X-ray Microanalysis
Download
1 / 56

- PowerPoint PPT Presentation


  • 136 Views
  • Updated On :

X-ray Microanalysis. An inelastic collision between a primary beam electron and an inner orbital electron results in the emission of that electron from the atom. The energy released from an electron replacement event produces a photon with an energy exactly equal to the drop in energy. .

loader
I am the owner, or an agent authorized to act on behalf of the owner, of the copyrighted work described.
capcha
Download Presentation

PowerPoint Slideshow about '' - nonnie


An Image/Link below is provided (as is) to download presentation

Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server.


- - - - - - - - - - - - - - - - - - - - - - - - - - E N D - - - - - - - - - - - - - - - - - - - - - - - - - -
Presentation Transcript
Slide1 l.jpg

X-ray Microanalysis

An inelastic collision between a primary beam electron and an inner orbital electron results in the emission of that electron from the atom.

The energy released from an electron replacement event produces a photon with an energy exactly equal to the drop in energy.


Slide2 l.jpg

X-rays can have an energy nearly equal to that of the primary beam electron and thus can escape from very deep within the specimen


Slide3 l.jpg

Energy Dispersive primary beam electron and thus can escape from very deep within the specimen

Spectroscopy (EDS or EDX)


Slide4 l.jpg

When an electron from a K-shell is replaced by one from the next closest shell (L), it is designated as a Kα event


Slide5 l.jpg

K next closest shell (L), it is designated as a Kα event b

Ka

When an electron from a K-shell is replaced by one from the second closest shell (M), it is designated as a Kβ event


Slide6 l.jpg

L next closest shell (L), it is designated as a Kα event a - When an electron from a L-shell is replaced by one from the next closest shell (M).

The K shell will never donate its electron as this would require an increase in energy, not a drop.


Slide7 l.jpg

Certain events such as M next closest shell (L), it is designated as a Kα event α, Lβ, and Kγ are only possible in atoms of sufficient atomic weight


Slide8 l.jpg

There are a wide variety of subsets of next closest shell (L), it is designated as a Kα event

X-rays since each electron shell has multiple orbitals


Slide9 l.jpg

# Counts next closest shell (L), it is designated as a Kα event

X-ray Energy in KeV

An X-ray spectrum for a sample is composed of all the possible signals for that given set of elements.

These will differ in terms of energies (KeV) and probabilities (likelihood) scored as number of such signals collected over a given period of time.



Slide11 l.jpg

Positive identification of an element is best done by evaluating the entire family of peaks for a given element.


Slide12 l.jpg

" evaluating the entire family of peaks for a given element.Bremsstrahlung" means "braking radiation" and comes from the original German to describe the radiation which is emitted when electrons are decelerated or "braked" when they interact with the specimen.

Although they contribute to the total X-ray signal they contain no useful information because their energies are nonspecific and therefore are considered as part of the background .


Slide13 l.jpg

Bremsstrahlung evaluating the entire family of peaks for a given element. X-rays are the major part of the continuum X-ray signal that can escape from the deepest portion of the interaction region.


Slide14 l.jpg

Chrysotile Asbestos Fibers evaluating the entire family of peaks for a given element.


Slide15 l.jpg

Bullet fragments (blue) can be identified on cloth fibers and distinguished from other metal pieces by their elemental composition


Slide16 l.jpg

Gunshot Residue (GSR) Analysis and distinguished from other metal pieces by their elemental composition


Slide17 l.jpg

Gunshot Residue (GSR) Analysis and distinguished from other metal pieces by their elemental composition

  • Particles are very characteristic, therefore presence of these particles forms evidence of firing a gun.

  • Particles normally consist of Pb (lead), Sb (antimony) and Ba (barium).

  • New ammunition: environmentally friendly (no Sb).


Slide18 l.jpg

The proportion of elements present in GSR differ slightly and databases of GSR from different manufacturers can be used to identify what ammunition was used in a crime.

GSR is often found on criminals and also on victims if shot at close range.


Slide19 l.jpg

X-ray Mapping and databases of GSR from different manufacturers can be used to identify what ammunition was used in a crime.


Slide20 l.jpg

X-ray analysis of paint fragments and databases of GSR from different manufacturers can be used to identify what ammunition was used in a crime.

The combined (a) backscatter image and X-ray maps of

(b) Au,

(c) Ba

(d) Ca

Different layers of paint can be identified


Slide21 l.jpg

X-ray Detection and databases of GSR from different manufacturers can be used to identify what ammunition was used in a crime.

EDS = Energy Dispersive Spectroscopy

WDS = Wavelength Dispersive Spectroscopy


Slide22 l.jpg

EDS and databases of GSR from different manufacturers can be used to identify what ammunition was used in a crime.

WDS


Slide23 l.jpg

Pulse Processor and databases of GSR from different manufacturers can be used to identify what ammunition was used in a crime.

Measures the electronic signals to determine the energy of each X-ray detected

Analyzer

Displays and interprets

the X-ray data

X-ray Detector

Detects and converts

X-rays into electronic signals


Slide24 l.jpg

Cut-away diagram showing and databases of GSR from different manufacturers can be used to identify what ammunition was used in a crime.

the construction of a typical EDS detector.

Crystal

Collimator

FET

Window


Slide25 l.jpg

Lithium doped Silicon ( and databases of GSR from different manufacturers can be used to identify what ammunition was used in a crime.SiLi) crystal detector

acts as a semiconductor that carries current in a rate proportional to the number of ionization events and acts as an indirect measurement of the energy contained in the X-ray.



Slide27 l.jpg

Each ionized atom of silicon absorbs 3.8 of a eV of energy, so an X-ray of 3.8 KeV will ionize approximately 1000 silicon atoms.


Slide28 l.jpg

Crystal of a

Collimator

FET

Window

Collimator to limit BSE and stray X-rays

Window usually made of beryllium (limited to sodium, atomic number 11) or thin plastic to detect down to boron (Atomic number 5) protects cooled crystal from air.


Slide29 l.jpg

Crystal of a

Collimator

Detector : crystal silicon wafer with lithium added in. For each 3.8 eV from an X-ray, produce an electron and hole. This produces a pulse of current, the voltage of which is proportional to the X-ray energy. Must keep the crystal at LN temperature to keep noise to a minimum.

FET : The field effect transistor is positioned just behind the detecting crystal. It is the first stage of the

amplification process that measures the charge liberated in the crystal by an incident X-ray and converts it to a voltage output.

FET

Window


Slide30 l.jpg

Multichannel Analyzer (MCA) of a

The changes in conductivity of the SiLi crystal can be counted for a given time and displayed as a histogram using a multichannel analyzer.


Slide31 l.jpg

Multichannel Analyzer (MCA) of a

Now

Then

MCA consists of an analog to digital converter which “scores” the analog signal coming from the field effect transistor (FET). Newer systems employ a digital pulse processor which converts the signal on the fly


Slide32 l.jpg

Factors affecting signal collection of a

Distance between detector and X-ray source

Angle at which detector is struck

Volume of signal collected.


Slide33 l.jpg

Take-off Angle of a

For a given angle of electron incidence, the length of the absorption path is directly proportional to the cosecant of the take-off angle, φ


Slide34 l.jpg

Solid Angle of a

The solid angle Ω of a detector is defined as angle of the cone of signal entering the detector. The greater the size of the detector surface area the greater will be the solid angle.


Slide35 l.jpg

Larger of a SiLi crystals will be able to sample a larger volume of signal (better Ω) but because of imperfections in the crystal they have slightly greater noise and thus slightly lower resolution.


Slide36 l.jpg

One can also increase the solid angle by placing the detector closer to the source.

One then tries to maximize both the solid angle and the take-off angle.


Slide37 l.jpg

One reason that the final lens of an SEM is conical in shape is so that the EDS detector can be positioned at a high take-off angle and inserted close to the specimen for a high solid angle.


Slide38 l.jpg

William Henry Bragg is so that the EDS detector can be positioned at a high take-off angle and inserted close to the specimen for a high solid angle.

1862 – 1942

Nobel Prize in Physics

1915

X-ray diffraction in a crystal.

Like an electron beam an X-ray has its own wavelength which is proportional to its energy


Slide39 l.jpg

Crystal: is so that the EDS detector can be positioned at a high take-off angle and inserted close to the specimen for a high solid angle. A solid formed by the solidification of a chemical and having a highly regular atomic structure. May be composed of a single element (C = diamond) or multiple elements.


Slide40 l.jpg

Hexagonal is so that the EDS detector can be positioned at a high take-off angle and inserted close to the specimen for a high solid angle.

Cubic


Slide41 l.jpg

If a wavelength enters a crystal at the appropriate angle it will be diffracted rather than being absorbed or scattered by the crystal


Slide42 l.jpg

For a given wavelength will be diffracted rather than being absorbed or scattered by the crystal λthere is a specific angle θ (Bragg’s angle) at which diffraction will occur.

Bragg’s angle is determined by the d-spacing (interplanar spacing) of the crystal and the order of diffraction (n = 1, 2, 3….).


Slide43 l.jpg

A WDS detector takes advantage of the fact that an X-ray of a given wavelength can be focused by a crystal if it encounters the crystal at the proper Bragg’s angle.

To better accomplish this crystals are bent and ground to form a curved surface which will bring all the diffracted X-ray wavelengths to a single focal point, thus the crystal acts as a focusing lens.


Slide44 l.jpg

To change the Bragg’s angle the diffracting crystal and detector can be moved together relative to the stationary specimen along a circle known as the Roland Circle.


Slide45 l.jpg

WDS detectors are quite large and must be positioned around the specimen chamber at an angle to take advantage of maximum take-off angle and maximum solid angle


Slide46 l.jpg

A microprobe is a specialized SEM that is outfitted with an EDS detector and array of several WDS detectors.


Slide47 l.jpg

Different diffracting crystals can only diffract certain wavelengths (even with the changes in Bragg’s angle) so an array of detectors must be used if one is to be able to detect K, L, and M events for many different elements. Since WDS detectors do not need to be cooled they are windowless and can detect down to Berylium

LiF = Lithium fluoride; PET =Pentaerythritol; and TAP = Thallium acid phthalate.


Slide48 l.jpg

Specimen preparation for WDS wavelengths (even with the changes in Bragg’s angle) so an array of detectors must be used if one is to be able to detect K, L, and M events for many different elements. Since WDS detectors do not need to be cooled they are windowless and can detect down to

Samples must be conductive since high KeV is used (Carbon coating if not naturally conductive)

Samples must be flat (polished) as geometry of sample to detector is crucial and also minimizes artifacts when doing quantitative measurements.


Slide50 l.jpg

A comparison of two spectra collected with EDS and WDS shows how peak overlap and energy spread can serve to obscure the information in an EDS spectrum


Slide51 l.jpg

Quantitative X-ray Analysis how peak overlap and energy spread can serve to obscure the information in an EDS spectrum

If one wants to quantify the relative amounts of different elements present in a complex sample one has to account for a number of factors and carry out a correction of the data


Slide52 l.jpg

One must account for other elements present in the sample and whether their individual peaks overlap with each other creating a “shoulder” that can mask the presence of one element or distort the midpoint of another.


Slide53 l.jpg

Several methods to correct the spectra. ZAF takes into account the Atomic Weight (Z), effects of Absorbance (A) and effects of Fluorescence (F) in adjusting the data to give the correct values.


Slide54 l.jpg

Applications of X-ray Microanalysis account the Atomic Weight (Z), effects of Absorbance (A) and effects of Fluorescence (F) in adjusting the data to give the correct values.

Secondary

Electron image


Slide55 l.jpg

EDS can be added as a component of a TEM account the Atomic Weight (Z), effects of Absorbance (A) and effects of Fluorescence (F) in adjusting the data to give the correct values.

Requires an angled detector (for take-off angle) and scan coils in the column to function as a Scanning Transmission Electron Microscope or STEM.


Slide56 l.jpg

EDS can be used to identify elements present vacuoles or inclusions. Must take into account elements present in the embedding medium