An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. A design study of a Cryogenic High Accurate Derotator.

  2. A design study of a Cryogenic High Accurate Derotator. Assignment Perform a design study a derotator to prevent the smearing of the image with such a precision that an object falling on one pixel does not shift more then 1/5th (6.2μm) in one hour of observation time.

  3. Introduction (E-ELT,METIS) Problem definition Concept design Feasibility test Conclusion and remarks A design study of a Cryogenic High Accurate Derotator.

  4. A design study of a Cryogenic High Accurate Derotator. Introduction (E-ELT,METIS)

  5. A design study of a Cryogenic High Accurate Derotator. Introduction (E-ELT,METIS) • Mid-infrared E-ELT Imager and spectrograph • Imaging/spectroscopy in the mid infrared range (wavelengths of 2.9-14 µm) • Environment cryogenic and vacuum

  6. A design study of a Cryogenic High Accurate Derotator. Introduction (E-ELT,METIS) • Mid-infrared E-ELT Imager and spectrograph • Imaging/spectroscopy in the mid infrared range (wavelengths of 2.9-14 µm) • Environment cryogenic and vacuum

  7. A design study of a Cryogenic High Accurate Derotator. Why is derotationnecessary • The detector of the metis instrument needs and integration time of at least 15 minutes to get an high enough signal to noise ratio. • When not derotating smearing on the detector will occur, due to the rotation of the earth. • Because of the altitude azimuth configuration of the E-ELT the rotation can not be compensated by the telescope

  8. A design study of a Cryogenic High Accurate Derotator. Optical configuration

  9. A design study of a Cryogenic High Accurate Derotator. Influence of DOF • Mirror influences • 6 degrees of freedom for each mirror • Only 3 influence the science beam • Derotator • 6 degrees of freedom for the derotator • 4 influence the science beam

  10. A design study of a Cryogenic High Accurate Derotator. Influence of DOF • Mirror influences • 6 degrees of freedom for each mirror • Only 3 influence the science beam • Derotator • 6 degrees of freedom for the derotator • 4 influence the science beam

  11. A design study of a Cryogenic High Accurate Derotator. Influence of DOF • Derotator • Specific property when rotating around the specific rotation point • Red (1 degree) • Blue (2 degrees)

  12. A design study of a Cryogenic High Accurate Derotator. Concluding optical analysis • When defining the rotation point as shown in the previous slide the problem will reduce to a 3 DOF problem. • In the x-z plane the axis of the derotator needs to be directed to the rotation point (β) • The angle of the axis of the derotator in the z-y plane needs to be zero (α). • The rotation around the axis needs to be controlled to control the rate of derotation (γ)

  13. A design study of a Cryogenic High Accurate Derotator. Problem definition The end goal for the derotator is to prevent the smearing of the image with such a precision that an object falling on one pixel does not shift more then 1/5th (6.2μm) in one hour of observation time.

  14. A design study of a Cryogenic High Accurate Derotator. Requirements mirrors • Requirementsmirrors • Rotation: 2,6 arcsecond • Translation: 5 micrometer • Environmental aspects • Working temperature: 25-90 K • Working pressure 10-7-1 bar

  15. A design study of a Cryogenic High Accurate Derotator. Requirements Derotator • Requirements derotator • rotation αandβ: 2,6 arcseconds • rotationγ: 2 arcseconds • Maximum rotation speed: 7,5 degrees/hour • Minimum rotation speed: 0 degrees/hour • Setup speed: 90 degrees/minute • Needs to rotate in both directions • MTBF: 36500 hours • Maximum allowed weight: 30 Kg • Environmental aspects • Working temperature: 25-90 K • Working pressure 10-7-1 bar

  16. A design study of a Cryogenic High Accurate Derotator. Concept design Derotator

  17. A design study of a Cryogenic High Accurate Derotator. Concept design Derotator

  18. A design study of a Cryogenic High Accurate Derotator. Concept design Derotator

  19. A design study of a Cryogenic High Accurate Derotator. Feasibility test • Purpose of test setup • Test the feasibility of the used principles • Test the accuracy of the capacitive sensors • Kept as simple as possible • Modelled as a pendulum

  20. A design study of a Cryogenic High Accurate Derotator. Feasibility test Calculating PID values

  21. A design study of a Cryogenic High Accurate Derotator. Feasibility test • Closed loop bode plot • Phase margin in zero degrees up to 1 hertz

  22. A design study of a Cryogenic High Accurate Derotator. Feasibility test

  23. A design study of a Cryogenic High Accurate Derotator. Feasibility test

  24. A design study of a Cryogenic High Accurate Derotator. Test results Required accuracy: 2,6 arcseconds = 12 micro radians Signal that needs to be followed is of very low frequency • Applying a sinusoidal reference signal with: • Amplitude 400 micro radians • Frequency 0,02 Hz

  25. A design study of a Cryogenic High Accurate Derotator. Test results • Proven that an accuracy of 2,6 arcsecond is feasibly with a low disturbing frequency (up to 0.02 Hz) • This was done with relative simple building block (capacitive sensor; Voice coil actuator)

  26. A design study of a Cryogenic High Accurate Derotator. Conclusion and remarks • With optical analyses the 4 DOF system is reduced to a 2 DOF system • This 2 DOF can steer the science beam on the detector • Errors in the mirrors position can be compensated with the derotator reducing the requirements on the mirror • Relative easy to build a system with a 2,6 arcsecond angular accuracy at low frequencies

  27. A design study of a Cryogenic High Accurate Derotator. Conclusion and remarks • There is still some work that needs to be done • The test setup can be improved with a discrete controller such that error at higher frequencies is reduced • A trade of has to be made between the amount of heat dissipation and the required force to reduce error at higher frequencies • A thermal analyses of the system needs to be done

  28. A design study of a Cryogenic High Accurate Derotator. Questions