An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. A few thoughts on gamma densitometry of powder jet Goran Skoro 27 October 2008

  2. *Here: ‘High’ activity gamma source (75-Se, 192-Ir, 137-Cs); Detector: NaI(Tl) or BGO; MCA with possibility to work in multiscaler regime with ms dwell time. Density of tungsten powder jet? Can the standard* gamma spectrometry equipment be used for this purpose? Is it possible to map the jet density profile? g DET 2

  3. Gamma densitometry The jet velocity is about 10 m/s, so it would be nice to have the counting time at the level of 1 ms –> the density map resolution will be about 1 cm. where N (N0) is the counting rate* with (without) jet, r is the density, NAv is Avogadro number, A is atomic mass of tungsten, s is (known) cross-section for gamma interaction with tungsten atom, d is shown above. Idea: If the jet diameter (size) is known, collimated gamma radiation can be precisely pointed… …and the difference in detector counting rate with/without jet will give us the density, because: Because of this we need a fast Multiscaler. *This is the counting rate of the spectral line which corresponds to the characteristic gamma energy –> the spectral analysis is needed before counting. g d DET Jet cross-section 3

  4. Gamma densitometry – gamma sources, detectors, geometry,… Possible gamma sources: 75-Se: T1/2 = 120 days, Eg = 265 keV, bf = 60%; 192-Ir: T1/2 = 74 days, Eg = 317 keV, bf = 81%; 137-Cs: T1/2 = 30 years, Eg = 661 keV, bf = 100%. Possible detectors: NaI(Tl), BGO,… 2x2” NaI(Tl) good enough Important: The difference N0-N must be big enough in order not to be interpreted as a statistical fluctuation (proportional to N). On the other hand, source activity has to be as low as possible (for obvious reasons). This means that sensitivity of the density measurements will have the lower limit. What if this minimal value of ‘observable’ density is 10% of W density (and d = 1 cm)? Possible geometry: d = 2 cm; l = 10 cm; ‘spot size’ -> r = 1 mm d l For comparison only! What is the minimal gamma source activity? Results -> next page Source DET 4

  5. Gamma densitometry – an example If > 1, the difference in counting rate is a result of absorption of gammas in tungsten jet, not a statistical fluctuation. For example, we need a 137-Cs source with activity of > 1 Ci to measure the density of 10% of W density. In the case of 192-Ir or 75-Se it is enough to have ~ 100 mCi source. r = 10% of W density, d = 1 cm Statistical significance = (N0 -N)/3s Source activity [Ci] 5

  6. 137-Cs: No! (of course); high activity needed, long half-life, high (~500 kHz) counting rate* 75-Se: OK! Counting rate below 50 kHz for 100 mCi source 192-Ir: OK! Counting rate below 50 kHz for 100 mCi source Gamma densitometry – which gamma source? 192-Ir preferable because of the shorter half-life and the price! ** £1100 for 20 Ci 192-Ir, £2000 for 20 Ci 75-Se. We don’t need such a strong source (it seems that the upper limit in this case is around 1 Ci*) so maybe a borrowing is a good idea… To conclude: In general, this approach to measure the density looks possible. More detailed optimisation needed… *Another limiting factor (from the detector point of view). NaI(Tl) starts to lack in the performance at counting rates between 200-500 kHz (dead time, pile-up, etc. corrections have to be performed). ** www.ndtequipment.co.uk, www.ndt-es.co.uk 6

  7. 7 Update I 11 January 2009

  8. Gamma densitometer – a scheme Detector and signal processing system Signal processing system Multichannel analyser (MCA) can be used to measure a counting rate of corresponding photopeak or, if we decide to use Single channel analyser (SCA) + Multiscaler for this, to monitor the gamma spectrum (to control the possible drift of the amplifier gain, to see the effect of the pile-up, etc…) NaI Scintillator Source Shielding & collimator Jet section Detector Multiscaler High voltage PM tube Preamplifier Amplifier MCA SCA 8