NIH PCKD/Emory University. MRI Imaging Report 1/31/2000. Overview. 2 Patients Scanned with NIH Protocol Both: Comparison with “Old Protocol” Visual comparison No SNR measurements performed No Breathhold Flow Quantification (Yet) Philips scanner should be capable

michon
Download Presentation

NIH PCKD/Emory University

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. NIH PCKD/Emory University MRI Imaging Report 1/31/2000

  2. Overview • 2 Patients Scanned with NIH Protocol • Both: Comparison with “Old Protocol” • Visual comparison • No SNR measurements performed • No Breathhold Flow Quantification (Yet) • Philips scanner should be capable • Little experience; Validation experiments?

  3. T2-Weighted Multi-Slice (#1) Old Protocol (3 mm) Multiple Breathholds NIH Protocol (5 mm) Single Breathhold

  4. T2-Weighted Multi-Slice (#2) Old Protocol (3 mm) Multiple Breathholds NIH Protocol (5 mm) Single Breathhold

  5. T2-Weighted Imaging: Remarks • 3 mm slice thickness resolves cysts better • Fat Suppression useful, works well • Multiple-breathhold: registration needed • Kidney “rigid object”: overlap + affine Xform • Avoid misregistration between interlaced stacks • Role in image analysis?

  6. T1-Weighted 3-D (PRE-#1) Old Protocol (2.5/5 mm) =40 Single Breathhold NIH Protocol (2.5/5 mm) =12 Single Breathhold

  7. T1-Weighted 3-D (PRE-#2) Old Protocol (2.5/5 mm) =40 Single Breathhold NIH Protocol (2.5/5 mm) =12 Single Breathhold

  8. T1-Weighted 3-D (POST-#1) Old Protocol (2.5/5 mm) =40 90 s post-Gado NIH Protocol (2.5/5 mm) =12 120 s post-Gado

  9. T1-Weighted 3-D (POST-#2) Old Protocol (2.5/5 mm) =40 60 s post-Gado NIH Protocol (2.5/5 mm) =12 120 s post-Gado

  10. T1-Weighted 3-D heart liver Ghost artifact due to heart motion (?) Apply pre-saturation slab anterior to volume to reduce intensity?

  11. T1-Weighted Imaging: Remarks • NIH protocol (=12) better overall SNR • Pre- & post-contrast: more complex image • Segmentation easier? (CNR measurements) • Coil placement important! Difficult? • Pre-saturation slabs? Added acq. time? • Older patients: • Many breathholds taxing to patient • Only 90 or 120 s post contrast?