Magnetic Field Finite-Element Calculations for the Upper Cryostat

Magnetic Field Finite-Element Calculations for the Upper Cryostat. S. Balascuta, R. Alarcon Arizona State University B. Plaster, B. Filippone, R. Schmid Caltech The nEDM Collaboration 11/02/2007. Design Considerations.

vesta
Download Presentation

Magnetic Field Finite-Element Calculations for the Upper Cryostat

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Magnetic Field Finite-Element Calculations for the Upper Cryostat S. Balascuta, R. Alarcon Arizona State University B. Plaster, B. Filippone, R. Schmid Caltech The nEDM Collaboration 11/02/2007

  2. Design Considerations • The magnetic field uniformity inside the collection volume has to be 510-4. • 3He spin relaxation time has to be bigger than 10000 seconds inside the collection volume. The loading and transfer time of 3He from the ABS source to the collection volume is less than 2000 seconds. • 3. The loss in the polarization of 3He from ABS source to measurement cells can be minimized if the field changes adiabatically in any point along the 3He path.

  3. The location of the 3He magnet and collection volume inside the upper cryostat

  4. The geometry of the coils and of the shield inside the upper cryostat. • 24 cos() coils (R=20 cm, L=134 cm) • Transport solenoid (r=5 cm, L=80 cm) • Magnetic shield R=21 cm, L=136 cm, 62 mils thick.

  5. The cos coils The position of each “cos” coil is given by the angle J K>0 y ’J ’J SJ SJ A’ B Coil j’ In the presence of the shield the field uniformity along OZ axis decreases but is still big along the OX and OY axis. Coil j J J A x On circle: A (xj, yj); On Ellipse: A’ (xj1, yj1 ) Coil j Coil j’ Project back on circle A’ to B (xJ2, yJ2 )

  6. The magnetic field distribution inside 24 coils (K) placed inside the shield with two gaps (12 cm wide and 64 cm long) with and without the transport solenoid. Shield: Metglas, 62 mils thick, R=21 cm, L=136 cm long, with two gaps 12 cm wide. Transport solenoid: 80 cm long, center at Xc=51.19, Yc=0 , Zc=47.73 cm

  7. The correction of the asymmetry in the OZ direction by adjusting the length and the center of the gap. Shield: Metglas, 62 mils thick, R=21 cm, L=138 cm long, with two gaps 12 cm wide. 68cm 68cm 42 cm A O C Z Lgap D 32.3cm 35.7cm Y

  8. N=24 + M=8 coils with the same radius (20cm) and length (134 cm) inside shield (R=21 cm, 136 cm long) with two gaps (12 cm wide, 60.6 cm long) Y N=24 coils with current I1=14.7 mA. M=8 coils with current I8 ( I2=I8+I1 ) 3He X 12 cm Transport solenoid (80 cm long, R= 5 cm), with a current Is=45.3 mA. For the N=24 (+M=8) coils, the optimum K= -0.03 with I2=I1+I8 =40 mA. K

  9. The longitudinal (T1) relaxation time for 3He inside the collection volume of the upper cryostat versus temperature. The volume average total gradients are:

  10. Conclusions • For the upper cryostat we have considered a system of N=24 coils (with current I1) and M=8 coils (with current I2) coils ( with radius R=20 cm and length L=134 cm). The M=8 coils overlap some of the N=24 coils. • The magnetic shield (R=21 cm, L=136 cm) has two gaps (12 cm wide, and 60.6 cm long). • In the presence of the transport solenoid (with a current 45.3 mA) the field inside the collection volume is asymmetric along OX and OZ. To eliminate the asymmetry along the OZ axis, the length of the two gaps is decreased to 60.6 cm and the Z coordinate of their center is moved to +2 cm. • For the optimum currents I8=25 mA and I1=14.7 mA through the M=8 coils and N=24 coils (with K=-0.03) the relaxation times T1 and T2 are bigger than 5·104 seconds (if the temperature in the collection volume is T<0.5 K).