1 / 16

FDA Recommended Study Design for CABG Anastomosis Devices

Cardica Proximal Anastomosis System 510k 0304034 Wolf Sapirstein MPH, FACS Clinical Review FDA/CDRH/ODE. FDA Recommended Study Design for CABG Anastomosis Devices. Observational Study 510K clearance Effectiveness: LCB for 95% CI, 80% angiographic patency at 6-months

Download Presentation

FDA Recommended Study Design for CABG Anastomosis Devices

An Image/Link below is provided (as is) to download presentation Download Policy: Content on the Website is provided to you AS IS for your information and personal use and may not be sold / licensed / shared on other websites without getting consent from its author. Content is provided to you AS IS for your information and personal use only. Download presentation by click this link. While downloading, if for some reason you are not able to download a presentation, the publisher may have deleted the file from their server. During download, if you can't get a presentation, the file might be deleted by the publisher.

E N D

Presentation Transcript


  1. Cardica Proximal Anastomosis System510k 0304034Wolf Sapirstein MPH, FACSClinical ReviewFDA/CDRH/ODE FDA Clinical Presentation

  2. FDA Recommended StudyDesign for CABG AnastomosisDevices • Observational Study • 510K clearance • Effectiveness: • LCB for 95% CI, 80% angiographic patency at 6-months • One year clinical evaluation with Stress-ECG. • Safety: • Procedure adverse events. Adverse events to 1 year • MACE (Mortality; TVR; Myocardial infarction) • Control: • Sutured anastomosis • Historical • FitzGibbon FDA Clinical Presentation

  3. PAS-Port Study 1 • Enrolled 55 patients • Failed Use 8 patients • Implanted 47 patients; 54 devices • Not Evaluated 2 withdrew; 1 discontinued • Angiograms 39 in 44 patients; 42 grafts 82.9% cases implanted • Patent grafts 36 of 42 grafts 85.7%, LCB 73.7% • Not followed 2 Deaths; 3 discontinued • Stress-ECG 36 of 42 patients/1 positive 78.2% living FDA Clinical Presentation

  4. PAS-Port Study 1Effectiveness • Observed Patency (6-Month Angio.) • 36 of 42 implants in 39 patients 85.7%, LCB 73.7% • Imputed Patency • 43 of 50 implants in 47 implanted patients 86.0%, LCB 75.3% • Intention to treat (Imputed) • CABG procedures patent in 50 of 60 grafts 83.3%, LCB 71.4% • Device grafts patent in 43 of 60 grafts 71.1%, LCB 59.0% FDA Clinical Presentation

  5. PAS-Port Study 1Safety • MACE (12/24 months 42 patients) • Deaths 2 (4.7%) • Myocardial Infarction 1 (2.3%) • Graft occlusion 6 (14%) • K-M estimate of rate at 24 Months: 20.2% (UCB 30.1%) of 47 patients per protocol 17.5% (UCB 26.2%) of 55 intent to treat patients • Non-Device Adverse events • Myocardial Ischemia 3; Infarction 1; Revasc 4 • Hospital event rates similar to historical data bases FDA Clinical Presentation

  6. Enrolled 54 patients (133 Consented and 118 enrolled in parent study) PAS-Port Intended 63 devices Failed Implant 2 patients Implanted 52 patients, 59 devices Hospital Death 2 patients Discontinued 4 patients Six-month Angio. 38 of 46 patients 45 of 47 grafts patent One-year Follow-up: 46 patients One-year Stress-ECG: 42 patients Negative Stress test 38 patients PAS-Port StudyCohort 2 FDA Clinical Presentation

  7. PAS-Port Study Cohort 2Effectiveness • Observed Patency (6-month angio) • 45 of 47 grafts patent in 38 patients • 95.7%, LCB 87.2% • Imputed Patency (All device grafts) • 55 of 59 grafts • 93.0%, LCB 85% • Imputed Patency (Intent to treat) • 57 of 63 CABG procedures (all intended) • 90.5%, LCB 80% FDA Clinical Presentation

  8. PAS-Port Study Cohort 2Safety • MACE • Death 4 (7.6%) • Myocardial Infarction 0 • Graft Occlusion/stenosis 5 (9.6%) • K-M estimated rates at 1 year: • Per Protocol (50 patients) 12.7% (UCB 20.7%) • Intent to Treat (54 patients) 15.5% (UCB 23.6%) • Adverse Events • Myocardial Ischemia 6 (11.5%) • Revascularization 9 (17.3%) • Graft stenosis 11 (21.1%) FDA Clinical Presentation

  9. PAS-PortCombined Study 1 and Cohort 2 • Enrollment 109 patients • Failed Implants 12 patients • Implants 97 patients • Hospital deaths 2 patients • Lost follow-up 7 patients • 6-month Angio 77 patients (90 grafts) • Angio patent 81 grafts (89 grafts) • Imputed Patency 17 patients • MSCT/MRI patent 10 patients • Stress-ECG negative 5 patients • CCS 2 patients • 12/24 Month Stress-ECG 78 patients • Ischemic 5 patients FDA Clinical Presentation

  10. PAS-PortCombined Study 1 and Cohort 2 Effectiveness • Observed Patency for 99 patients • 81 of 89 grafts in 77 patients • 91%, LCB 84% • Imputed Patency Device for 107 patients • 98 of 109 grafts • 89.9%, LCB 83.8% • Imputed Patency Intent to Treat (grafts) • 107 patent CABG grafts in 123 grafts • 87%, LCB FDA Clinical Presentation

  11. PAS-Port Patency Imputation • Study 1: • Implants (47 patients) • Angio: 39 patients evaluated 36 (85.7%) of 42 grafts patent 72% of all 50 implants • MRI 5 patients • Stress ECG 2 patients • Died 1 patient • Suture Conversions (8 patients) • Stress ECG 7 patients • CCS 1 patient FDA Clinical Presentation

  12. PAS-PortImputed Patency • Cohort 2 • Implants (50 patients) • Angio 38 patients (73.7 % all implants) 45 grafts of 47 implants (95.7%) • MSCT 5 patients • Stress-ECG 3 patients • CCS 2 patients • Suture Conversions (2 patients) • CCS 2 patients FDA Clinical Presentation

  13. Study Design Limitations • Modification to study protocols after study initiation • Cohort 2 sample stratified from a study enrolled under a different protocol for a different CABG device • Problems with Pooling of Study 1 and Cohort 2: • Inclusion criteria different • Cohort 2 retrospective enrollment • Cohort 2 employed 2 experimental devices • Baseline variability • Instrument modification FDA Clinical Presentation

  14. Safety and Effectiveness Concerns • Angiography • Study 1: 39 of 47 patients implanted (82.9%) • Cohort 2: 38 of 52 patients implanted (73%) • Imputation of Patency (Combined groups) • 20 of 97 implanted patients (20%) • 26 of 109 intent to treat (23.8%) • Imputation Criteria (Combine groups) • MRI/MSCT 10 • Stress ECG 11 • CCS 5 FDA Clinical Presentation

  15. Safety and Effectiveness Concerns (cont’d) • Stress ECG • 36 of 45 survivors  1 Ischemia • 39 of 48 survivors  4 Ischemia • Thrombo-Embolism • Two events • Stenosis/Occlusion • 11 Stenoses • 9 Ischemia • 1 Myocardial infarction • MACE • PAS-Port CABG supplied <= 22% Left Ventricle FDA Clinical Presentation

  16. PAS-Port Anastomosis System Study Conclusions • Study 1 did not meet the pre-specified effectiveness hypothesis. • The poolability of Study 1 and Cohort 2 is problematic. • Angiographic follow-up of patients is incomplete. The Sponsor relied on imputations with non-invasive technologies to assess patency. • The attribution of several adverse events as “non-device related” is questionable. FDA Clinical Presentation