1 / 21

Introducing the New Range of PC-based Ion-Analysers

Introducing the New Range of PC-based Ion-Analysers. ELIT 9801 – ELIT 9804 – ELIT 9808. From NICO 2000 Ltd. Document No. IAP 05-03-c.ppt compiled by Heinz Kreuzberg, Shree Barve, Ameet Barve & Chris Rundle London, UK - May/June 2003. A Multitude of Applications for Ion Analysers.

Download Presentation

Introducing the New Range of PC-based Ion-Analysers

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Introducing the New Range of PC-based Ion-Analysers ELIT 9801 – ELIT 9804 – ELIT 9808 From NICO 2000 Ltd. Document No. IAP 05-03-c.ppt compiled by Heinz Kreuzberg, Shree Barve, Ameet Barve & Chris Rundle London, UK - May/June 2003

  2. A Multitude of Applications for Ion Analysers AGRICULTURE COSMETICS INDUSTRIAL PRODUCTION WATER MONITORING WASTE WATER TREATMENT FISHERIES POLLUTION CONTROL BLOOD TESTING

  3. Contents • Introduction to Ion Measurement • pH Measurement • Product Description • Features – Benefits – Advantages • Applications • Hardware Details • Software Details • Using the ISE / pH Ion-Analyser • Using the 4- & 8- Channel Analysers • The NICO 2000 Product Range

  4. Introduction to IonMeasurement An Ion-Analyser or Ion Meter is required to convert the electrical signal from the ISE into a relevant unit of concentration (ppm or mol/litre). Ion-selective Electrodes (ISEs) allow us to determine the concentration of a specific ion in aqueous (& in rare cases non-aqueous) solutions. Measurements with ISEs are temperature sensitive. An exact temperature compensation for measurements taken at different temperatures is not possible. Therefore, calibration and sample measurement should be done within the same temperature range (+/- 10). The measurement made is not an absolute one but rather a comparison of a sample measurement with a calibration measurement (calibration curve) The calibration procedure uses Ion-Standards (solutions with a precisely defined ion concentration) and specifically diluted calibrating solutions. Usually, 1:10 serial dilutions from a 1000 ppm ion-standard.

  5. Typical ISE Calibration Graph NB: X-axis units are the logarithm of the Molar Activity of the Ion

  6. ISE Measurement with ELIT ISE/pH Ion-Analyser For educational purposes, sample results are plotted on the Calibration Graph

  7. pH Measurement with ELIT ISE/pH Ion Analyser 2-Channel pH Display “Virtual Instrumentation” • pH is a measure of the Acidity or Alkalinity of a solution. • Actually measures the concentration of Hydrogen Ions. • Is based on a logarithmic relationship between the sensor • signal (mV) & the Hydrogen ion concentration (mol/L). • Calibration standards for pH measurements are called • Buffers due to their ability to keep the pH value stable when small amounts of acid or alkaline substance added. • The most common Buffers are pH 4, pH 7 & pH 10. • Like all Ion-selective measurements, pH measurements are influenced by temperature changes. Therefore, instruments have manual or automatic temperature compensation facilities. • To simplify the handling of results of pH measurements, the 14 decades of hydrogen ion concentration are converted into a linear (pH) scale from 0 to 14 using the Soerensen equation: pH value = negative Logarithm of the Hydrogen ion concentration. pH 0 is extremely acidic, pH14 is extremely alkaline & pH 7 is neutral.

  8. Typical pH Values

  9. Product Description • 3 Types of Ion-Analyser are available: ELIT 9801 = ISE/pH Ion-Analyser ELIT 9804 = 4-Channel Ion-Analyser ELIT 9808 = 8-Channel Ion-Analyser • They are powerful tools for measuring & monitoring Ion Concentrations, pH, Redox (ORP) & Temperature • They contain high quality Electrode Amplifiers, High Resolution Analogue-to-Digital converters, a Powerful Microprocessor & a state-of-the-art opto-isolated Serial interface • They areprimarily PC-based Instruments - but the hardware is also compatible with any computer with a serial port (RS232) • They are powered by sophisticated software which runs on any Microsoft Windows Operating System; - Windows 98 or later (2000, ME, XP). • Menu-driven software guides the user reliably through the measurement process & supports GLP effectively. • They can perform Single Measurements,on a series of samples or Continuous Monitoring of changes in individual solutions . • Acquired data is displayed online in Tables and Graphs. • Data are stored as ASCII text files which are easily Exportable to other software packages such as Excel, Labview etc.

  10. Features & Benefits • Compared to many other analytical methods, Ion-selective measurements are remarkably analyte -specific,very cost effective & suitable for continuous monitoring of time dependent changes in analyte concentration. • ELIT Ion-Analysers incorporate the Latest Technology & enable the user to realise the full potential of Ion-Selective measurements. • ELIT Analyser software ensures error-free data acquisition, storing & processing, and provides the user with reliable and clear guidance during the measuring process. • ELIT Multi-channel Analysers connect up to 8 Electrodes to one computer. • During Continuous Monitoring, the Sampling time can be varied over a wide range. • An Averaging Algorithm smoothes unstable and noisy measurements. • ELITISE/pH Analyser:- - Enables the detection limit to be significantly lowered by using three different algorithms to obtain optimal results in the non-linear range of the calibration curve. - Allows the user to select the most suitable method from: Direct Potentiometry; Standard-Addition; Sample-Addition.

  11. Worldwide Applications • Pollution Control • pH of acid rain, soil, surface water • Contamination of surface water and ground water with ammonium and nitrate • Contamination of waste water with Cyanide, Cadmium, Mercury & Copper • Pharmaceutical & Cosmetics Industry • Fluoride & pH of Tooth Paste • pH of Hair shampoo • Many other Ions determined in Pharmaceutical and Cosmetic Research & Development • Water Quality Monitoring • pH, Calcium & various ions in surface water (rivers, lakes & ponds) • Ammonium in ground water • pH, Ammonium & Nitrate in fish tanks & fish ponds The ELIT Range of Ion-Analysers • Industrial Production • Salinity & pH of Boiler feed water • Cyanide in Plating baths • Process-specific Ions • Food Quality Control • Nitrate & Nitrite in meat & vegetables • Chloride, Sodium, Nitrate & Nitrite in baby food • Fluoride & Calcium in milk & other milk products • Cadmium in fish • Waste Water Treatment • pH, Ammonium, Nitrate & other Ions used during processing • Research & Education • pH, Redox in Higher Education • All types of Ion Measurements & Continuous Monitoring in Schools and Colleges • Agriculture & Fishery • Soil & Fertilizer Analysis for Nitrate, Ammonium & Potassium to optimize the use of fertilizer (cost savings & highest yield) • Dissolved Oxygen and pH in Ponds for Fish Breeding • Medical Diagnosis • & Hygiene Control • Potassium in urine • Contamination Control for various Ions

  12. Solid all-ABS plastic enclosure ensures good resistance to corrosive spills and atmosphere. Plug in AC / DC Power Adapters with country-specific plugs. Optional small Battery Power Pack for use in the field. BNC connectors for Electrodes. 3.5mm connectors for Temperature sensors. 2mm connectors for external Reference electrodes (if required). Supplied with data cable and adapter for connection to 25-way or 9-way serial port on the computer(RS232). Optional special “active” cable for connecting to USB ports on the computer Stainless steel temperature sensors with optional Teflon coating. Electrode Amplifiers guarantee the best results with all types of ion-selective & pH electrodes (standard, special & micro). Powerful Microprocessor allows high resolution measurements & wide range of sampling times for continuous monitoring. Signal ground completely isolated, therefore no Ground Loops. Hardware Details

  13. State-of-the-art, User-friendly “Virtual Instrumentation” using the power of modern computing for operator convenience, efficiency and data security/protection. Storage of up 100,000 data points for one continuous measurement (ELIT 9804 & 9808). Easy to integrate into Laboratory Management Systems (LIMS). Acquired Data well presented in Graphical & Table format. Very Flexible 2D & 3D Graphical Data Display (ELIT 9804 & 9808) Online Data Display Selectable Smoothing of noisy measurements by selecting the Averaging number. Good suppression of 50 Hz & 60 Hz interference Easy-to-implement Communication Protocol if Ion-Analysers are connected to operating systems other than MS Windows Software Details

  14. Using the ISE / pH Ion-Analyser

  15. Calibration Procedure

  16. Calibration Curve

  17. Incremental ISE Determination

  18. Typical Graphical display from 4 or 8-Channel AnalyserNB: Options Menu allows plotting of mV, Mol/L, pH or Temperature and any combination of Channels; individually or together. Graphic Tools enable wide range of options for graph styles,e.g. 3D, bar chart, labels, axes, ticks, etc.

  19. 3D Display of Data

  20. 3D Bar Chart Data Display

  21. Other NICO 2000 Products

More Related