An introduction to Ultraviolet/Visible Absorption Spectroscopy

An introduction to Ultraviolet/Visible Absorption Spectroscopy. Lecture 24. Instrumentation Light source  - selector Sample container Detector Signal processing Light Sources (commercial instruments) D 2 lamp (UV: 160 – 375 nm) W lamp (vis: 350 – 2500 nm).

jabari
Download Presentation

An introduction to Ultraviolet/Visible Absorption Spectroscopy

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. An introduction to Ultraviolet/Visible Absorption Spectroscopy Lecture 24

  2. Instrumentation • Light source •  - selector • Sample container • Detector • Signal processing • Light Sources (commercial instruments) • D2 lamp (UV: 160 – 375 nm) • W lamp (vis: 350 – 2500 nm)

  3. SourcesDeuterium and hydrogen lamps (160 – 375 nm) D2 + Ee → D2* → D’ + D’’ + h

  4. Deuterium lamp UV region (a) A deuterium lamp of the type used in spectrophotometers and (b) its spectrum. The plot is of irradiance Eλ (proportional to radiant power) versus wavelength. Note that the maximum intensity occurs at ~225 m.Typically, instruments switch from deuterium to tungsten at ~350 nm.

  5. Visible and near-IR region • A tungsten lamp of the type used in spectroscopy and its spectrum (b). Intensity of the tungsten source is usually quite low at wavelengths shorter than about 350 nm. Note that the intensity reaches a maximum in the near-IR region of the spectrum.

  6. The tungsten lamp is by far the most common source in the visible and near IR region with a continuum output wavelength in the range from 350-2500 nm. The lamp is formed from a tungsten filament heated to about 3000 oC housed in a glass envelope. The output of the lamp approaches a black body radiation where it is observed that the energy of a tungsten lamp varies as the fourth power of the operating voltage.

  7. Tungsten halogen lamps are currently more popular than just tungsten lamps since they have longer lifetime. Tungsten halogen lamps contain small quantities of iodine in a quartz envelope. The quartz envelope is necessary due to the higher temperature of the tungsten halogen lamps (3500 oC). The longer lifetime of tungsten halogen lamps stems from the fact that sublimed tungsten forms volatile WI2 which redeposits on the filament thus increasing its lifetime. The output of tungsten halogen lamps are more efficient and extend well into the UV.

  8. Tungsten lamps (350-2500 nm) Why add I2 in the lamps? W + I2 → WI2 • Low limit: 350 nm • Low intensity • Glass envelope

  9. 3. Xenon Arc Lamps Passage of current through an atmosphere of high pressured xenon excites xenon and produces a continuum in the range from 200-1000 nm with maximum output at about 500 nm. Although the output of the xenon arc lamp covers the whole UV and visible regions, it is seldom used as a conventional source in the UV-Vis. The radiant power of the lamp is very high as to preclude the use of the lamp in UV-Vis instruments. However, an important application of this source will be discussed in luminescence spectroscopy which will be discussed later.

  10. Sample Containers Sample containers are called cells or cuvettes and are made of either glass or quartz depending on the region of the electromagnetic spectrum. The path length of the cell varies between 0.1 and 10 cm but the most common path length is 1.0 cm. Rectangular cells or cylindrical cells are routinely used. In addition, disposable polypropylene cells are used in the visible region. The quality of the absorbance signal is dependent on the quality of the cells used in terms of matching, cleaning as well as freedom from scratches.

  11. Instrumental Components • Source •  - selector (monochromators) • Sample holders • Cuvettes (b = 1 cm typically) • Glass (Vis) • Fused silica (UV+Vis) • Detectors • Photodiodes • PMTs

  12. Types of Instruments Instrumental designs for UV-visible photometers or spectrophotometers. In (a), a single-beam instrument is shown. Radiation from the filter or monochromator passes through either the reference cell or the sample cell before striking the photodetector.

  13. 1. Single beam • Place cuvette with blank (i.e., solvent) in instrument and take a reading  100% T • Replace cuvette with sample and take reading  % T for analyte (from which absorbance is calc’d)

  14. Most common spectrophotometer: Spectronic 20. • On/Off switch and zero transmission adjustment knob • Wavelength selector/Readout • Sample chamber • Blank adjustment knob • Absorbance/Transmittance scale

  15. End view of the exit slit of the Spectronic 20 spectrophotometer pictured earlier

  16. Single-Beam Instruments for the Ultraviolet/Visible Region

  17. Inside of a single-beam spectrophotometer connected to a computer. • Single-Beam Computerized Spectrophotometers

  18. 2. Double beam (most commercial instruments) • Light is split and directed towards both reference cell (blank) and sample cell • Two detectors; electronics measure ratio (i.e., measure/calculate absorbance) • Advantages: • Compensates for fluctuations in source intensity and drift in detector • Better design for continuous recording of spectra

  19. General Instrument Designs Double Beam: In - Space Needs two detectors

  20. General Instrument Designs Double Beam: In - Time

  21. Merits of Double Beam Instruments • Compensate for all but the most short term fluctuation in radiant output of the source • Compensate drift in transducer and amplifier • Compensate for wide variations in source intensity with wavelength

  22. Dual Beam Instruments