1 / 15

A Design of PET detector using Microchannel Plate PMT with Transmission Line Readout

A Design of PET detector using Microchannel Plate PMT with Transmission Line Readout. Heejong Kim 1 , Chien-Min Kao 1 , Chin-Tu Chen 1 , Jean-Francois Genat 2 , Fukun Tang 2 , Henry Frisch 2 , Woong-Seng Choong 3 , William Moses 3 1. Department of Radiology, University of Chicago, IL

hall
Download Presentation

A Design of PET detector using Microchannel Plate PMT with Transmission Line Readout

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. A Design of PET detector using Microchannel Plate PMT with Transmission Line Readout Heejong Kim1, Chien-Min Kao1, Chin-Tu Chen1, Jean-Francois Genat2, Fukun Tang2, Henry Frisch2, Woong-Seng Choong3, William Moses3 1. Department of Radiology, University of Chicago, IL 2. Enrico Fermi Institute, University of Chicago, IL 3. Lawrence Berkeley National Laboratory, Berkeley, CA

  2. 1. Introduction • The advantages of using Microchannel Plate(MCP) PMT. • Position sensitiveness. • Fast time response. • Compact size than conventional PMT. • LSO scintillator • High Light yield( 25000~30000/MeV) • Fast decay time( ~40ns) • Transmission Line readout scheme. • Readout both ends of the strip. • Position measurement by time difference • Efficient reduction of # of readout channel( NxN -> 2N) • Readout at both ends( Scintillator sandwiched by MCPs) • Possible to extract Depth of Interaction(DOI) • A PET detector design, using pixelated array of LSO scintillator with MCP PMT, has been investigated. Fast timing characteriscs of MCP combined with high sensitivity LSO makes this design suitable for TOF PET application. By design, DOI information is available by reading out the signals at both ends of scintillator. The preliminary results of Geant4 simulation study are presented here. The real tests to validate the simulation has been conducted with Photonis planacon MCP(XP85022) and the results are also shown.

  3. MCP & Transmission Line board Fig.1 Photonis Planacon MCP(XP85022) with 1024(32x32) anodes(left) and Transmission line(TL) baord with 32 microstrip (right). One microstrip is connected to one raw of MCP anode(32) and signals are readout at both ends of a TL.

  4. 2.Material and Methods • One detector module consists of 24x24 array of LSO scintillator and 2 MCP assemblies. • Two detector modules facing each other. • 5cm distance between them. • LSO pixel dimension : 4x4x25mm3. • Crystal pitch : 4.25mm • MCP assembly dimension : 102x102x9.15mm3. It includes photocathode and TL structure. (MCP with 8’’x8’’ area is under development.)‏ • MCP is coupled to LSO at both front and back ends. A. Detector configuration

  5. Scintillators sandwiched by MCPs Fig. 2 Simulation set-up with two detector modules. Each module consist of 24x24 array of pixelated LSO scintillators and two MCPs coupled to the scintillators at both front and back side.

  6. B. Simulation Setup • Optical Photon generation and transport was simulated by Geant4. • Two 511keV gammas are generated back to back at the middle of two detector modules and sent to the detector centers. • The reflective media was inserted between crystals. • The surface between LSO slab and MCP glass was optically coupled with the optical grease. • LSO characteristics( simulation input parameters) • Light yield : 30,000/MeV • Decay time : 40ns • Re‏solution : 10.4%( FWHM)

  7. Signal Readout Scheme • Electrical signal was formed based on the measured XP85022 characteristics combining with the Geant4 simulation outputs: optical photon’s position and arrival time at photocathode. • For each individual photo electron, the measured single photo electron response was assigned. Convolute pulses due to all the photo-electron within the area of TL strip. • TL signal then propagates to both ends of TL. • In the forward MCP, 24 TL strips run vertically. By applying Anger logic to measured TL signals, X coordinate can be obtained. • TLs runs horizontally in the backward MCP to get Y coordinate in the same way. • The position also can be measured from the measured time difference at both ends of TL.

  8. 3. Experimental Tests • XP85022 Chevron type, 10um pore • Textronix DPO7354 Digital Oscilloscope recorded the waveform of TL at 10GS/s. • The charge of pulse are obtained by integrating TL waveform. The test set-up was built using a XP85022 MCP and TL board to measure the characteristics of the MCP. The measured single photo-electron response(SER) was fed to the simulation for the electrical signal. Fig. 3 MCP/TL assembled for the real test. 4 TL channels were connected through SMA to the DPO7354 Oscilloscope. A LSO crystal with 1x1x10mm3 was placed on top of the XP85022 MCP surface.

  9. A. Single Electron Response(SER) • SER was measured using the pulsed LED as a light source. • The rise time of SER was measured ~560ps. • The SER signal was spread in ~5 TL. • The XP85022 gain at HV = -2300V : 1.5 x 106 mV Fig. 4 Integrated charge of SER waveforms(left). Averaged waveform of SER; the maximum TL signal only (middle). XP85022 MCP gain as a function of HV

  10. B. Responses to 511keV photon • MCP/TL coupled to 1x1x10mm3 LSO crystal. • Hamamatsu R9800PMT with 6.2x6.2x25mm3 LSO for coincidence • Use Na22 for positron source.‏ • Waveform recorded by Tektronix DPO7354 scope E resolution = 13.8% fwhm Fig.5 Test set-up for 511keV gamma coincidence (left). Energy distribution of R9800PMT(right).

  11. Energy( real test)‏ • Charge sum of 3 TL signal : only left side of TL. • Compton + 511keV peak structure is clearly found. • Discrepancy between the real test and simulation. • E resolution : 22.3% vs 15.8% ( at 511keV peak) • Shape of compton continuum. • Due to simplified simulation setup( gamma direction). Test set-up simulation Real Test 15.8% fwhm 22.3% fwhm Fig. 6 Energy sum of 3 TL signal by 511keV photon.

  12. Coincidence Timing ( real test)‏ • Event selection requirement for the coincidence timing. R9800PMT : 400 < E < 600 keV MCP 3TL Sum : 35 < Int. Charge < 60pC • Coincidence timing resolution = ~416ps( FWHM)‏ contribution from R9800PMT side = ~200ps (FWHM)‏ Real Test ~416ps Test set-up simulation ~398ps Fig. 7 coincidence time distribution.

  13. 4. Results – Energy & Timing • Sum of 5 TL signals around the maximum amplitude. • Energy resolution : ~11% • Use the measured XP85022 SER for the TL signal. • The event time was extracted by Leading Edge(LE) to the maximum TL signal. ( Threshold : 3mV)‏ • Energy window [450, 600] keV required for coincidence event. • The detection efficiency : ~40%( ~63% for one module). • Coincidence timing resolution : ~323 ps. E_res = 11%(fwhm) ~323ps(fwhm) Fig 8. Energy (left) and Coincidence timing distribution (right)

  14. Results - DOI • 511keV gamma injected from side of detector with 1mm step along Z axis. • Energy asymmetry and time difference of front and back due to different interaction depth. • (EFront – EBack)/(EFront + EBack) • Clear correlations were found. Fig. 9 Energy asymmetry( left) and time difference( right) measured at both front and back MCP as a function of depth of interaction.

  15. 5. Summary • A PET detector design using pixelated array of LSO scintillator and MCP PMT with Transmission Line readout was studied. • Geant4 was used for optical photon simulation. • Real test setup using XP85022 MCP and TL board was built to measure SER of MCP. The measurement from the test set-up was fed to the simulation for TL signal forming. • The preliminary results from the study show promising results. • Energy resolution~11% at 511keV was obtained. • The coincidence time resolution ~323ps with ~40% detection efficiency were estimated. • Readout at both ends of scintillator makes it possible to extract the DOI information.

More Related