1 / 2

Horiba ORP Electrode Solution Table

Find mV values for temperatures from 5°C to 40°C. Polish platinum wire surface for accuracy. Valid for 2 hours.

Download Presentation

Horiba ORP Electrode Solution Table

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. MODEL 160-51 SOLUTION ORP mV Value vs. Temperature 5° C +111.9 10°C +106.9 15°C +101.0 20°C +95.0 25°C +89.0 30°C +82.7 35°C +76.2 40°C +69.0 The accuracy of this solution is +/- 15mV, valid only for 2 hours. This table applies to Horiba ORP electrodes. Polish the platinum wire surface if readings are out of range.