Electron probe microanalysis epma
This presentation is the property of its rightful owner.
Sponsored Links
1 / 28

Electron Probe Microanalysis EPMA PowerPoint PPT Presentation


  • 92 Views
  • Uploaded on
  • Presentation posted in: General

Revised 10-19-2005. UW- Madison Geology 777. Electron Probe Microanalysis EPMA. Wavelength Dispersive Spectrometry (WDS) I. UW- Madison Geology 777. Generic EMP/SEM. Electron gun. Column/ Electron optics. Optical microscope. EDS detector. Scanning coils. SE,BSE detectors.

Download Presentation

Electron Probe Microanalysis EPMA

An Image/Link below is provided (as is) to download presentation

Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server.


- - - - - - - - - - - - - - - - - - - - - - - - - - E N D - - - - - - - - - - - - - - - - - - - - - - - - - -

Presentation Transcript


Electron probe microanalysis epma

Revised 10-19-2005

UW- Madison Geology 777

Electron Probe MicroanalysisEPMA

Wavelength Dispersive Spectrometry (WDS) I


Generic emp sem

UW- Madison Geology 777

Generic EMP/SEM

Electron gun

Column/ Electron optics

Optical microscope

EDS detector

Scanning coils

SE,BSE detectors

WDS spectrometers

Vacuum pumps

Faraday current measurement


Key points

UW- Madison Geology 777

Key points

X-rays are dispersed by crystal with only one wavelength (nl) reflected (=diffracted), with only one wavelength (nl) passed to the detector

Detector is a gas-filled (sealed or flow-through) tube where gas is ionized by X-rays, yielding a massive multiplication factor (‘proportional counter’)

X-ray focusing assumes geometry known as the Rowland Circle

Key features of WDS are high spectral resolution and low detection limits


Spectrometers

UW- Madison Geology 777

Spectrometers

An electron microprobe generally has 3-5 spectrometers, with 1-4 crystals in each. Here, SP4 (spectro #4) of our SX51 (#485) with its cover off.

Crystals (2 pairs)

Proportional Counting Tube (note tubing for gas)

PreAmp


X ray path

UW- Madison Geology 777

BSE detectors alternate

X-ray path

Only a small % of X-rays reach the spectrometer. They first must pass thru small holes (~10-15 mmdia; red arrows) in the top of the chamber (above, looking straight up), then thru the column windows (below, SP4).Thus, in our EMP, there are different vacuum regimes in the chamber vs the spectrometer.

Crystal


Wavelength dispersion

BA’ = A’C = d sinq for constructive reinforcement of a wave, the distance BA’ must be one half the wavelength. Thus, 2d sinq = l and by similar geometric construction = nl

UW- Madison Geology 777

Wavelength Dispersion

Of the small % of X-rays that reach the crystal, only those that satisfy Braggs Law will be diffracted out of the crystal.

nl = 2d sinq

Note that exact fractions of l will satisfy the conditions for defraction. Thus, there is a possibility of “higher order” (n=2,3,...11,?) interference in WDS (but there also is the means electronically to discriminate against the interference).


What is n l

BA’ = A’C = d sinq for constructive reinforcement of a wave, the distance BA’ must be one half the wavelength. Thus, 2d sinq = l and by similar geometric construction = nl

UW- Madison Geology 777

nl = 2d sinq

What is nl ?

This is sometimes difficult to comprehend. Assume you have your spectrometer set to one particular position, which means for that sinq and that 2d (let’s say they = 12 Å), there are several possible signals that the spectrometer can tune in to at that position: (1) an x-ray with wavelength of 12 Å, (2) an x-ray of 6 Å, (3) an x-ray of 4 Å, (4) an x-ray of 3 Å, etc --of course, if and only if such x-rays are being generated in the sample.


Lots of crystals

UW- Madison Geology 777

Lots of Crystals

Over the course of the first 30 years of EPMA, ~50 crystals and pseudocrystals have been used.


Crystals and pc lsms

UW- Madison Geology 777

Crystals and PC/LSMs

  • Consider the order of 2d in Braggs Law: sin q varies from .2-.8, and l varies over a wide range from hundreds to fractions of an A. Thus, we need diffractors that cover a similiar range of 2d, from around 1 Å to hundreds of Å. In our SX51, we utilize TAP, PET and LIF crystals for the shorter wavelengths. For longer wavelengths, there are 2 options:

  • pseudocrystals (PCs), produced by repeated dipping of a substrate in water upon which a monolayer (~soap film) floats,progressively adding layer upon layer, or

  • layered synthetic microstructures (LSMs; also LDEs, layered diffracting elements), produced by sputtering of alternating light and heavy elements, such as W and Si, or Ni and C.

  • Both these are periodic structures that diffract X-rays. The LSMs yield much higher count rates; however, peaks are much broader, which have good/bad consequences, discussed later.

  • In reality, people interchange the words PC, LSM, LDE, etc. Cameca uses PC and JEOL uses LDE, for same things.


Crystals and pcs on the uw sx51

UW- Madison Geology 777

Crystals and PCs on the UW SX51

There is a more precise form of Braggs Law, that takes into account refraction, which is more pronounced in the layered synthetic diffractors,

nl = 2d sinq(1-k/n2)

k is refraction factor, n is order of diffraction


Pseudocrystals lsms

Pseudocrystals/LSMs

Goldstein et al, p. 280


Crystals and pcs which to use

UW- Madison Geology 777

Crystals and PCs: Which to use?

  • The EPMA user may have some control over which crystal to use; some element lines can be detected by either of 2 crystals (e.g. Si Ka by PET or TAP, V Ka by PET or LIF), whereas other elements can only be detected by one (e.g. S Ka by PET). Each probe has its own (unique?) set of crystals and the user has to work out the optimal configuration, taking into account concerns such as

  • time and money

  • interferences vs counting statistics (sharper peaks usually have lower count rates)

  • stability (thermal coefficient of expansion)


Crystal comparison

The class project in 2002 was to collect data that will be put in a chart to compare the efficiency of different crystals/ PCs for certain elements.

Crystal comparison

Å

Å

TAP gives a higher count rate, and wider peak for Si Ka vs. using PET

Si Ka on TAP sin q = 27714 P/B= 4862/40=122 FWHM=0.038Å

Si Ka on PET sin q = 81504 P/B= 207/1.3=159 FWHM=0.006Å


Full width half max

UW- Madison Geology 777

Max

4862 cts

Peak (spectral) resolution is described by FWHM

Full Width Half Max

Full Width

Half max

2431 cts

counts

Si Ka on TAP sin q = 27714 P/B= 4862/40=122 FWHM=0.038Å

Å


Wds detector

UW- Madison Geology 777

P10 gas (90% Ar - 10% CH4) is commonly used as an ionization medium. The X-ray enters through the thin window and 3 things can occur: (1) the X-ray may pass thru the gas unabsorbed (esp for high keV X-rays); (2) it may produce a trail of ion pairs (Ar+ + e), with number of pairs proportional to the X-ray energy; and (3) if the X-ray is >3206 eV it can knock out an Ar K electron, with L shell electron falling in its place. There are also 3 possibilities that can result from this new photon:

WDS detector

(3a) internal conversion of the excess energy with emission of Auger electron (which can produce Ar+ + e pairs); (3b) Ar Ka X-ray itself can knock out electron of another Ar molecule, producing Ar+ + e pair; or (3c) the Ar Ka X-ray can escape out thru a window, reducing the number of Ar+ + e pairs by that amount of energy (2958 eV)


Detector amplification

UW- Madison Geology 777

Nominally, it takes 16 eV to produce one Ar+- electron pair, but the real (effective) value is 28 eV. For Mn Ka (2895 eV), 210 ion pairs are initially created per X-ray. However, there is a multiplier effect (Townsend avalanching). For our example of Mn Ka, all 210 electrons are accelerated toward the center anode (which has a positive voltage [bias] of1200-2000v) and produces many secondary ionizations. This yields a very large amplification factor (~105), and has a large dynamic range (0-50,000 counts/sec).

Detector amplification


Detector windows

UW- Madison Geology 777

Detector windows

Thin (polypropylene) windows are used for light X-rays (e.g. those detected by TAP and PC crystals). Since the windows are thin, the gas pressure must be low (~0.1 atm). And being thin windows, some of the gas molecules can diffuse out through them -- so the gas is replenished by having a constant flow. For thicker windows (mylar), 1 atm gas pressure is used (with higher counts resulting). Sealed detectors with higher pressure gas (e.g. Xe) are also used by some.


Wds detector1

UW- Madison Geology 777

WDS detector

The bias on the anode in the gas proportional counter tube needs to be adjusted to be in the proportional range. Too high bias can produce a Geiger counter effect. Too low produces no amplification.


Wds pulse processing

UW- Madison Geology 777

WDS pulse processing

The small electron pulses (charges) generated in the tube are first amplified in the pre-amp (top) located just outside the vacuum on the outside of the spectrometer, then sent to the PHA board where they are amplified (center) and shaped (bottom). Each figure is for one (the same) pulse.

Goldstein et al Fig. 5.10


Ar escape peak

UW- Madison Geology 777

Ar-escape peak

There is a probability that a small number of Ar Ka X-rays produced by the incident X-ray (here, Fe Ka) will escape out of the counting tube. If this happens, then those affected Fe Ka X-rays will have pulses deficient by 2958 eV. Fig 7.8 is an unusual plot of this (for teaching purposes); what is normally seen is the Pulse Height distribution where the pulses are plotted on an X-axis of a maximum of 5 or 10 volts.

Reed, 1993, p. 90


Electron probe microanalysis epma

UW- Madison Geology 777

Actual PHA plot for Fe Ka: note the Ar-escape peak


Electron probe microanalysis epma

UW- Madison Geology 777

Actual PHA plot for Si Ka: there is no Ar-escape peak.

WHY?


Electron probe microanalysis epma

UW- Madison Geology 777

Higher order reflections

Recall nl = 2d sinq. Higher order reflections are possible in your specimens, and must be avoided to prevent errors in your analyses.

Reed (1993) reports that LIF can show a strong second order peak, up to 10% of the first order peak.

In 1999, the 777 class examined the higher order reflections of Cr Ka lines. On PET, 2nd and 3rd order peaks were present, and decreased ~an order of magnitude with increasing n. On LIF, up to the 8th order peak was present. Here, the intensity of the odd numbered orders was less than the following even order, e.g. the 5th order Ka had 30% fewer counts than the 6th order line.

Differential mode of pulse height analysis (PHA) may be used to ignore counting higher order X-rays.


Electron probe microanalysis epma

window

baseline

UW- Madison Geology 777

Pulse Height Analysis

If a higher order reflection falls upon a peak (or unavoidable background) position, the analyst has the option of using Pulse Height Analysis, i.e. setting up a window and not counting any X-rays that have energies greater than the window’s upper limit.

There is a lower limit (baseline, usually 0.5 volts). The window stretches above it (here 4 volts),

and thus a second order reflection should have a pulse height around 6 volts, and would not be counted.

There are some situations where operation in differential mode can

introduce errors (pulse height depression), so the normal mode is integral. And

if differential is used, it is good to do some tests first to get a feeling of how

comprehensive its filtering action is.


Integral vs differential pha

UW- Madison Geology 777

Integral vs Differential PHA

Analysis of ‘light elements’ such as C is complicated because of the long wavelength (44 Å) which means that higher order reflections of many elements can interfere. At top, where the PHA is set to the “count everything” integral mode, the 3rd order reflection of Ni La1 falls very close to C Ka and adds some to C peak counts. Note also the 2 and 3 order Fe L lines. By setting the detector electronics to the discrimination mode (“differential”), bottom, the higher order lines are strongly (but not totally) suppressed.

Spectrometer scans

Goldstein et al, p.507-8


Electron probe microanalysis epma

UW- Madison Geology 777

Putting it all together: Spectrometer = Crystal + Detector moving in a highly choreographed danceDance floor =Rowland Circle


Rowland circle

C

B

A

UW- Madison Geology 777

Rowland Circle

For most efficient detection of X-rays, 3 points must lie upon the focusing circle known as the Rowland Circle. These points are

the beam impact point on the sample (A),

the active central region of the crystal (B), and

the detector -- gas-flow proportional counter (C).


Rowland circle1

UW- Madison Geology 777

The loci of 3 points must always lie on the Rowland Circle. Starting at the top position (blue), there is a shallow angle of the X-ray beam with the analyzing crystal. To be able to defract a longer wavelength X-ray on the same spectrometer, the crystal travels a distance further out, and effectively the (green) Rowland Circle “rolls”, pinned by the beam-specimen interaction point.

Rowland Circle


  • Login