1 / 43

Quality Control in Diagnostic Radiology

Quality Control in Diagnostic Radiology. Factors driving Q.C. Why do we do it?. Legal Requirements Accreditation JCAHO ACR Clinical improvement equipment performance image quality. Medical Physicists at Work. Q.C. Goals. Minimize dose to patients staff Optimize image quality

amiel
Download Presentation

Quality Control in Diagnostic Radiology

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Quality Control in Diagnostic Radiology

  2. Factors driving Q.C.Why do we do it? • Legal Requirements • Accreditation • JCAHO • ACR • Clinical improvement • equipment performance • image quality Medical Physicists at Work

  3. Q.C. Goals • Minimize dose to • patients • staff • Optimize image quality • Establish baselines • More on this in a moment

  4. Why is Q.C. Important? Without a QC program the only way to identify problems is on patient images. And some problems don’t show up on images. Yeah, that’s what I always say.

  5. QC can detect • Malfunctions • Unpredictability • may be hard to isolate clinically • Inefficient use of Radiation • high fluoroscopic outputs • Radiation not reaching receptor • inadequate filtration • oversized collimation

  6. Goals of a Q.C. Program • Obtain acceptable image with least possible radiation exposure to • patients • staff • Attempt to identify problems before they appear on patient films • without QC problems only detected on patient films

  7. “Acceptable” Image • Image containing information required by radiologist for correct interpretation • goal: minimize exposure while maintaining acceptability • high exposure images often have excellent appearance • Low noise

  8. Q.C. & Baselines • Baselines • quantitative data on equipment obtained during normal operations • Baselines useful for troubleshooting • isolating problem component, for example • generator • processor • Allows efficient use of engineering / repair personnel

  9. X-Ray Quality Control • Filtration • Focal Spot Size • Collimation • Maximum Fluoroscopic Output • Calibration Verification • Phototimer Performance

  10. Why is Filtration Important? • Tube emits spectrum of x-ray energies • Filtration preferentially attenuates low energy photons • low energy photons expose patients • do not contribute to image • low penetration

  11. Half Value Layer (HVL) • We don’t measure filtration • We measure HVL • HVL: amount of absorber that reduces beam intensity by exactly 50%

  12. kVp HVL (mm Al) 30 0.3 40 0.4 49 0.5 50 1.2 60 1.3 70 1.5 71 2.1 80 2.3 90 2.5 100 2.7 110 3.0 120 3.2 130 3.5 140 3.8 150 4.1 Half Value Layer • Depends upon • kVp • waveform (single/three phase) • inherent filtration • Minimum HVL regulated by law • Maximum HVL regulated only in mammography Georgia State Rules & Regulations for X-Ray

  13. Radiographic HVL Setup

  14. Checking HVL Compliance(Radiographic) • How much aluminum must be placed in beam to reduce intensity by exactly 50%? 90 kVp Measurements; 2.5 mm Al minimum HVL filter mR (mm Al) ------------------- 0 250 2.5 133 filter mR (mm Al) ------------------- 0 250 2.5 125 filter mR (mm Al) ------------------- 0 250 2.5 111 Not OK! Must remove Al to reduce beam to exactly 50% OK! Must add Al to reduce beam to exactly 50% Acceptable HVL > 2.5 mm Marginal HVL = 2.5 mm Unacceptable HVL < 2.5 mm

  15. Checking HVL Compliance(Radiographic) • Is this machine legal? • 2.5 mm Al minimum filtration at 90 kVp 90 kVp Measurements filter mR (mm Al) ------------------- 0 450 2.5 205

  16. Fluoroscopic HVL Setup

  17. Fluoroscopic HVL • Set desired kilovoltage manually • measure exposure rates instead of exposure • Move absorbers into beam as needed

  18. Actual Focal Spot Apparent Focal Spot Focal Spot Size • We measure apparent focal spot • Trade-off • smaller spot reduces geometric unsharpness • larger spot improves heat ratings

  19. Focal Spot Size (cont.) • Focal spot size changes with technique • Standard technique required • 75 kV (typical) • 50% maximum mA for focal spot at kV used • direct exposure (no screen) • NEMA Standardsdefines tolerances Nominal Size Tolerance ------------------------------------- >1.5 mm 30% >0.8 and <=1.5 mm 40% <0.8 mm 50%

  20. Focal Spot Measuring Tools • Direct MeasurementPin Hole Camera • Slit Camera • Indirect Measurement of Resolving Power • Star Test Pattern • Bar Phantom

  21. Direct Focal Spot Measurement • Measure focal spot directly in each direction • Use triangulation to correct for distances • formula corrects for finite tool size • two exposures required for slit Slit Camera Pinhole Camera

  22. Star Test Pattern • Measures resolving power • infers focal spot size • Dependent on focal spot energy distribution • measure • largest blur diameter (in each direction) • magnification • use equation to calculate focal spot size

  23. Bar Phantom • Measures resolving power • Find smallest group where you can count three bars in each direction

  24. Bar Phantom Setup

  25. Radiographic Collimation • X-Ray / Light Field Alignment • Beam Central Axis • should be in center of x-ray beam • Collimator field size indicators • PBL (automatic collimation) • field automatically limited to size of receptor • Bucky Alignment • Using longitudinal bucky light & transverse detent, x-ray field should be centered on bucky film

  26. X-Ray / Light Field Alignment • Mark light field on table top with pennies

  27. Radiographic X-Ray / Light Field Alignment

  28. Fluoroscopic Collimation • image field is scale seen on monitor • expose film on table above scale • compare visual field (monitor) with x-ray field on film • must check all magnification modes

  29. Fluoroscopic Collimation

  30. Fluoroscopic Collimation

  31. Maximum Fluoro Output • put chamber in beam on tabletop • block beam with lead above chamber • fools generator into providing maximum output • 10 R/min. limit for ABS fluoro

  32. Maximum Fluoro Output Lead

  33. CalibrationPerformance Parameters • Timer Accuracy • Repeatability • Linearity/Reciprocity • Kilovoltage accuracy • mA • must be measured invasively

  34. Calibration 120 kVp • mR/mAs should stay constant for all combinations of mA & kVp at any particular kVp mA time mAs mR mR / mAs (msec) ------------------------------------------------------ 100 .1 10 240 24 200 .05 10 ? ? 50 .2 10 ? ? Constant mAs

  35. Calibration 120 kVp • mR/mAs should stay constant for all combinations of mA & time at any particular kVp mA time mAs mR mR / mAs (msec) ----------------------------------------------------- 100 .1 10 240 24 200 .1 20 ? ? 100 .4 40 ? ? Double mAs Double mAs again

  36. Phototiming(check with output or film) • Reproducibility • Density Controls • Field Placement • Field Balance Phototiming Operation should be Predictable

  37. Density Control -2 -1 0 1 2 41 49 62 76 96 Phototimer Density Control Settings R R T a b l e t o p

  38. Phototiming Density Steps should be predictable & approximately even

  39. M e a s u r e m e n t o f P h o t o t i m e r F i e l d P l a c e m e n t / B a l a n c e Lead for checking field placement R R T a b l e t o p Phototimer Field Placement / Balance • Placement • cover desired field with lead • select field as indicated • Balance • no fields covered • select field as indicated

  40. Phototimer Field Placement / Balance

  41. Phototimingchecked with Exposure Index • kV Response • phototimer pick-up attenuation may vary with kV • phototimer must track kV response of rare-earth film • Rate Response • Check with varying • phantom (lucite) thickness • mA

  42. kV Response 4 Optical Density 2 0 70 81 90 kilovoltage kV/Rate Response kV 70 81 90 Lucite 17.5 4.5 4.9 5.2 Depth 12.5 4.7 (cm) 7.5 4.7 Thickness Tracking 4 Density Optical 2 0 17.5 12.5 7.5 Lucite Thickness

  43. The End Any questions, you varmints?

More Related