Quality Assessment for Routine Clinical MR Spectroscopy. Diana M. Lindquist, PhD Assistant Professor of Radiology. Purpose. Clinical indication: apparently abnormal metabolite ratios in infant with normal imaging findings.

loader
I am the owner, or an agent authorized to act on behalf of the owner, of the copyrighted work described.
capcha
Download Presentation

PowerPoint Slideshow about 'Quality Assessment for Routine Clinical MR Spectroscopy' - cheyenne-turner


An Image/Link below is provided (as is) to download presentation

Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server.


- - - - - - - - - - - - - - - - - - - - - - - - - - E N D - - - - - - - - - - - - - - - - - - - - - - - - - -
Presentation Transcript
quality assessment for routine clinical mr spectroscopy

Quality Assessment for Routine Clinical MR Spectroscopy

Diana M. Lindquist, PhD

Assistant Professor of Radiology

purpose
Purpose
  • Clinical indication: apparently abnormal metabolite ratios in infant with normal imaging findings.
  • Quality control is as important in clinical spectroscopy as in research
  • Develop a simple short protocol for quality control in clinical MRS
    • Localization
    • Spectral data
methods
Methods
  • Standard head coil and manufacturer’s phantom
  • Protocol
    • Axial localizer
    • 3 voxels (8 cm3)
    • PRESS
  • Total time: 30 minutes
results voxel localization
Results: Voxel Localization
  • Mean voxel size was 20.2 (± 0.8) by 18.9 (± 0.4) mm2 in-plane
  • One voxel is adequate; scan time reduced from 30 to 15 minutes
sample spectra
Sample spectra

NAA

Cr

Cho

Lac

TE 35 ms

TE 144 ms

conclusions
Conclusions
  • Following implementation of this protocol, we are now able to detect and correct scanner instabilities.
  • QC data are useful as a monitor of system performance
  • Scheduling still difficult even with a 15 minute exam.