1 / 67

NEUSOFT NSM-P035 Super Open

NEUSOFT NSM-P035 Super Open Magnetic Resonance Imaging System. Outline. MRI Introduction principles 、 features 、 clinical applications 、 development trend Features of NEUSOFT MRI NSM-P035 hardware : magnet 、 RF 、 gradient 、 computer

tyrell
Download Presentation

NEUSOFT NSM-P035 Super Open

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. NEUSOFT NSM-P035 Super Open Magnetic Resonance Imaging System

  2. Outline • MRI Introduction • principles、features、clinical applications、development trend • Features of NEUSOFT MRI NSM-P035 • hardware:magnet、RF、gradient、computer • software:imaging technology、technical specifications、sequence、operating interface

  3. Principles of MRI • Nuclear magnetic resonance phenomenon • Hydrogen nucleus —proton

  4. Spin and Magnetic Moment Principles of MRI

  5. Magnetization Principles of MRI • Proton Precession

  6. Resonance • Relaxation Principles of MRI

  7. MR Signal Principles of MRI

  8. Features of MRI • Non-invasive examination • Multi-type of images • High image contrast • Slicing at any orientation • No contrast agent required in MRA

  9. Non-invasive examination • CT、X-ray、Nuclear medicine– ionizing radiation • Radiation energy of MRI is 1/1012 that of X-ray system • No definite trauma has been certified since more than 20 years clinical application • Pregnant woman can be inspected by MRI, not CT • MRI interventional therapeutic procedure,no trauma to doctor and patient Features of MRI

  10. IR T2 weighted T2 weighted Water suppression T1 weighted IR T1 weighted Multi-type of images • To get more diagnosis information by comparing various images • To improve detectable rate and accuracy of clinical diagnosis Features of MRI

  11. High Image Contrast • Water • Fat • Protein • Skeleton • Gas Features of MRI

  12. Slicing at any orientation Features of MRI

  13. No Contract Agent Need in MRA • In-flow and out-flow effect • TOF • PC Features of MRI

  14. MRI Clinical Applications MRI Diagnose Advantage Parts • Nervous system • Spine 、extremity • Muscle、convective tissue • Angiography system • Pelvis • Substantive organs

  15. Coronal FLAIR Coronal T2 • Head • Brain Stem Infarct MRI Clinical Applications

  16. Coronal T1 Arrow T1 • Pituitary Tumor MRI Clinical Applications

  17. Spine and Spinal Cord MRI Clinical Applications

  18. Sagittal T1 Sagittal T2 • Spinal Cord Malacosis MRI Clinical Applications

  19. Axial T1 Sagittal T2 • Lung Cancer Lumber Transfer MRI Clinical Applications

  20. Extremity 、Joint MRI Clinical Applications

  21. Angioma of Liver • Abdomen MRI Clinical Applications

  22. Pelvis MRI Clinical Applications

  23. Breast MRI Clinical Applications

  24. Strongpoint of MRI Imaging • No trauma——meet for patient requirement for health care • More detectable rate——decrease misdiagnosis • More accurate diagnosis --reduce wrong diagnosis • More quantitive information • Increase estimation prognosis accuracy • Assist making better therapy plan • More impersonality evidence——decrease law risk MRI Clinical Applications

  25. MRI Development Trend • MRI Market Status • High field • more than 1.5T 30% • Permanent magnet open 70% • Medium field quits basically from market

  26. Comparison of the Kinds of Magnet MRI Development Trend

  27. Clinical Application Comparison of Permanent and Superconducting Magnet Low field permanent 5% Claustrophobic patient Fat patient Less embeddedmetal Joint motion analysis MRI intervention High field superconducting 5% Imperceptible pathological changes Functional imaging Spectrum analysis 95% Conventional clinical application Majority advanced function MRI Development Trend

  28. Evolvement History of Permanent Open MRI MRI Development Trend

  29. Sparkler 1.5T Superconducting 1.5T NAM-P023A C-type Super Open 0.23T NSM-P035 C-type Super Open 0.35T NEUSOFT MRI Technology Development MRI Development Trend

  30. NSM-P035 Permanent Magnet MRI Most successful magnetic field strength in permanent magnet systems

  31. NSM-P035 facts • SNR is 2.66 times more that 0.2T ,improve greatly image quality Double scan speed of 0.2T, increase scan speed • Increase spatial resolution, 0.2mm slice thickness • Provide firm foundation for future upgrade NSM-P035 Permanent Magnet MRI

  32. Relation between Image SNR and magnet strength NSM-P035 Permanent Magnet MRI

  33. Noise Effect upon Image Quality NSM-P035 Permanent Magnet MRI

  34. Advanced Magnet Technology • Type:permanent magnet • Advanced permanent magnet material --NdFe • Whole integrated foundry • Self-shield technology • Eddy current free technology • Self-constant temperature technology • Insure long stable operation NSM-P035 Permanent Magnet MRI

  35. The most advanced open magnet—C-type 2-column super open • Super open—4 sides open • Decrease patient claustrophobic feeling , expand patient volume • Positioning of any part of patient at the center of magnet ensures the high image quality • Convenient for interventional treraphy NSM-P035 Permanent Magnet MRI

  36. The most convenient design of patient couch • Couch is separate from magnet • 2 D movement with maximum distance 160cm • Accurate laser localization system • Easy system positioning • Space saving NSM-P035 Permanent Magnet MRI

  37. Minimal Installation Area Less Environment Requirement • 30m2 installation area only • Lowest height 3m • Temperature control only need popular air conditioner NSM-P035 Permanent Magnet MRI

  38. Powerful Gradient System • Gradient Amplifier • Maximum gradient strength:16mT/m,ensure high spacial resolution • Rising time:300μs( up to 98%) • Slew rate:52mT/m/ms,increase the scanning speed • air cooling,no additional expenditure • Gradient coils • plane fingerprint design • assure accurate spacial positioning NSM-P035 Permanent Magnet MRI

  39. High Efficient RF System • RF Amplifier • RF power reach to 5KW, which leads to the high signal frequency • RF transmitter coil • Quanture(QD) design • Signal-to-Noise ratio is 40% higher than normal coil • longitudinal resonance signal efficiency is higher than transversal resonance signal efficiency by 40% NSM-P035 Permanent Magnet MRI

  40. Comprehensive Receiving coils • 7 standard configuration: QD head coil QD Neck Coil QD Body Coil NSM-P035 Permanent Magnet MRI QD Extremity Coil Flat Spine Coil Breast Coil

  41. Computer System • High configuration and performance • CPU: Pentium IV 2.4GHz • RAM:1.5GB • HD:120GB--capability of storing 0.9 million images • peripheral storage:2.3GB MO or 640MBCD-R • monitor:21” SONY high rosolution color display monitor resolution:1600×1200 • Windows operating system, easy to learn and use • Abundant in backup source, minimum cost in maintenance and upgrade NSM-P035 Permanent Magnet MRI

  42. Optional Accessories • Mutual intercom system • Hi-Fi stereo sound • Laser Imager NSM-P035 Permanent Magnet MRI

  43. Advanced Imaging Techniques • Multi-slice/Multi-angle • Gating Technology • Partial echo • Flow compensation • Anti-wrap-around artifact elimination • Multi-region pre-saturation • Optimized bandwidth • Optimized data acquisition NSM-P035 Permanent Magnet MRI

  44. Thickness 2mm FOV 50mm Performance Specifications • TE(min): 2ms • TR(min): 6ms • FOV: 26-400mm • Slice thickness(min) 2D: 2mm • Slice thickness(min) 3D: 0.2mm • Number of slices : 64(2D) 128(3D) NSM-P035 Permanent Magnet MRI

  45. Abundant Imaging Sequences • Spin Echo (SE) • Gradient Echo(GE) • IR, FLAIR, STIR • MSME • MRA • FSE • MRU、MRCP) • 3D GE,3D SE • DWI NSM-P035 Permanent Magnet MRI

  46. neck sag. knee joint sag. Spin Echo(SE) • The most common sequence in clinical routine • Clear image • Display anatomic image • Only apply to T1 weighted imaging at present NSM-P035 Permanent Magnet MRI

  47. Liver T2/W Head T2/W Fast Spin Echo (FSE) • fast image acquisition • commonly used for T2 weighted imaging NSM-P035 Permanent Magnet MRI

  48. T1/W T2/W Gradient Echo (GE) • Fast image acquisition • Sensitive to bleeding NSM-P035 Permanent Magnet MRI

  49. T2/W Water suppression IR , FLAIR • Suppress free water • Infarct displayed clear • Judge focal composition NSM-P035 Permanent Magnet MRI

  50. T1/W Fat suppression IR, STIR • Fat suppression signal • Judge focus composition • Other parts displayed much clearer NSM-P035 Permanent Magnet MRI

More Related