An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Production of Insulin Reverse Phase – High Pressure Liquid Chromatography Unit (RP-HPLC) Presented by: Justin McComb Rachelle Bolton Young Chang

  2. Overview • Purpose of the Unit • Principles of RP-HPLC • Design Validation Equations • Organic Modifiers • Resin Design Options • Process Design Considerations • Cost Analysis • Final Process • Final Design

  3. Purpose of Unit • The unit purifies native insulin by removing impurities such as: • insulin ester • denatured insulin • partially cleaved precursor components • The second RP-HPLC used in the production of insulin is used to purify the human insulin that has been produced.

  4. Principles of RP-HPLC • RP-HPLC is a technique by which differences in polarity of compounds can be used to separate them from a mixture into their components • Chromatography functions through mass transfer between a mobile and stationary phase • Stationary phase (packing): non-polar resin • Mobile phase (solvent): polar liquid • As the mobile phase passes through the column, the components within that phase will have different affinities for the stationary phase.

  5. Principles of RP-HPLC • This will affect the elution time of each compound, and will cause the mixture to separate into its components.

  6. Principles of RP-HPLC

  7. Design Validation Equations Ergun Equation • Rep = Reynolds • fp = friction factor • Q = volumetric flowrate • A = x-sectional area • ρ = density • μ = viscosity • L = column length • ΔP = pressure drop • ε = void fraction • Dp = resin diameter Laminar Flow Validation Pressure Drop Calculation

  8. Summary Table of Organic Modifiers

  9. Resin Design Option #1

  10. Resin Design Option #2

  11. Process Design Considerations Scale Up Constant Length Constant Linear Flowrate

  12. Process Design Considerations

  13. Cost Analysis • Capital Cost (Hamilton estimates): • 20 units x $20000/unit = $400,000 • Operating Costs • Resin Cost: $10,000/unit • Solvent Cost: Encompasses 80% of total operating cost • Energy Cost: • Cold water • Pump (vs. pressure drop)

  14. Final Process • 20,000 mg of the insulin solution is dissolved in 1.5 L water, 10% 2-propanol • Column is regenerated with 0.5N NaOH, washed with water, then washed with 80% isopropyl alcohol containing 0.1% trifluoroacetic acid • Column is equilibrated with 5 column volumes of Buffer A • Insulin solution is applied at 100 cm/h flow rate • Column washed with 3 column volumes of 20% Buffer B and 80% Buffer B • Buffer B increased from 20% to 40% in 1 column volume • Native insulin eluted in a linear gradient of 40-50% buffer B in 30 column volumes • 16,000 mg insulin (98% purity) generated

  15. Final Design

  16. Questions • ????