approach Introduction ●●● Probe kinematics based rigid presearch Refinement of rigid estimate Prostate Tissue Tracking ●● Elastic estimation ¯ Registration Framework ●●●●●●● Experiments and Results ○○○ variational optimization parametric local optimization parametric systematic search Discussion ○ loss-containing multi-resolution techniques Conclusion/Applications ○○ SSD with local intensity shift multivariate correlation coefficient linear elasticity bio-mechanical probe insertion endorectal probe kinematics inverse consistency

  30. Experiments and results

  31. Experiments and Results Outline • Experiments • on real patient data • Pitie-Salpétrière Hospital, Paris, urology department • P. Mozer, G. Chevreau, S. Bart, J.-C. Bousquet • 3D ultrasound images (GE Voluson, RIC5-9 probe) • acquired before biopsies and after each sample acquisition • targeting carried out under 2D US control • Registration example Introduction ●●● Prostate Biopsy Tracking ●● Registration Framework ●●●●●●● ¯ Experiments and Results ●○○ Discussion ○ Conclusion/Applications ○○

  32. Experiments and Results Outline • Rigid Registration • Algorithm tested on 785 image pairs from 47 patients • 27 mis-registrations (success-rate 96.5 %) • Conclusion • probe movement model works fine! Introduction ●●● Prostate Biopsy Tracking ●● Registration Framework ●●●●●●● ¯ Experiments and Results ●○○ Discussion ○ Conclusion/Applications ○○ ultrasound depth ultrasound quality partial contact

  33. Experiments and Results Outline • Accuracy study • 208 registrations on data from 14 patients • manual point fiducial segmentation (calcifications, dark spots) • error computed on Euclidean distances of corresponding fiducials • Registration accuracy • rigid optimization performed on resolution levels 5 to 3 • elastic optimization performed on resolution levels 6 to 3 • Conclusion • accuracy sufficient for many clinical applications Introduction ●●● Prostate Biopsy Tracking ●● Registration Framework ●●●●●●● ¯ Experiments and Results ●●○ Discussion ○ Conclusion/Applications ○○

  34. Experiments and Results Outline • First Application: Biopsy maps • show targeting difficulties • P. Mozer, M. Baumann, G. Chevreau, A. Moreau-Gaudry [Mozer’08] • apex and base targets more difficult to reach than central gland • operator learning curve measured Introduction ●●● Prostate Biopsy Tracking ●● Registration Framework ●●●●●●● ¯ Experiments and Results ●●● Discussion ○ Conclusion/Applications ○○

  35. Discussion

  36. Discussion Outline • Automatic registration validation • visual validation time-consuming and operator-dependent • open issue: automatic detection of failures! • necessary for guidance! • Registration and real-time • requires 5 – 15 seconds • stream parallelization: • algorithm mainly consists of image convolutions • can be parallelized on a voxel per voxel basis • well suited for latest graphic card architectures (stream processors) • registration times of 1 second or less should be feasible • Similarity measures • Good performance for intra-series registration • Still to be evaluated for inter-series registration • only one patient with two biopsy series for instance • Intensity shift model • depends strongly on parameter σof Gaussian convolution Introduction ●●● Prostate Biopsy Tracking ●● Registration Framework ●●●●●●● Experiments and Results ●●● ¯ Discussion ● Conclusion/Applications ○○

  37. Discussion Outline • Probe movement model (rigid registration) • very good success rate • no probe tracking necessary • less hardware in OR! Simpler workflow and logistics! • improvements with model to data fitting possible • should further improve success rate • Bio-mechanical probe insertion model (elastic registration) • for about 50% image pairs, the model improves elastic registration • but: sometimes inadequate model of reality Introduction ●●● Prostate Biopsy Tracking ●● Registration Framework ●●●●●●● Experiments and Results ●●● ¯ Discussion ● Conclusion/Applications ○○

  38. Discussion Outline • Clinical acceptability • only slight modification of classical acquisition protocol • bounding box placement • registration validation (probably post-op step) • no additional instruments/hardware in operation room • cost effective: cost similar to current procedure Introduction ●●● Prostate Biopsy Tracking ●● Registration Framework ●●●●●●● Experiments and Results ●●● ¯ Discussion ● Conclusion/Applications ○○

  39. Conclusion and Applications

  40. Conclusion Outline • Scientific contributions • probe movement model • robust • no probe tracking hardware required • completely solves patient movement problem • reusable for many endocavitary US interventions! • loss-containing multi-resolution filtering and interpolation • robust optimization on very sparse resolution levels • hybrid model- and image-based elastic deformation framework • novel voxel similarity measure for elastic registration • remarkably robust • simple • proof of concept on large set of patient data • Medical contributions • biopsy accuracy study on biopsy maps • more difficult to reach apex/base than mid-gland • operator learning curve proven Introduction ●●● Prostate Biopsy Tracking ●● Registration Framework ●●●●●●● Experiments and Results ●●● Discussion ● ¯ Conclusion/Application ●○

  41. Future work/Prospects Outline • Potential Applications : Biopsies • biopsy maps • immediate feed-back, post-interventional quality control • cancer maps • map histological results on 3D biopsy map • guidance • assist clinician during targeting • requires automatic registration validation and real-time registration • guidance  MRI target mapping • reach MRI targets under ultrasound control • requires • MRI to ultrasound registration • guidance  repeated biopsy series • avoid multiple sampling • visualize already sampled tissues • guidance  cancer atlas targets • define targets with cancer probability atlas (Shen’01) • map them onto anchor volume • requires atlas to ultrasound volume registration Introduction ●●● Prostate Biopsy Tracking ●● Registration Framework ●●●●●●● Experiments and Results ●●● Discussion ● ¯ Conclusion/Applications ●●

  42. Future work/Prospects Outline • Potential applications : Therapy • improve accuracy of ultrasound-guided therapy • brachytherapy, HIFU, cryotherapy, … • focal therapy? • currently: two unknowns after positive biopsy findings • shape of the tumor • exact location of the biopsy • not accurate enough for focal therapy • we solve 2! • sufficient for focal therapy? • in combination with statistical tumor atlas? Introduction ●●● Prostate Biopsy Tracking ●● Registration Framework ●●●●●●● Experiments and Results ●●● Discussion ● ¯ Conclusion/Applications ●●

  43. Publications and References • Publications • [Baumann’07] M. Baumann, P. Mozer, V. Daanen, J. Troccaz. Towards 3D Ultrasound Image Based Soft Tissue Tracking: a Transrectal Ultrasound Prostate Image Alignment System. MICCAI'07, Brisbane, Australia, 2007. Springer LNCS 4792. • [Mozer’07] P. Mozer, M.Baumann, G. Chevreau, J. Troccaz. “Fusion d’images : application au contrôle de la distribution des biopsies prostatiques,” Progrès en Urologie (les Cahiers de la Formation Continue), vol. 18 (1), 2008 • [Baumann’08] M. Baumann, P. Mozer, V. Daanen, J. Troccaz. “Fast and robust elastic registration of endorectal 3D ultrasound prostate volumes for transrectal prostate needle puncture tracking,”In proceedings of CARS’08, Barcelona, 2008 • References • [Shen’04] D. Shen, Z. Lao, J. Zeng, W. Zhang, I. A. Sesterhenn, L. Sun, J. W. Moul, E. H. Herskovits, G. Fichtinger, and C. Davatzikos. “Optimization of biopsy strategy by a statistical atlas of prostate cancer distribution,” Medical Image Analysis, vol. 8, no. 2, pp. 139–150, 2004. • [Zhang’05] Z. Zhang, Y. Jiang, and H. Tsui. “Consistent multi-modal non-rigid registration based on a variational approach,” Pattern Recognition Letters, pp. 715–725, 2006.

  44. Acknowledgements • Urology department Pitié-Salpétrière • Pierre Mozer, Grégoire Chevreau, Stéphane Bart • Koelis SA • Antoine Leroy • Vincent Daanen • TIMC • GMCAO group • Jocelyne Troccaz • and everyone else who supported this project during the last three years! • Funding: • 2004-06: ”Programme Hospitalier de Recherche Clinique - Prostate-Echo”, French ministry of research • 2005-07: “Surgétique Minimalement Invasive (SMI)”, Agence Nationale de Recherche (ANR) • 2005-08: Association Nationale de la Recherche Technique, bourse CIFRE

  45. Inadequate probe model • possible explanation

  46. Separate elastic estimation • first step: • estimate deformations caused by probe forces • second step: • estimate deformations caused by image forces • start optimization with probe deformation as initial guess

  47. Elastic Regularization probe kinematics elastic rigid refinement Outline • Elastic regularization: solution scheme (ctd) • von Neumann stability analysis of numerical scheme yields • Stability criterion and elasticity parameters • forces in our framework are not physical • derived from distances • how to calibrate them with the elastic forces? • Young’s modulus E has no physical meaning • interpret it as free parameter • control elasticity parameters with Poisson’s coefficient v and ∆t • seek best balance between smoothness and convergence rate • balance elastic smoothing and maximally admitted deformation Introduction ●●● Prostate Tissue Tracking ●● ¯ Registration Framework ●●●●●●○ Experiments and Results ○○○ Discussion ○ Conclusion/Applications ○○ Young’s modulus Poisson’s coefficient

  48. Elastic Registration probe kinematics elastic rigid refinement Outline • Elastic regularization: solution scheme (ctd) • solved with Gauss-Seidel and full multigrid strategy • Boundary conditions • bending side-walls, fixed edges • good model for probe insertion Introduction ●●● Prostate Tissue Tracking ●● ¯ Registration Framework ●●●●●●○ Experiments and Results ○○○ Discussion ○ Conclusion/Applications ○○

  49. Introduction • Biopsy acquisition • patient in dorsal or lateral position • local anesthesia • 12 acquisitions