Determination of Tritium in Hydrosphere Samples by Ultra Low Level Liquid Scintillation Quantulus 1220 Barbora Sedlářová. Water Research Institute T. G. Masaryka, Podbabská 30, 160 62 Prague 6, e- mail: barbora_sedlarova@vuv.cz. Intoduction. Quantulus 1220

loader
I am the owner, or an agent authorized to act on behalf of the owner, of the copyrighted work described.
capcha
Download Presentation

Water Research Institute T. G. Masaryka, Podbabská 30, 160 62 Prague 6,


An Image/Link below is provided (as is) to download presentation

Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server.


- - - - - - - - - - - - - - - - - - - - - - - - - - E N D - - - - - - - - - - - - - - - - - - - - - - - - - -
Presentation Transcript
slide1

Determination of Tritium in Hydrosphere Samples by Ultra Low Level Liquid Scintillation Quantulus 1220Barbora Sedlářová

Water Research Institute

T. G. Masaryka, Podbabská 30, 160 62 Prague 6,

e-mail: barbora_sedlarova@vuv.cz

intoduction
Intoduction
  • Quantulus 1220
  • Project requisted by the Ministry of Environment: The Contamination of Hydrosphere by radionuclides
  • Determination of tritium in surface water: Labe Hřensko, Odra Bohumín,Dyje Pohansko
  • Determination of tritium in rain-fall: Závišín
method of 3 h determination
Method of 3H determination
  • SOP-RA-7: ČSN ISO 9698
  • 20 ml PE vial: 8 ml water sample + 12 ml Optiphase HiSafe – 3
  • Counting efficiency: ~ 22%
  • Calibration: standard of tritium, CMI IIZ
  • Counting time: t = 800 min
  • Background: ~ 0,6 CPM
  • Detection limit: level of significancy

a = b = 0,05 (P = 95%), cND 1,2 Bq/l, cNV 0,6 Bq/l

  • Control of quality: verification of instrument byCMI IIZ, interlaboratorycomparison test, laboratory quality system is assessed by ASLAB
areas of tritium detection
Areas of tritium detection

Labe Hřensko

Odra Bohumín

Závišín

Dyje Pohansko

slide14

Deposition of tritium in river-basin Labe and outfall of tritium in Labe – Hřensko

during 1993 – 2002 include highlight increase of ETE

(nuclear power station Temelín) inpact in 2002

Rain-fall

ETE

Labe

slide16

Thank you for your attention