TURKEY REPUBLIC MINISTRY OF ENVIRONMENT AND FORESTRY TURKISH STATE METEOROLOGICAL SERVICE DIRECTORY OF METEOROLOGICAL INSTRUMENTS AND CALIBRATION PowerPoint PPT Presentation


  • 204 Views
  • Uploaded on
  • Presentation posted in: General

TURKEY REPUBLIC MINISTRY OF ENVIRONMENT AND FORESTRY TURKISH STATE METEOROLOGICAL SERVICE DIRECTORY OF METEOROLOGICAL INSTRUMENTS AND CALIBRATION. ECO CALIBRATION CENTER. Calibration Center.

Download Presentation

TURKEY REPUBLIC MINISTRY OF ENVIRONMENT AND FORESTRY TURKISH STATE METEOROLOGICAL SERVICE DIRECTORY OF METEOROLOGICAL INSTRUMENTS AND CALIBRATION

An Image/Link below is provided (as is) to download presentation

Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server.


- - - - - - - - - - - - - - - - - - - - - - - - - - E N D - - - - - - - - - - - - - - - - - - - - - - - - - -

Presentation Transcript


Slide1 l.jpg

TURKEY REPUBLIC

MINISTRY OF ENVIRONMENT AND FORESTRY

TURKISH STATE METEOROLOGICAL SERVICE

DIRECTORY OF METEOROLOGICAL INSTRUMENTS AND CALIBRATION

ECO CALIBRATION

CENTER

CalibrationCenter - Ankara 2010


Slide2 l.jpg

Calibration Center

At the Calibration Center: Automated Weather Observing System's sensors that is installed to all around the Türkiye are calibrated by applying the 17025 standards.

Calibration Center is accreditedfor the calibrations of tempetature, relative humidity, air pressure and wind speed sensors.

Calibration Center - Ankara 2010


Slide3 l.jpg

TSMS’s Calibration Center acts as Economic Cooperation Organisation (ECO) calibration center. It would provide training facilities for the member states.

Member States:

1- The Islamic Republic of Afghanistan .

2- The Republic of Azerbaijan.

3- The Islamic Republic of Iran .

4- The Republic of Kazakhstan .

5- The Kyrgyz Republic .

6- The Islamic Republic of Pakistan .

7- The Republic of Tajikistan .

8- The Republic of Turkey .

9- The Republic of Turkmenistan .

10-The Republic of Uzbekistan .

Calibration Center - Ankara 2010


Slide4 l.jpg

CALIBRASTION CENTER ORGANIZATIONAL CHART

Calibration Center - Ankara 2010


Slide5 l.jpg

DIRECTORY OF METEOROLOGICAL INSTRUMENTS and CALIBRATION

PRECIPITATION

WIND

Calibration Center operates under

Meteorological Instruments and Calibration Division.

TEMP&HUM

SOLAR RADIATION

Mr. Nur DOĞAN

Manager of Meteorological Instruments

and Calibration Division

PRESSURE

Calibration Center - Ankara 2010


Slide6 l.jpg

QUALITY MANAGER

Calibration Center of all activities;providing work in accordance with the rules of the17025 standard, providing continuous development of Laboratory and providing relations with the TURKAK duty and authorized to Institution to represent in TURKAK .

Mr. Yusuf Özcan CANDEMİR

QUALITY MANAGEMENT

Calibration Center - Ankara 2010


Slide7 l.jpg

CHIEF OF CALIBRATION LABORATORIES

  • Temperature & Humidity Calibration Laboratory,

  • Pressure Calibration Laboratory,

  • Wind Speed Calibration Laboratory,

  • Radiation Calibration Laboratory,

  • Precipitation Calibration Laboratory.

Mr. Tekin SOYLU

Seniorchief of theLaboratories

Calibration Center - Ankara 2010


Slide8 l.jpg

CALIBRATION CENTER

PERSONNEL CHART

Calibration Center - Ankara 2010


Slide9 l.jpg

INSTRUMENT ACCEPTANCE & REGISTRATION AND ARCHİVE

Instruments accepted to the Calibration Center are registered and related documents are archivedin accordance with the rules of 17025 standards.

Calibration Center - Ankara 2010


Slide10 l.jpg

CALIBRATION CENTER

ENVIRONMENTAL CONDITIONS OF THE LABRATORIES

Environmental conditions of the Laboratories are controlled by using precision air-conditioner. The measured data of temperature and humuidity are automatically displayed and stored.

Calibration Center - Ankara 2010


Slide11 l.jpg

CALIBRATION CENTER

ENVIRONMENTAL CONDITIONS OF THE LABRATORIES

Devices to control the laboratory environmental conditions:

Fluke Hart Scientific 1620A

Measuring, displaying and recording.

Liebert Hiross HPM Precision Air conditioner

Calibration Center - Ankara 2010


Slide12 l.jpg

TEMPERATURE & HUMIDITY LABORATORY

  • The thermometer typesto be calibrated:

  • Liquid glass thermometers,

  • Resistance thermometers,

  • Analog output thermometers,

  • Thermometerswithdisplay.

All types of relative humidity sensors are calibrated.

Calibration Center - Ankara 2010


Slide13 l.jpg

TEMPERATURE & HUMIDITY LABORATORY

Scope of Accreditation

Calibration Center - Ankara 2010


Slide14 l.jpg

TEMPERATURE & HUMIDITY LABORATORY

Reference devices for relative humidity calibration:

Thunder Scientific 2500 Humidity Generator.

Operation range: -10˚C / +70˚C

10%RH / 90%RH

Vaisala HMP76

Calibration Center - Ankara 2010


Slide15 l.jpg

TEMPERATURE & HUMIDITY LABORATORY

Climate provider for Relative Humidity Calibration:

Weiss SB22 500

Climate Chamber.

Operation Range:

-50˚C / +50˚C

10%RH / 90%RH

Calibration Center - Ankara 2010


Slide16 l.jpg

TEMPERATURE & HUMIDITY LABORATORY

Reference devices for Temperature Calibration:

Isotech 909Q Standard Platinum Resistance Thermometer

Hart Scientific 5615 Platinum Resistance Thermometer

Calibration Center - Ankara 2010


Slide17 l.jpg

TEMPERATURE & HUMIDITY LABORATORY

Climate provider for Temperature Calibration:

Weiss SB22 500

Climate Chamber.

Operation Range: -50˚C / +50˚C

10%RH / 90%RH

Fluke 7381 Liquid bath.

Operation range:

-80˚C /+110˚C

Calibration Center - Ankara 2010


Slide18 l.jpg

TEMPERATURE & HUMIDITY LABORATORY

Secondary devices for Temperature & Humidity Laboratory:

Fluke PRT measuring unit, Zeno 3200 Data Logger and Fluke 8846A 6 ½ digit multimeter are used to measure the data.

Calibration Center - Ankara 2010


Slide19 l.jpg

TEMPERATURE & HUMIDITY LABORATORY

Calibration at The Triple Point of Water (TPW)

and

Ice Point Calibration :

Calibration Center - Ankara 2010


Slide20 l.jpg

PRESSURE CALIBRATION LABORATORY

  • Pressure sensors typesto be calibrated:

  • Frequency output barometers,

  • Digital output barometers,

  • Barometerswith digital display,

  • Mechanical barometers.

Calibration Center - Ankara 2010


Slide21 l.jpg

PRESSURE CALIBRATION LABORATORY

Scope of Accreditation

Calibration Center - Ankara 2010


Slide22 l.jpg

PRESSURE CALIBRATION LABORATORY

Reference device for Pressure Calibration:

Parosceintific 785 digiquartz barometer

Calibration Center - Ankara 2010


Slide23 l.jpg

PRESSURE CALIBRATION LABORATORY

Climate providers for Pressure Calibration:

Fluke PPC4-ui A100K Calibrator.

Dry air are used (Azoth gas).

Pressure Chamber

Calibration Center - Ankara 2010


Slide24 l.jpg

PRESSURE CALIBRATION LABORATORY

Secondary devices of Pressure Laboratory:

Druck DPI 145

Measurement device for frequency output barometer.

Measurement device for digital output barometer.

Calibration Center - Ankara 2010


Slide25 l.jpg

PRESSURE CALIBRATION LABORATORY

Secondary devices of the Pressure Laboratory:

Paroscientific 765 digiquartz barometer.

Ruska 6200 barometer.

Calibration Center - Ankara 2010


Slide26 l.jpg

WIND SPEED CALIBRATION LABORATORY

  • WindSpeed sensor typesto be calibrated (Cup, hotwire, ultrasonic, pitotand propelleretc.):

  • Electronics wind speed sensors,

  • Mechanical wind speed sensors.

Calibration Center - Ankara 2010


Slide27 l.jpg

WIND SPEED CALIBRATION LABORATORY

Scope of Accreditation

Calibration Center - Ankara 2010


Slide28 l.jpg

WIND SPEED CALIBRATION LABORATORY

Reference devicesfor Wind Speed Calibration:

Furness Control FCO510 Micromanometer and Pitot tube.

Thies 2D 4.3810.00.340 Ultrasonic wind speed sensor

Calibration Center - Ankara 2010


Slide29 l.jpg

WIND SPEED CALIBRATION LABORATORY

Medium provider for Wind Speed Calibration:

Wind Speed Tunnel is the Göttinger type, closed loop and open test area ( 0 – 40 m/s).

Calibration Center - Ankara 2010


Slide30 l.jpg

WIND SPEED CALIBRATION LABORATORY

Devices of the Wind Speed Calibration Laboratory:

Hotwire Testo 435 Wind Speed Sensor.

Druck PTX7517-1 Pressure (600–1100mbar)

PR 5333A Temperature (0-40ºC) sensors.

Calibration Center - Ankara 2010


Slide31 l.jpg

WIND SPEED CALIBRATION LABORATORY

Devices of the Wind Speed Calibration Laboratory:

Vaisala WMS302M and WAA151, Thies Clima 4.3303.22.007, Qualimetrics ECN4738, Siap handel type, Vector İnst. A100M/HE-1 and Lastem DN002 Wind Speed sensors are used.

Calibration Center - Ankara 2010


Slide32 l.jpg

RADIATION CALIBRATION LABORATORY

Pyranometers are calibrated at the Radiation Calibration Laboratory under the constant irradiance and new sensitivity coefficients for the calibrated sensors are calculated.

Calibration Center - Ankara 2010


Slide33 l.jpg

RADIATION CALIBRATION LABORATORY

Scope of the Calibration

Calibration Center - Ankara 2010


Slide34 l.jpg

RADIATION CALIBRATION LABORATORY

Reference device for Pyranometer Calibration:

Kipp&zonen CMP22 Pyranometer

Calibration Center - Ankara 2010


Slide35 l.jpg

RADIATION CALIBRATION LABORATORY

Medium provider for Pyranometers Calibration:

Kipp&Zonen Calibration Facility for Pyranometers.

Indoor Pyranometer Calibration System under constant irradiance light (0 - 700W/m²).

Calibration Center - Ankara 2010


Slide36 l.jpg

RADIATION CALIBRATION LABORATORY

Devices of the Pyranometers Calibration Laboratory:

Kipp&Zonen CMP21 Pyranometer

Keithley 2000 6 ½ digit multimeter

Calibration Center - Ankara 2010


Slide37 l.jpg

PRECIPITATION CALIBRATION LABORATORY

  • Precipitation (Rain Gauge) sensors typesto be calibrated:

  • Tipping bucket rain gauges,

  • Weighing type rain gauges.

Calibration Center - Ankara 2010


Slide38 l.jpg

PRECIPITATION CALIBRATION LABORATORY

Scope of calibration

Calibration Center - Ankara 2010


Slide39 l.jpg

PRECIPITATION CALIBRATION LABORATORY

Reference devices for Rain Gauges Calibration:

Mettler Toledo XP4002S

Precision Electronic Balances

800 gr. / 4100 gr.

Baykon F1 Class Weigths

1 gr / 2 kg

Calibration Center - Ankara 2010


Slide40 l.jpg

PRECIPITATION CALIBRATION LABORATORY

Medium provider for Rain Gauges Calibration:

Precipitation simulation is prepared by using the Watson Marlow 520Du peristaltic pump which calculates the necessary amount of water for the desired amount of Rain İntensity.

Calibration Center - Ankara 2010


Slide41 l.jpg

PRECIPITATION CALIBRATION LABORATORY

Secondary devices of the Precipitation Laboratory:

Zeno 3200 Data Logger

Caps and tubeswith different type and dimensions.

Calibration Center - Ankara 2010