Ventricular Assist Devices

Ventricular Assist Devices. Brian Schwartz, CCP February 25, 2003. Criteria for Ventricular Assist Devices. Cardiac Index < 2.0 L/m2/min SVR > 2,100 dyn/sec/cm2 Systolic Pressure < 80 mmHg Atrial Pressure > 20 mmHg Assisted (diuretics) Urine Output < 20 ml/hr Metabolic Acidosis.

roth-golden
Download Presentation

Ventricular Assist Devices

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Ventricular Assist Devices Brian Schwartz, CCP February 25, 2003

  2. Criteria for Ventricular Assist Devices • Cardiac Index < 2.0 L/m2/min • SVR > 2,100 dyn/sec/cm2 • Systolic Pressure < 80 mmHg • Atrial Pressure > 20 mmHg • Assisted (diuretics) Urine Output < 20 ml/hr • Metabolic Acidosis

  3. Criteria for Ventricular Assist Devices • Failure to separate from CPB • Irreversible cardiac injury

  4. Short Term Support • Cardiopulmonary bypass Resuscitation • 15 % survival rate • Immediately able to support patient • Need to fully heparinize patient • CPR • Provides support temporarily

  5. Devices Used to Assist the Ventricle(Moderate Setting) • Intra-aortic balloon pump • Cardiopulmonary Support (CPS) • Centrifugal Pump “Bio-Head” • Abiomed (BVS-5000) Bi-Vad

  6. Intra-aortic Balloon Pumps • The least complicated means of circulatory assistance • Effects of IABP • Augmentation of Diastolic Pressure • Decrease Afterload • Decrease myocardial oxygen consumption • Augments C.O. by 10% (500-800 cc/min) • Relatively inexpensive

  7. Intra-aortic Balloon Pumps(Indications) • Cardiogenic shock following MI • Unstable Angina • Left Main Disease • Ventricular Dysrhythmias • Septic Shock

  8. Intra-aortic Balloon Pumps(Contraindications) • AI • Aortic Aneurysm • Severe Femoral Disease

  9. Cardiopulmonary Support (CPS) • Percutaneous insertion • Need oxygenator and heat exchanger • Cannulate both femoral artery and femoral vein • Needs continuous monitoring, therefore very labor intensive • Maximum support…48 hours

  10. Centrifugal Pump • Easy to prime and set up • Requires continuous monitoring • Kinetic assisted venous drainage • ACT’s around 180-200 seconds • Moderate cost

  11. Abiomed • Quick set-up • Minimal bedside monitoring • Supports large children and adults • Flow rates up to 5 L/Min • Maximum use….1 week • Patients are not mobile • High cost

  12. Long Term Devices for Ventricular Support • TCI ( Heartmate IP 1000) Pneumatic • LVAD only • TCI (VE) Vented Electric • LVAD only • Novacor (N 100P) Electric • LVAD only • Thoratec Pneumatic • LVAD, RVAD, Bi-VAD

  13. Heartmate Pneumatic LVAD • Allows blood flows to exceed 10 liters per minute • Inserted during CPB • Minimum BSA required…1.7 • Very costly to insert

  14. Heartmate Vented Electric LVAD • Allows flows exceeding 10 liters • Need CPB for placement • BSA requirement…greater than 1.7 • Patients are able to go home • Minimal anti-coagulation • High cost

  15. Total Artificial Heart • CardioWest ( C-70 ) Pneumatic total artificial heart • C.O. is approximately 7.0 L/M • BSA>1.7 • Need CPB for implant • Native heart not excised • Need Anti-coagulation • Patient in-house but mobile

  16. Total Artificial Heart • Abiomed’s total artificial heart • Still in clinical trials • First patient lasted several months on device • If successful, will save hundreds of thousands of live because there will be no waiting like the transplant list

  17. Signs indicating Left Ventricular Failure • Decreased contractility • Elevated left ventricular filling pressures • Elevated pulmonary capillary wedge pressures • Decrease pulmonary oxygenation

  18. Signs of Right Ventricular Failure • Cardiac Index less than 1.8 L/min/m2 • Aortic pressure less than 90 mmHg • Atrial pressure greater than 20 mmHg • Pulmonary capillary wedge pressure less than 10 mmHg

  19. Cannulation sites for LVAD’s • Inlet • Left atrium • Left ventricle • Left superior pulmonary vein • Outlet • Aorta

  20. Cannulation sites for RVAD’s • Inlet • Right atrium • Outlet • Pulmonary artery

  21. Heparin Management in Patients with VAD’s • ACT’s are maintained around 180-200 seconds…as long as the cardiac output is above three liters per minute • ACT’s are maintained above 300 seconds while the cardiac output is below three liters per minute • WHEN IS THIS IMPORTANT???????

  22. Discontinuing a VAD • If possible, the heart is allowed to rest for 48-72 hours • Weaning the patient off the assist devices is then performed. ( Patient needs to have a native C.I. of at least 2.2 L/min/m2) • ACT’s are increased to compensate for the low flows • If the patient tolerates the low flows, he/she is separated from the assist devices and the cannulae are removed

  23. VAD Protocol for Your Institution • Know your protocol • Know your equipment • Be able to predict those patients at risk • Be able to prime in an orderly fashion • Be prepared to assist the surgeons • Help educate all staff working with the patient