slide1
Download
Skip this Video
Download Presentation
Measurements of CO 2 Molar Mixing Ratio by Infrared Absorption Spectroscopy C, mole fraction ( μ mole mole -1 = ppm)

Loading in 2 Seconds...

play fullscreen
1 / 12

Measurements of CO2 Molar Mixing Ratio by Infrared Absorption Spectroscopy C, mole fraction mole mole-1 ppm LI-C - PowerPoint PPT Presentation


  • 224 Views
  • Uploaded on

Measurements of CO 2 Molar Mixing Ratio by Infrared Absorption Spectroscopy C, mole fraction ( μ mole mole -1 = ppm) LI-COR analyzers measure absorbance at 4.26 μ m V = kA (voltage proportional to absorbance) ρ c = P c /RT = CP/RT (molar density of absorber)

loader
I am the owner, or an agent authorized to act on behalf of the owner, of the copyrighted work described.
capcha
Download Presentation

PowerPoint Slideshow about 'Measurements of CO2 Molar Mixing Ratio by Infrared Absorption Spectroscopy C, mole fraction mole mole-1 ppm LI-C' - grover


An Image/Link below is provided (as is) to download presentation

Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author.While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server.


- - - - - - - - - - - - - - - - - - - - - - - - - - E N D - - - - - - - - - - - - - - - - - - - - - - - - - -
Presentation Transcript
slide1
Measurements of CO2 Molar Mixing Ratio

by Infrared Absorption Spectroscopy

C, mole fraction (μmole mole-1 = ppm)

LI-COR analyzers measure absorbance at 4.26 μm

V = kA (voltage proportional to absorbance)

ρc = Pc/RT = CP/RT (molar density of absorber)

ρcL = CPL/RT (absorber amount in absorption cell)

slide5

Virial Equation of State (REAL not IDEAL):

PV = nRT(1 + nB(T)/V + nC(T)/V2 + …)

neglect higher order terms

Solve for n:

n = -(V/2B){1 – (1 + 4PB/RT)1/2}

B (10-6 m3 mole-1) (in air at T = 300 K)

air -7.7

CO2 -42.3

Using this, we find the equivalence of (at 300 K and 1 Bar):

volume mixing ratio 370.0 x 10-6 m3 CO2 / m3 air

molar mixing ratio 370.5 x 10-6 mole CO2 / mole air

slide6
LiCor Analyzer Response Curve:

C = [a0 + a1 (V P0/P) + a2 (V P0/P)2] T/T0

C is CO2 mole fraction

P0, T0 are pressure and temperature during calibration

slide8
Pressure Broadening

effective pressure: Pe = PN2 + Σbi Pi

Gas Coef (bi) % of air

N2 1.00 78.084

O2 0.81 20.946

Ar 0.78 0.934

H2O ≈1.57 ≈1

slide9

Pressure Broadening

Example calibration curve in air (PO2 = 20 kPA, PN2 = 80 kPA):

Pe = 80 kPa x 1.00 + 20 kPa x 0.81 = 96.2 kPa

C = 326.61 + 0.1738 V + 4.5095 x 10-5 V2

A LiCor response of 300 mV implies CO2 = 382.82 ppm

What if you calibrated using pure N2? (PN2 = Pe = 100 kPa)

C = 326.61 + 0.1807 V + 4.8723 x 10-5 V2

Now a response of 300 mV gives CO2 = 385.21 ppm !

What error is made using synthetic (no Argon) vs. real air?

slide10
Water Vapor
  • Pressure broadening of CO2 line

bH2O≈ 1.57

  • Dilution of air – adding 1% H2O displaces 1% of CO2 (≈ 3.7 ppm)

Solution: DRY THE AIR! How dry is dry enough?

0.1 ppm / 370 ppm = 0.027% → H2O ≤ 0.00027 m3/m3 air

Error in CO2Dewpoint

0.1 ppm -36.5oC

0.2 ppm -29.4oC

slide12

The results differ for different LiCor CO2

analyzers under the same conditions!!!

ad