An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. KCL GAS INSTRUMENTS KSS (tower): Licor7500 (CO2, H2O) Licor840_18 on tower (CO2, H2O) WXT510 (H2O) KSK (tower): WXT510 (H2O) LHR (met station): RH instrument (H2O) 30 km

  2. CO2 Profile: Hardware Li840_16/1 (KSS tower) Rotometer Array Valve Array CR1000 Datalogger TOWER ROTOMETER ARRAY VALVE ARRAY GAS ANALYSER PUMP DATA LOGGER COMPUTER

  3. STRAND ROOF • Laptop (Roof13) • Program: Li840.pyc • Saves to folder: C:\Li840_1\dataFilename: LI-1yyyyddd_hh_raw.txt • Li840_reader.cmd • Ports • Log Files • Program: KSS GasProfile 19_11_09.CR1 • Saves to folder: • C:\Profile\dataProfile_valve_array • Split to folder: C:\Profile\data\ • Filename: PVyyyyddd_hh.txt WXTGasGrabCeilAB_Move_DriveI 00:11 hourly D:\Roof\CO2\WXTGasGrabCeilAB_yyyy_v#.m CopyLicor???? D:\Roof\CO2\CopyLicor_v#.m Cannot alter data on drive I, copy to drive A Licor840_16 USB hub CR 1000 I:\yyyy\KSS\ LI840-#\mm\RAW\ LI-#yyyyddd_hh_raw.txt CProfile\Valve_data\mm\ VAyyyyddd_hh.txt A:\works\yyyy\London\Level_0\KSS\DAY\ddd\ LI-#yyyyddd_hh_raw.txt Valve_Array_Split 00:08 hourly D:\Roof\Split\HourlyData Split_Valve_Array_v#.bat Splits valve data into hourly files ReadLicor.sh Data02 home/micromet/ script/cronJobs/ Strips out xml code and overwrites file LI840NCDF 02:15 daily D:\Roof\CO2\Li840NCDF_yyyy_v#.m Merges valve and Li-840 files. Interpolates missing values. Averages L0 (2 Hz) data to 1.5 min data (L1) Level 0 (2Hz) merged data A:\data\yyyy\London\L0\KSS\DAY\ddd\ Li840-#_KSS_yyyyddd_0_5s.nc All variables plotted per valve over 3 months. A:\data\2011\London\L1\KSS\PROFILES Li840-#_KSS_yyyy _1_5min_####.pdf mergeLi840.sh Data02/home/micromet/script/cronJobs/ Merges daily L1 files into an annual file. NetcdfProfiler_v# 05:16 daily D:\Roof\netCDF\NetcdfProfiler_v#.m Generates .pdf data summaries. Level 1 (1.5min) merged data A:\data\yyyy\London\L1\KSS\DAY\ddd\ Li840-#_KSS_yyyyddd_1_5min.nc Pdf2web.sh Data02 /home/micromet/ script/cronJobs/ Uploads pdfs to web. MustRunLast_Error_Summary06:35 daily D:\Roof\ErrorSummary\ ErrorSummary_v#.m Diagnoses problems, emails micromet. Level 1 (1.5min) annual data A:\data\yyyy\London\L1\KSS\ANNUAL\ddd\ Li840-#_KSS_yyyyddd_1_5min.nc GasCompPlots6:15 daily D:\Roof\TOM\CO2H2Ocomp_yyyy_v#.m Creates daily CO2/H2O plots. Uploads to web. http://geography.kcl.ac.uk/ micromet/index.htm KUMA-Data Archive Diagnostics http://geography.kcl.ac.uk/micromet/CO2_H2O_plot.htm

  4. CO2 Profile: Troubleshooting Check which files are present and determine at which stage of processing the system has failed, e.g. if the raw data in the works folder on the A drive still has xml tags, ReadLicor.sh needs to be run (look at list of netcdf programs to get the correct directory). If a failure occurs early on in the stages of processing e.g. Point 1, all subsequent scheduled tasks will need to be rerun. If the data was never collected from roof13, check the computer is on and the relevant programs are running (see next 4 slides). If this fails, email teama@kcl.ac.uk

  5. The programs and data for the LICOR are stored here This must be the same version of Python as that used to compile the LICOR programs, i.e. 2.6

  6. The program and script are stored in this folder Data is output to this folder Double left click to run the script. A command window should appear but check under Task Manager. Right click and select ‘Edit’ to change the settings

  7. The prefix denotes whether the information is coming from LI-849_16 (1) or LI-840_18 (2). LI-2 is no longer in use Controls where the data is stored on roof13 Program to run the script Script to run The COM port to which the instrument is connected. This may be determined by: • Start • Control Panel • System • Hardware • Device Manager • Ports (COM and LPT) Some systems may allow you to bypass some of these steps.

  8. Check that the data is being stored by noting down the active file size and refreshing the window after a couple of minutes.