UNCERTAINTY IN RELATIVE HUMIDITY – CALIBRATION OF PSYCHROMETERS

Tempmeko 2004 - Workshop on Uncertainty in Humidity Measurements. UNCERTAINTY IN RELATIVE HUMIDITY – CALIBRATION OF PSYCHROMETERS. Anders Kentved NATIONAL REF. LAB. FOR HUMIDITY DELTA, DK Cavtat – Dubrovnik, 21 June 2004.

berke
Download Presentation

UNCERTAINTY IN RELATIVE HUMIDITY – CALIBRATION OF PSYCHROMETERS

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Tempmeko 2004 - Workshop on Uncertainty in Humidity Measurements UNCERTAINTY IN RELATIVE HUMIDITY – CALIBRATION OF PSYCHROMETERS Anders Kentved NATIONAL REF. LAB. FOR HUMIDITY DELTA, DK Cavtat – Dubrovnik, 21 June 2004

  2. Uncertainty in Relative Humidity – Calibration of Psychrometers:Summary • Introduction • Calibration of psychrometers at DELTA • Psychrometer basics – a brief review • Psychrometer equations and sensitivity coefficients • Sources of uncertainty related to calibration of psychrometers • Worked example – Calibration of an Electropsychrometer • Sources of uncertainty related to use of psychrometers • Typical calibration results • Conclusion • Discussion

  3. Calibration of psychrometers at DELTA:Background – Capability • Need for reliable instruments for measuring relative humidity during climatic testing of electronics • Typical test conditions:10 %RH – 97 %RH / 15°C – 85°CTolerance level < ± 2-3 %RH • Calibration method:Comparison in climatic chamber using dew-point hygrometer and thermometers • Capability:10 %RH – 97 %RH / 15°C – 85°CWorst case uncertainty < ± 1 %RH (k=2) for entire range(3 psychrometers calibrated in parallel)

  4. Psychrometer basics – a brief review:Measuring principle • Two thermometers:- A “dry”, for measuring the air temperature t [°C]- A “wet”, sheathed in wet wicking, for measuring the “wet bulb temperature”tw [°C] • A fan draws air across the thermometers • Evaporation from the wick cools the wet thermometer • Relative Humidity is calculated from t and tw

  5. Psychrometer basics – a brief review:Psychrometric chart

  6. t = Air temp. [°C] e’s(tw,P) = Sat. vap. pres. at tw and P [Pa] tw = Wet bulb temp. [°C] e’s(t,P) = Sat. vap. pres. at t and P [Pa] P = Total pressure [Pa] A and B = Constants Psychrometer equations and sensitivity coefficients:Typical equations

  7. Psychrometer equations and sensitivity coefficients:Sensitivity coefficients – Effect on RH of 1 °C change of t and tw

  8. Psychrometer equations and sensitivity coefficients:Sensitivity coefficients – Effect on RH of 1 hPa change of P

  9. Sources of uncertainty related to calibration of psychrometers:Most significant contributions – Typical values • Uncertainty of reference temperature measurement in calibration chamber – typical increase of uncertainty due to psychrometers:± 20 mK (stability in time) • Uncertainty of reference dew-point measurements in calibration chamber – typical increase of uncertainty due to psychrometers:± 20 mK (stability in time) • Resolution of psychrometer temperature measurement, typically:± 10 mK to ± 100 mK • Stability (short term) of psychrometer temperature measurement, typically: ± 10 mK • Readings of psychrometer, typically: ± 0.1 %RH • Reproducibility of psychrometer, typically: ± 0.2 to ± 0.5 %RH(if caused by contamination of wick, typically:+ 0.2 % to + 0.5 %RH)

  10. Sources of uncertainty related to calibration of psychrometers:Often overlooked contributions – Typical values • Uncertainty of reference temperature measurement in calibration chamber – typical increase of uncertainty due to psychrometers:± 100 mK (stability in calibration space) • Changes in atmospheric pressure (P) during calibration can cause uncertainty of psychrometer reading up to ± 1 %RH – measure atmospheric pressure during calibration! • Psychrometers are very non-linear, uncertainty calculation is valid only for the specific calibration point (t, %RH) • Readings of mercury thermometers, typically: ± 50 mK • Improper fitting of psychrometer wick, typically:± 0.5 %RH to ± 1.0 %RH

  11. Worked example – Calibration of an Electropsychrometer:Uncertainty budget for calibration at 30° / 70 %RH Calculated reference value = 70.3 %RH Average of psychrometer readings = 68.9 %RH Correction to be applied = 70.3 %RH – 68.9 %RH = 1.4 %RH Expanded uncertainty (k=2) of correction = 0.7 %RH

  12. Sources of uncertainty related to use of psychrometers: Most significant contributions – Typical values • Long term stability – drift of thermometers and measuring electronics, typically: ± 0,5 %RH to ± 1.0 %RH • Thermal drift of measuring electronics during use at high or low temperatures – correlated and cancel out to some extent, typically ± 0.2 %RH to ± 0.5 %RH • Uncertainty of readings from psychrometric charts • Changes in atmospheric pressure (P) during use causes uncertainty of psychrometer reading up to ± 1 %RH – correction possible if pressure is known • Improper fitting of psychrometer wick, typically:± 0.5 %RH to ± 1.0 %RH • Contamination of psychrometer wick – don’t reuse wick, change it!

  13. Typical calibration results:DELTA Electropsychrometer EP02 - Corrections

  14. Uncertainty in Relative Humidity – Calibration of Psychrometers:Conclusion • Psychrometers increase the uncertainty of reference temperature measurement in calibration chambers. Reasons: Power dissipated in fan and evaporation of cooled air/water vapour from wet bulb • Psychrometers increase the uncertainty of reference dew-point measurement in calibration chambers: Reason: Water evaporation from wet bulb • The influence of a psychrometer on calibration uncertainty depends significantly on the size of the calibration chamber, stirring and where the reference temperature is measured • The calibration measurement capability could be based on a small psychrometer capable of stable and high resolution temperature measurement

  15. Uncertainty in Relative Humidity – Calibration of Psychrometers:Discussion • Which uncertainty contributions do we need to include in CMC’s? • How many different calibration methods exists for calibration of psychrometers? • Is it common practice to calibrate the psychrometer coefficient A instead of direct comparison of %RH reading? • Is it common practice to include readings of atmospheric pressure (P) in calibration certificates? • ----