Minimizing Beam-On Time in Cancer Radiation Treatment Using Multileaf Collimators. Natashia Boland Horst W. Hamacher Frank Lenzen January 4, 2002. GOALS. 1. apply radiation to tumor (target volume) sufficient to destroy it while maintaining the functionality of the surrounding organs

sagittarius
Download Presentation

Minimizing Beam-On Time in Cancer Radiation Treatment Using Multileaf Collimators

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Minimizing Beam-On Time in Cancer Radiation Treatment Using Multileaf Collimators Natashia Boland Horst W. Hamacher Frank Lenzen January 4, 2002

  2. GOALS 1. apply radiation to tumor (target volume) sufficient to destroy it while maintaining the functionality of the surrounding organs (organs at risk) 2. Minimize amount of time patient spends positioned and fixed on the treatment couch. 3. Minimize beam-on time (time in which radiation is applied to patient)

  3. Designing the treatment plan: decisions to be made • Location of tumor (target volume) and organs at risk • A discretization of the radiation beam head into bixels • A discretization of the tumor (target volume) and risk organs into voxels • Gantry stops • Amount of radiation released at each stop and in each bixel (intensity function) • How to achieve the intensity function using a multileaf collimator

  4. bixel

  5. Designing the treatment plan: decisions to be made • Location of tumor (target volume) and organs at risk • A discretization of the radiation beam head into bixels • A discretization of the tumor (target volume) and risk organs into voxels • Gantry stops • Amount of radiation released at each stop and in each bixel (intensity function) • How to achieve the intensity function using a multileaf collimator

  6. The amount of radiation released at each stop and in each bixel (the intensity function) can be written as a system of linear equations Px = D dosage vector Di is the radiation of each voxel i accumulated as cumulative radiation from all bixels j. bixel-voxel unit radiation matrix Pij is amount of radiation reaching voxel i if one unit of radiation is released at bixel j. xj is the amount of time radiation is sent off at bixel j Note: this will later be written as two-dimensional intensity matrix, I Must satisfy contraints: eg. lower bound - must destroy cancer upper bound - maintain functionality of organs at risk Note: in general, these constraints are inconsistent and mathematical programming Methods must be used to minimize deviation From the bounds

  7. At this point we assume that 1-5 have been dealt with. So all that remains is to decide on a modulation of the uniform radiation. For each stop of the gantry we have an intensity function, I, where Iij is the amount of time uniform radiation is released in bixel (i,j). For example, if we have chosen a discretization of the beam head into a 6x6 grid, I = is a possible intensity matrix.

  8. This paper focus on using multileaf collimators to achieve modulation. Here, each row of I has an associated pair of leaves - a right leaf and a left leaf. If I has n columns the left leaf may be positioned in column 0,1,…,n, and the right leaf may be placed in columns 1,…,n,n+1, where columns 0 and n+1 are notational columns used to represent the respective leaf’s fully retracted position. I = column n+1 column 0 left leaf positions right leaf positions left leaf < right leaf

  9. This paper focus on using multileaf collimators to achieve modulation. Here, each row of I has an associated pair of leaves - a right leaf and a left leaf. If I has n columns the left leaf may be positioned in column 0,1,…,n, and the right leaf may be placed in columns 1,…,n,n+1, where columns 0 and n+1 are notational columns used to represent the respective leaf’s fully retracted position. I = column n+1 column 0 left leaf positions right leaf positions left leaf < right leaf

  10. This paper focus on using multileaf collimators to achieve modulation. Here, each row of I has an associated pair of leaves - a right leaf and a left leaf. If I has n columns the left leaf may be positioned in column 0,1,…,n, and the right leaf may be placed in columns 1,…,n,n+1, where columns 0 and n+1 are notational columns used to represent the respective leaf’s fully retracted position. Shape matrix S = column n+1 column 0 left leaf positions right leaf positions left leaf < right leaf

  11. K I = S ak Sk k=1 ak > 0 is time the linear accelerator is opened to release uniform radiation Sk is shape matrix

  12. I = S1 = S2 = S3 = I = 3S1 + 1S2 + 2S3

  13. S1 = S2 = S3 =

  14. Multileaf Collimator (MLC) problem with minimal beam-on time min S at subject to S at St = I at >= 0 where t is an element of the index set of all possible shape matrices t t

  15. Multileaf Collimator (MLC) problem with minimal beam-on time K min S ak + (K - 1)Tc subject to S at St = I at >= 0 where t is an element of the index set of all possible shape matrices k = 1 t

  16. Multileaf Collimator (MLC) problem with minimal beam-on time K min S (ak + c(Sk,Sk+1)) subject to S at St = I at >= 0 where t is an element of the index set of all possible shape matrices c(SK,SK+1) = 0 k = 1 t

  17. D 1021 1131 1,0 1,1 1,2 1,3 1022 1132 2031 2131 2,1 2,0 2,2 2,3 2132 2032 3021 3031 3,0 3,1 3,3 3,2 3032 3022 4131 4021 4,1 4,3 4,0 4,2 4022 4132 D’

  18. D 101 102 103 112 113 123 201 202 203 212 213 223 301 302 303 312 313 323 401 402 403 412 413 423 D’